Feasibility of photon-counting CT for femoroacetabular impingement syndrome evaluation: lower radiation dose and improved diagnostic confidence
被引:6
|作者:
Ferrero, Andrea
[1
]
Powell, Garret M.
[1
]
Adaaquah, Dennis K.
[1
]
Rajendran, Kishore
[1
]
Thorne, Jamison E.
[1
]
Krych, Aaron J.
[2
]
Horst, Kelly K.
[1
]
McCollough, Cynthia H.
[1
]
Baffour, Francis I.
[1
]
机构:
[1] Mayo Clin, Dept Radiol, 200 1st St SW, Rochester, MN 55902 USA
[2] Mayo Clin, Dept Orthoped Surg & Sports Med, Rochester, MN USA
关键词:
Photon-counting CT;
Femoroacetabular impingement;
Radiation dose reduction;
Pelvis CT;
COMPUTED-TOMOGRAPHY;
HIP ARTHROSCOPY;
COMMON;
D O I:
10.1007/s00256-023-04325-4
中图分类号:
R826.8 [整形外科学];
R782.2 [口腔颌面部整形外科学];
R726.2 [小儿整形外科学];
R62 [整形外科学(修复外科学)];
学科分类号:
摘要:
Objective The feasibility of low-dose photon-counting detector (PCD) CT to measure alpha and acetabular version angles of femoroacetabular impingement (FAI). Material and methods FAI patients undergoing an energy-integrating detector (EID) CT underwent an IRB-approved prospective ultra-high-resolution (UHR) PCD-CT between 5/2021 and 12/2021. PCD-CT was dose-matched to the EID-CT or acquired at 50% dose. Simulated 50% dose EID-CT images were generated. Two radiologists evaluated randomized EID-CT and PCD-CT images and measured alpha and acetabular version angles on axial image slices. Image quality (noise, artifacts, and visualization of cortex) and confidence in non-FAI pathology were rated on a 4-point scale (3 = adequate). Preference tests of standard dose PCD-CT, 50% dose PCD-CT, and 50% dose EID-CT relative to standard dose EID-CT were performed using Wilcoxon Rank test. Results 20 patients underwent standard dose EID-CT (similar to CTDIvol, 4.5 mGy); 10 patients, standard dose PCD-CT (4.0 mGy); 10 patients, 50% PCD-CT (2.6 mGy). Standard dose EID-CT images were scored as adequate for diagnostic task in all categories (range 2.8-3.0). Standard dose PCD-CT images scored higher than the reference in all categories (range 3.5-4, p < 0.0033). Half-dose PCD-CT images also scored higher for noise and cortex visualization (p < 0.0033) and equivalent for artifacts and visualization of non-FAI pathology. Finally, simulated 50% EID-CT images scored lower in all categories (range 1.8-2.4, p < 0.0033). Conclusions Dose-matched PCD-CT is superior to EID-CT for alpha angle and acetabular version measurement in the work up of FAI. UHR-PCD-CT enables 50% radiation dose reduction compared to EID while remaining adequate for the imaging task.
引用
收藏
页码:1651 / 1659
页数:9