An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of methotrexate in human serum and plasma

被引:6
|
作者
Engel, Anett [1 ]
Ruhe, Lena [1 ]
Singh, Neeraj [2 ]
Wright, Jo Anne [2 ]
Liesch, Franziska [2 ]
Bauland, Friederike [3 ]
Ostermann, Annika I. I. [1 ]
Sumalowitsch, Tamara [1 ]
Schweinsberg, Vincent J. T. [4 ]
Geistanger, Andrea [2 ]
Hegel, Johannes Kolja [1 ]
Geletneky, Christian [2 ]
Taibon, Judith [2 ,5 ]
机构
[1] Lab Berlin Charite Vivantes Serv GmbH, Collaborat & Innovat Management, Dept Studies, Berlin, Germany
[2] Roche Diagnost GmbH, Berlin, Germany
[3] Chrestos Concept GmbH & Co KG, Essen, Germany
[4] Lab Berlin Charite Vivantes GmbH, Dept Lab Med & Toxicol, Berlin, Germany
[5] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
关键词
isotope dilution LC-MS/MS; methotrexate; qNMR; reference measurement procedure; SI units; traceability; LC-MS/MS ASSAY; PHARMACOKINETICS; VALIDATION; METABOLITE; CHILDREN;
D O I
10.1515/cclm-2022-1001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To develop an isotope dilution-liquid chromatography-tandem mass spectrometry-(ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for quantification of methotrexate in human serum and plasma.Methods: Quantitative nuclear magnetic resonance (qNMR) was used to determine absolute methotrexate content in the standard. Separation was achieved on a biphenyl reversed phase analytical column with mobile phases based on water and acetonitrile, both containing 0.1% formic acid. Sample preparation included protein precipitation in combination with high sample dilution, and method validation according to current guidelines. The following were assessed: selectivity (using analyte-spiked samples, and relevant structural-related compounds and interferences); specificity and matrix effects (via post-column infusion and comparison of human matrix vs. neat samples); precision and accuracy (in a five-day validation analysis). RMP results were compared between two independent laboratories. Measurement uncertainty was evaluated according to current guidelines.Results: The RMP separated methotrexate from potentially interfering compounds and enabled measurement over a calibration range of 7.200-5,700 ng/mL (0.01584-12.54 mu mol/L), with no evidence of matrix effects. All pre-defined acceptance criteria were met; intermediate precision was <= 4.3% and repeatability 1.5-2.1% for all analyte concentrations. Bias was -3.0 to 2.1% for samples within the measuring range and 0.8-4. 5% for diluted samples, independent of the sample matrix. RMP results equivalence was demonstrated between two independent laboratories (Pearson correlation coefficient 0.997). Expanded measurement uncertainty of target value-assigned samples was <= 3.4%.Conclusions: This ID-LC-MS/MS-based approach provides a candidate RMP for methotrexate quantification. Traceability of methotrexate standard and the LC-MS/MS platform were assured by qNMR assessment and extensive method validation.
引用
收藏
页码:1917 / 1929
页数:13
相关论文
共 50 条
  • [41] Candidate reference measurement procedure for determination of urea in serum by liquid chromatography-tandem mass spectrometry
    Han, Liqiao
    Huang, Xiaoting
    Zhang, Lu
    Zhang, Qiaoxuan
    Wang, Jianbing
    Lin, Haibiao
    Yan, Jun
    Zhuang, Junhua
    Huang, Xianzhang
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2019, 162 : 124 - 129
  • [42] Evaluation and clinical application of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for serum theophylline
    Du, Yuanyuan
    Yan, Qiaofang
    Zhan, Min
    Zhang, Qiaoxuan
    Huang, Di
    Zhang, Pengwei
    Yan, Jun
    Wang, Jianbing
    Huang, Xianzhang
    Han, Liqiao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (22) : 4897 - 4906
  • [43] Quantification of psilocin in human whole blood using liquid chromatography-tandem mass spectrometry (LC-MS/MS)
    Gomonit, Munchelou M.
    Skillman, Britni
    Swortwood, Madeleine J.
    JOURNAL OF FORENSIC SCIENCES, 2024, 69 (02) : 678 - 687
  • [44] Development of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry
    Lee, Hwashim
    Park, Sangryoul
    Kim, Byungjoo
    Lee, Gaeho
    ANALYTICAL SCIENCES, 2007, 23 (03) : 299 - 303
  • [45] Development of a Candidate Reference Method for Determination of Tyrosine in Serum by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Hwashim Lee
    Sangryoul Park
    Byungjoo Kim
    Gaeho Lee
    Analytical Sciences, 2007, 23 : 299 - 303
  • [46] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [47] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [48] Simultaneous Determination of Creatinine and Creatine in Human Serum by Double-Spike Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) and Gas Chromatography-Mass Spectrometry (GC-MS)
    Fernandez-Fernandez, Mario
    Rodriguez-Gonzalez, Pablo
    Anon Alvarez, M. Elena
    Rodriguez, Felix
    Alvarez Menendez, Francisco V.
    Garcia Alonso, J. Ignacio
    ANALYTICAL CHEMISTRY, 2015, 87 (07) : 3755 - 3763
  • [49] Measurement of Plasma Free Metanephrines Using a Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) Method
    MacMahon, M. M.
    Mulready, K. J.
    Fitzgibbon, M. C.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2013, 182 : S406 - S406
  • [50] Development of isotope dilution-liquid chromatography/tandem mass spectrometry as a candidate reference method for the determination of folic acid
    Jung, Minyoung
    Kim, Byungjoo
    Boo, Doo Wan
    So, Hun-Young
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2007, 28 (05): : 745 - 750