SAFIR-I: first NEMA NU 4-2008-based performance characterization

被引:0
|
作者
Bebie, Pascal [1 ]
Lustermann, Werner [1 ]
Debus, Jan [1 ]
Ritzer, Christian [1 ,4 ]
Dissertori, Guenther [1 ]
Weber, Bruno [2 ,3 ]
机构
[1] Swiss Fed Inst Technol, Inst Particle Phys & Astrophys, Otto Stern Weg 5, CH-8093 Zurich, Switzerland
[2] Univ Zurich, Inst Pharmacol & Toxicol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[3] Univ Zurich, Neurosci Ctr Zurich, Winterthurerstr 190, CH-8057 Zurich, Switzerland
[4] Swiss Fed Inst Technol, Winterthurerstr 190, CH-8057 Zurich, Switzerland
关键词
Positron Emission Tomography; PET/MRI; Preclinical imaging; SAFIR; NEMA NU 4-2008; Performance characterization; PET;
D O I
10.1186/s40658-023-00603-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Small Animal Fast Insert for MRI detector I (SAFIR-I) is a novel Positron Emission Tomography insert for a 7 T Bruker BioSpec 70/30 Ultra Shield Refrigerated Magnetic Resonance Imaging (MRI) system. It facilitates truly simultaneous quantitative imaging in mice and rats at injected activities as high as 500 MBq. Exploitation of the resulting high count rates enables quick image formation at few seconds per frame. In this investigation, key performance parameters of SAFIR-I have been determined according to the evaluations outlined in the National Electrical Manufacturers Association (NEMA) Standards Publication NU 4-2008 (NEMA-NU4) protocol. Results: Using an energy window of 391 to 601 keV and a Coincidence Timing Window of 500 ps, the following performance was observed: The average spatial resolution at 5mm radial offset (Full Width at Half Maximum) is 2.54mm when using Filtered Backprojection, 3D Reprojection reconstruction. For the mouse- and rat-like phantoms, the maximal Noise-Equivalent Count Rates (NECRs) are 1368 kcps at the highest tested average effective concentration of 14.7 MBq cc-1, and 713 kcps at the highest tested average effective concentration of 1.72 MBq cc-1, respectively. The NECR peak is not yet reached for either of these cases. The peak sensitivity is 1.46%. The Image Quality phantom uniformity standard deviation is 4.8%. The Recovery Coefficient for the 5mm rod is (1.08 +/- 0.10). The Spill-Over Ratios are (0.22 +/- 0.03) and (0.22 +/- 0.02), for the waterand air-filled cylinder, respectively. An accuracy of 4.3% was achieved for the quantitative calibration of reconstructed voxel values. Conclusions: The measured performance parameters indicate that the various design goals have been achieved. SAFIR-I offers excellent performance, especially at the high activities it was designed for. This facilitates planned experiments with fast tracer kinetics in small animals. Ways to potentially improve performance can still be explored. Simultaneously, further performance gains can be expected for a forthcoming insert featuring 2.7 times longer axial coverage named Small Animal Fast Insert for MRI detector II (SAFIR-II).
引用
收藏
页数:20
相关论文
共 50 条
  • [1] SAFIR-I: first NEMA NU 4-2008-based performance characterization
    Pascal Bebié
    Werner Lustermann
    Jan Debus
    Christian Ritzer
    Günther Dissertori
    Bruno Weber
    EJNMMI Physics, 10
  • [2] Comparison of NEMA NU 4-2008 vs NEMA NU 2-2001 for the performance evaluation of the microPET R4 system
    Popota, F. D.
    Aguiar, P.
    Fernandez, Y.
    Lois, C.
    Pareto, D.
    Ros, D.
    Pavia, J.
    Gispert, J. D.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2706 - +
  • [3] SAFIR-I: Design and Performance of a High-Rate Preclinical PET Insert for MRI
    Bebie, Pascal
    Becker, Robert
    Commichau, Volker
    Debus, Jan
    Dissertori, Gunther
    Djambazov, Lubomir
    Eleftheriou, Afroditi
    Fischer, Jannis
    Fischer, Peter
    Ito, Mikiko
    Khateri, Parisa
    Lustermann, Werner
    Ritzer, Christian
    Ritzert, Michael
    Roser, Ulf
    Tsoumpas, Charalampos
    Warnock, Geoffrey
    Weber, Bruno
    Wyss, Matthias T.
    Zagozdzinska-Bochenek, Agnieszka
    SENSORS, 2021, 21 (21)
  • [4] Performance Evaluation of a Newly Designed SiPM-based Preclinical Scanner Based on NEMA-NU4 2008
    Amirrashedi, M.
    Sarkar, S.
    Ghafarian, P.
    Shahraki, R. Hashemi
    Geramifar, P.
    Zaidi, H.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S813 - S814
  • [5] Variations on the NEMA NU4-2008 Testing Procedures and Effect on the Performance Measurement Results
    Elhami, Esmat
    Boulding, Josh
    Goertzen, Andrew L.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3244 - 3246
  • [6] NEMA NU-4 2008 performance evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner
    Amirrashedi, Mahsa
    Sarkar, Saeed
    Ghafarian, Pardis
    Shahraki, Reza Hashemi
    Geramifar, Parham
    Zaidi, Habib
    Ay, Mohammad Reza
    MEDICAL PHYSICS, 2019, 46 (11) : 4816 - 4825
  • [7] Performance evaluation of the Mediso PET scanner to image three mice simultaneously based on the NEMA NU4-2008 standards
    Li, Guiyu
    Ma, Xiaowei
    Wang, Jing
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [8] NEMA NU 4-2008 Comparison of Preclinical PET Imaging Systems
    Goertzen, Andrew L.
    Bao, Qinan
    Bergeron, Melanie
    Blankemeyer, Eric
    Blinder, Stephan
    Canadas, Mario
    Chatziioannou, Arion F.
    Dinelle, Katherine
    Elhami, Esmat
    Jans, Hans-Sonke
    Lage, Eduardo
    Lecomte, Roger
    Sossi, Vesna
    Surti, Suleman
    Tai, Yuan-Chuan
    Jose Vaquero, Juan
    Vicente, Esther
    Williams, Darin A.
    Laforest, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (08) : 1300 - 1309
  • [9] Comparison of the Performance Evaluation of the MicroPET R4 Scanner According to NEMA Standards NU 4-2008 and NU 2-2001
    Popota, Fotini D.
    Aguiar, Pablo
    Raul Herance, J.
    Pareto, Deborah
    Rojas, Santiago
    Ros, Domenec
    Pavia, Javier
    Domingo Gispert, Juan
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) : 1879 - 1886
  • [10] Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs
    Patrick Hallen
    David Schug
    Volkmar Schulz
    EJNMMI Physics, 7