Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra

被引:5
|
作者
Zeimpekis, Konstantinos G. [1 ]
Mercolli, Lorenzo [1 ]
Conti, Maurizio [2 ]
Sari, Hasan [3 ]
Prenosil, George [1 ]
Shi, Kuangyu [1 ]
Rominger, Axel [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Nucl Med, Inselspital, Bern, Switzerland
[2] Siemens Healthineers, Mol Imaging, Knoxville, TN USA
[3] Siemens Healthcare AG, Adv Clin Imaging Technol, Lausanne, Switzerland
关键词
Whole-body PET; CT; Yttrium-90; NEMA; Image quality; Quadra; RADIOEMBOLIZATION; DOSIMETRY; PET/CT; PERFORMANCE; QUALITY;
D O I
10.1007/s00259-022-06074-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The image quality characteristics of two NEMA phantoms with yttrium-90 (Y-90) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after Y-90 radioembolization.Methods:Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 x 440 and 220 x 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR.Results: The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 x 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively.Conclusions: This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with Y-90. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR.
引用
收藏
页码:1168 / 1182
页数:15
相关论文
共 50 条
  • [21] Radioembolization of Renal Cell Carcinoma Using Yttrium-90 Microspheres
    Hamoui, Nabeel
    Gates, Vanessa L.
    Gonzalez, Jose
    Lewandowski, Robert J.
    Salem, Riad
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (02) : 298 - 300
  • [22] Yttrium-90 radioembolization using MIRD dosimetry with resin microspheres
    Sarwar, Ammar
    Kudla, Alexei
    Weinstein, Jeffrey L.
    Ali, Aamir
    Malik, Raza
    Bullock, Andrea
    Khwaja, Khalid O.
    Curry, Michael
    Faintuch, Salomao
    Ahmed, Muneeb
    EUROPEAN RADIOLOGY, 2021, 31 (03) : 1316 - 1324
  • [23] Image quality analysis for PET imaging using Yttrium-90
    Molina-Duran, F.
    Attarwala, A.
    Buesing, K.
    Tapner, M.
    Schoenberg, S.
    Willowson, K.
    Glatting, G.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2014, 190 : 148 - 148
  • [24] Radioassay of yttrium-90 radiation using the radionuclide dose calibrator
    Salako, QA
    DeNardo, SJ
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 723 - 726
  • [25] Reproducibility of Yttrium-90 activity determined using PET/CT
    Doss, Mohan
    Yu, Jian
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [26] Yttrium-90 radioembolization using MIRD dosimetry with resin microspheres
    Ammar Sarwar
    Alexei Kudla
    Jeffrey L. Weinstein
    Aamir Ali
    Raza Malik
    Andrea Bullock
    Khalid O. Khwaja
    Michael Curry
    Salomao Faintuch
    Muneeb Ahmed
    European Radiology, 2021, 31 : 1316 - 1324
  • [27] Quantitative Yttrium-90 PET/CT acquisition and reconstruction optimization: An anthropomorphic phantom validation.
    Gnesin, S.
    Paterne, L.
    Boubaker, A.
    Adib, S.
    Pappon, M.
    Kosinski, M.
    Prior, J.
    Baechler, S.
    Verdun, F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S150 - S150
  • [28] Radioembolization of Colorectal Hepatic Metastases Using Yttrium-90 Microspheres
    Mulcahy, Mary F.
    Lewandowski, Robert J.
    Ibrahim, Saad M.
    Sato, Kent T.
    Ryu, Robert K.
    Atassi, Bassel
    Newman, Steven
    Talamonti, Mark
    Omary, Reed A.
    Benson, Al, III
    Salem, Riad
    CANCER, 2009, 115 (09) : 1849 - 1858
  • [29] A TECHNIQUE FOR DESTRUCTION OF PINEAL BODY USING YTTRIUM-90 SPHERES
    HOLMGREN, U
    NOTTER, G
    NATURE, 1961, 189 (476) : 773 - &
  • [30] Use of a Gamma Camera for the Evaluation of the In-Vivo Distribution of YTTRIUM-90
    Abulizi, M.
    Zanotti-Fregonara, P.
    Dumont, A.
    Askienazy, S.
    Devaux, J. Y.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S69 - S69