Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra

被引:5
|
作者
Zeimpekis, Konstantinos G. [1 ]
Mercolli, Lorenzo [1 ]
Conti, Maurizio [2 ]
Sari, Hasan [3 ]
Prenosil, George [1 ]
Shi, Kuangyu [1 ]
Rominger, Axel [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Nucl Med, Inselspital, Bern, Switzerland
[2] Siemens Healthineers, Mol Imaging, Knoxville, TN USA
[3] Siemens Healthcare AG, Adv Clin Imaging Technol, Lausanne, Switzerland
关键词
Whole-body PET; CT; Yttrium-90; NEMA; Image quality; Quadra; RADIOEMBOLIZATION; DOSIMETRY; PET/CT; PERFORMANCE; QUALITY;
D O I
10.1007/s00259-022-06074-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The image quality characteristics of two NEMA phantoms with yttrium-90 (Y-90) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after Y-90 radioembolization.Methods:Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 x 440 and 220 x 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR.Results: The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 x 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively.Conclusions: This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with Y-90. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR.
引用
收藏
页码:1168 / 1182
页数:15
相关论文
共 50 条
  • [41] Multi institutional quantitative phantom study of yttrium-90 PET in PET/MRI: the MR-QUEST study
    Maughan, Nichole M.
    Eldib, Mootaz
    Faul, David
    Conti, Maurizio
    Elschot, Mattijs
    Knesaurek, Karin
    Leek, Francesca
    Townsend, David
    DiFilippo, Frank P.
    Jackson, Kimberly
    Nekolla, Stephan G.
    Lukas, Mathias
    Tapner, Michael
    Parikh, Parag J.
    Laforest, Richard
    EJNMMI PHYSICS, 2018, 5
  • [42] Multi institutional quantitative phantom study of yttrium-90 PET in PET/MRI: the MR-QUEST study
    Nichole M. Maughan
    Mootaz Eldib
    David Faul
    Maurizio Conti
    Mattijs Elschot
    Karin Knešaurek
    Francesca Leek
    David Townsend
    Frank P. DiFilippo
    Kimberly Jackson
    Stephan G. Nekolla
    Mathias Lukas
    Michael Tapner
    Parag J. Parikh
    Richard Laforest
    EJNMMI Physics, 5
  • [43] Voxel-Based Dosimetry and Validation of Yttrium-90 Microsphere Therapy Using Emission Tomography Imaging
    Potrebko, P.
    Shridhar, R.
    Biagioli, M.
    Cross, M.
    Lazarakis, T.
    Andl, G.
    Fox, T.
    MEDICAL PHYSICS, 2017, 44 (06) : 3154 - 3154
  • [44] Evaluation of the delivered activity of yttrium-90 resin microspheres using sterile water and 5 % glucose during administration
    Hojjat Ahmadzadehfar
    Carsten Meyer
    Claus Christian Pieper
    Ralph Bundschuh
    Marianne Muckle
    Florian Gärtner
    Hans Heinz Schild
    Markus Essler
    EJNMMI Research, 5
  • [45] Yttrium-90 radibembolization using TheraSphere® in the management of primary and secondary liver tumors
    Riaz, A.
    Lewandowski, R. J.
    Kulik, L.
    Salem, R.
    QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 53 (03): : 311 - 316
  • [46] Radioembolization of Symptomatic, Unresectable Neuroendocrine Hepatic Metastases Using Yttrium-90 Microspheres
    Philipp M. Paprottka
    Ralf-T. Hoffmann
    Alexander Haug
    Wieland H. Sommer
    Franziska Raeßler
    Christoph G. Trumm
    Gerwin P. Schmidt
    Nima Ashoori
    Maximilian F. Reiser
    Tobias F. Jakobs
    CardioVascular and Interventional Radiology, 2012, 35 : 334 - 342
  • [47] Evaluation of Early Response to Treatment of Hepatocellular Carcinoma with Yttrium-90 Radioembolization Using Quantitative Computed Tomography Analysis
    Kim, Sungwon
    Kim, Do-Young
    An, Chansik
    Han, Kyunghwa
    Won, Jong Yun
    Kim, Gyoung Min
    Kim, Myeong-Jin
    Choi, Jin-Young
    KOREAN JOURNAL OF RADIOLOGY, 2019, 20 (03) : 449 - 458
  • [48] Persistent synovitis treated with radiation synovectomy using yttrium-90: A retrospective evaluation of 83 procedures for 45 patients
    Jahangier, ZN
    Jacobs, JWG
    VanIsselt, JW
    Bijlsma, JWJ
    BRITISH JOURNAL OF RHEUMATOLOGY, 1997, 36 (08): : 861 - 869
  • [49] Radioembolization Using YTTRIUM-90 Microspheres for Advanced Primary and Metastatic Liver Malignancies
    Uslu, L.
    Arslan, E.
    Ozhan, M.
    Nisli, S.
    Sager, S.
    Kabasakal, L.
    Gulsen, F.
    Cantasdemir, M.
    Numan, F.
    Uslu, I.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S372 - S373
  • [50] LOCO-REGIONAL TREATMENT OF BRAIN TUMORS USING YTTRIUM-90 COLLOID
    Chiti, A.
    Gaetani, P.
    Tosi, G.
    Scorsetti, M.
    ANTICANCER RESEARCH, 2004, 24 (5D) : 3453 - 3453