Comparative study of sub-second temporal resolution 4D-MRI and 4D-CT for target motion assessment in a phantom model

被引:1
|
作者
Wang, Tianyuan [1 ]
Sofue, Keitaro [2 ]
Shimada, Ryuji [3 ]
Ishihara, Takeaki [1 ]
Yada, Ryuichi [1 ]
Miyamoto, Masanori [3 ]
Sasaki, Ryohei [1 ]
Murakami, Takamichi [2 ]
机构
[1] Kobe Univ Hosp, Dept Radiat Oncol, Kobe, Japan
[2] Kobe Univ, Grad Sch Med, Dept Radiol, 7-5-2 Kusunoki Cho,Chuo Ku, Kobe 6500017, Japan
[3] Kobe Univ Hosp, Ctr Radiol & Radiat Oncol, Kobe, Japan
基金
日本学术振兴会;
关键词
GD-EOB-DTPA; RESPIRATORY MOTION; HEPATOCELLULAR-CARCINOMA; RADIATION-THERAPY; LIVER-TUMORS; LESION SIZE; MRI; RADIOTHERAPY; CT; RECONSTRUCTION;
D O I
10.1038/s41598-023-42773-z
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
To develop and investigate the feasibility of sub-second temporal resolution volumetric T1-weighted four-dimensional (4D-) MRI in comparison with 4D-CT for respiratory-correlated motion assessment using an MRI/CT-compatible phantom. Sub-second high temporal resolution (0.5 s) gradient-echo T1-weighted 4D-MRI was developed using a volumetric acquisition scheme with compressed sensing. An MRI/CT-compatible motion phantom (simulated liver tumor) with three sinusoidal movements of amplitudes and two respiratory patterns was introduced and imaged with 4D-MRI and 4D-CT to investigate the geometric accuracy of the target movement. The geometric accuracy, including centroid position, volume, similarity index of dice similarity coefficient (DSC), and Hausdorff distance (HD), was systematically evaluated. Proposed 4D-MRI achieved a similar geometric accuracy compared with 4D-CT regarding the centroid position, volume, and similarity index. The observed position differences of the absolute average centroid were within 0.08 cm in 4D-MRI and 0.03 cm in 4D-CT, less than the 1-pixel resolution for each modality. The observed volume difference in 4D-MRI/4D-CT was within 0.73 cm3 (4.5%)/0.29 cm3 (2.1%) for a large target and 0.06 cm3 (11.3%)/0.04 cm3 (11.6%) for a small target. The observed DSC values for 4D-MRI/4D-CT were at least 0.93/0.95 for the large target and 0.83/0.84 for the small target. The maximum HD values were 0.25 cm/0.31 cm for the large target and 0.21 cm/0.15 cm for the small target. Although 4D-CT potentially exhibit superior numerical accuracy in phantom studies, the proposed high temporal resolution 4D-MRI demonstrates sub-millimetre geometric accuracy comparable to that of 4D-CT. These findings suggest that the 4D-MRI technique is a viable option for characterizing motion and generating phase-dependent internal target volumes within the realm of radiotherapy.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Comparison of the radiotherapy plan with the image data approach of a conventional and 4D-CT for movement target volume:: A phantom study
    Block, A.
    Noertersheuser, T.
    Waletzko, O.
    Rohn, R.
    Mathias, K.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2008, 184 : 41 - 41
  • [22] Assessment of motion and SUV recovery in 3D and 4D PET/CT: a multicentre phantom study
    Marin, K. J. Ortega
    Lambrecht, M.
    Arends, A. J.
    Adebahr, S.
    Nestle, U.
    Vogel, W.
    Verheij, M.
    Sonke, J. J.
    Hurkmans, C. W.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S540 - S540
  • [23] Development of a Computerized 4-D MRI Phantom for Liver Motion Study
    Wang, Chunhao
    Yin, Fang-Fang
    Segars, W. P.
    Chang, Zheng
    Ren, Lei
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 1051 - 1059
  • [24] Development of a Computerized 4D MRI Phantom for Liver Motion Study
    Wang, C.
    Yin, F.
    Segars, P.
    Chang, Z.
    Ren, L.
    MEDICAL PHYSICS, 2017, 44 (06) : 3027 - 3027
  • [25] 4D-CT Simulation Using Individually Optimized Contrast Enhancement (CE): A Phantom Study
    Xue, M.
    D'Souza, W.
    Lasio, G.
    Klahr, P.
    Lu, W.
    MEDICAL PHYSICS, 2012, 39 (06) : 3903 - 3903
  • [26] Is a single respiratory correlated 4D-CT study sufficient for evaluation of breathing motion?
    Guckenberger, Matthias
    Wilbert, Juergen
    Meyer, Juergen
    Baier, Kurt
    Richter, Anne
    Flentje, Michael
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05): : 1352 - 1359
  • [27] Target volume and motion position evaluation of four-dimensional cone-beam CT: comparison with 4D-CT using dynamic thorax phantom
    Doi, Yasuhiro
    Shimohigashi, Yoshinobu
    Yotsuji, Yohei
    Maruyama, Masato
    Kai, Yudai
    Toya, Ryo
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (03):
  • [28] Quantitative evaluation of a new 4D MRI sequence: a motion phantom study
    Parent, L.
    Nemtanu, T.
    Celicanin, Z.
    Bieri, O.
    Cattin, P.
    Ken, S.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1210 - S1210
  • [29] Interactive 4D-(3D+animation) cardiac CT: A new application ECG-gated sub-second helical CT
    Higashino, H
    Mochizuki, T
    Murase, K
    Miyagawa, M
    Miki, H
    Ikezoe, J
    RADIOLOGY, 1999, 213P : 136 - 137
  • [30] A Spatiotemporal-Constrained Sorting Method for Motion-Robust 4D-MRI: A Feasibility Study
    Wang, Chunhao
    Subashi, Ergys
    Yin, Fang-Fang
    Chang, Zheng
    Cai, Jing
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 103 (03): : 758 - 766