An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of methotrexate in human serum and plasma

被引:6
|
作者
Engel, Anett [1 ]
Ruhe, Lena [1 ]
Singh, Neeraj [2 ]
Wright, Jo Anne [2 ]
Liesch, Franziska [2 ]
Bauland, Friederike [3 ]
Ostermann, Annika I. I. [1 ]
Sumalowitsch, Tamara [1 ]
Schweinsberg, Vincent J. T. [4 ]
Geistanger, Andrea [2 ]
Hegel, Johannes Kolja [1 ]
Geletneky, Christian [2 ]
Taibon, Judith [2 ,5 ]
机构
[1] Lab Berlin Charite Vivantes Serv GmbH, Collaborat & Innovat Management, Dept Studies, Berlin, Germany
[2] Roche Diagnost GmbH, Berlin, Germany
[3] Chrestos Concept GmbH & Co KG, Essen, Germany
[4] Lab Berlin Charite Vivantes GmbH, Dept Lab Med & Toxicol, Berlin, Germany
[5] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
关键词
isotope dilution LC-MS/MS; methotrexate; qNMR; reference measurement procedure; SI units; traceability; LC-MS/MS ASSAY; PHARMACOKINETICS; VALIDATION; METABOLITE; CHILDREN;
D O I
10.1515/cclm-2022-1001
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To develop an isotope dilution-liquid chromatography-tandem mass spectrometry-(ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for quantification of methotrexate in human serum and plasma.Methods: Quantitative nuclear magnetic resonance (qNMR) was used to determine absolute methotrexate content in the standard. Separation was achieved on a biphenyl reversed phase analytical column with mobile phases based on water and acetonitrile, both containing 0.1% formic acid. Sample preparation included protein precipitation in combination with high sample dilution, and method validation according to current guidelines. The following were assessed: selectivity (using analyte-spiked samples, and relevant structural-related compounds and interferences); specificity and matrix effects (via post-column infusion and comparison of human matrix vs. neat samples); precision and accuracy (in a five-day validation analysis). RMP results were compared between two independent laboratories. Measurement uncertainty was evaluated according to current guidelines.Results: The RMP separated methotrexate from potentially interfering compounds and enabled measurement over a calibration range of 7.200-5,700 ng/mL (0.01584-12.54 mu mol/L), with no evidence of matrix effects. All pre-defined acceptance criteria were met; intermediate precision was <= 4.3% and repeatability 1.5-2.1% for all analyte concentrations. Bias was -3.0 to 2.1% for samples within the measuring range and 0.8-4. 5% for diluted samples, independent of the sample matrix. RMP results equivalence was demonstrated between two independent laboratories (Pearson correlation coefficient 0.997). Expanded measurement uncertainty of target value-assigned samples was <= 3.4%.Conclusions: This ID-LC-MS/MS-based approach provides a candidate RMP for methotrexate quantification. Traceability of methotrexate standard and the LC-MS/MS platform were assured by qNMR assessment and extensive method validation.
引用
收藏
页码:1917 / 1929
页数:13
相关论文
共 50 条
  • [31] State-of-the-art of serum testosterone measurement by isotope dilution-liquid chromatography-tandem mass spectrometry
    Thienpont, Linda M.
    Van Uytfanghe, Katleen
    Blincko, Stuart
    Ramsay, Carol S.
    Xie, Hui
    Doss, Robert C.
    Keevil, Brian G.
    Owen, Laura J.
    Rockwood, Alan L.
    Kushnir, Mark M.
    Chun, Kelly Y.
    Chandler, Donald W.
    Field, Helen P.
    Sluss, Patrick M.
    CLINICAL CHEMISTRY, 2008, 54 (08) : 1290 - 1297
  • [32] Development of a simplified sample pretreatment procedure as part of an isotope dilution-liquid chromatography/tandem mass spectrometry candidate reference measurement procedure for serum total thyroxine
    Van Uytfanghe, K
    Stöckl, D
    Thienpont, LM
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (13) : 1539 - 1540
  • [33] Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma
    Zhang, Qiaoxuan
    Zhang, Lu
    Lin, Haibiao
    Cai, Zhiliang
    Yan, Jun
    Wang, Qiqin
    Han, Liqiao
    Wang, Jianbing
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (27) : 7095 - 7104
  • [34] An isotope dilution LC-MS/MS-based candidate reference method for the quantification of androstenedione in human serum and plasma
    Gradl, Katrin
    Taibon, Judith
    Singh, Neeraj
    Albrecht, Eva
    Geistanger, Andrea
    Pongratz, Stephan
    Hutzler, Stefan
    Mayer, Magdalena
    Kleinschmidt, Christine
    Geletneky, Christian
    Hofmann, Verena
    Koeppl, Daniel
    Rauh, Manfred
    Kobold, Uwe
    CLINICAL MASS SPECTROMETRY, 2020, 16 : 1 - 10
  • [35] Quantification of mycophenolic acid in plasma by isotope dilution liquid chromatography-tandem mass spectrometry candidate reference methodQuantification of mycophenolic acid in plasma by isotope dilution liquid chromatography-tandem mass spectrometry candidate reference methodJ. Lin et al.
    Jing Lin
    Min Shen
    Ting Yu
    Huimin Wang
    Jingjue Pan
    Gaipeng Huang
    Quanle Li
    Analytical and Bioanalytical Chemistry, 2025, 417 (8) : 1607 - 1617
  • [36] ESTIMATION OF MELOXICAM IN HUMAN PLASMA BY LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY (LC-MS/MS)
    Ganesh, Mani
    Hemalatha, Pushparaj
    Peng, Mei Mei
    Sakthimanigandan, Kalaimani
    Jang, Hyun Tae
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2013, 36 (07) : 867 - 880
  • [37] Simultaneous measurement of testosterone, androstenedione and dehydroepiandrosterone (DHEA) in serum and plasma using Isotope-Dilution 2-Dimension Ultra High Performance Liquid-Chromatography Tandem Mass Spectrometry (ID-LC-MS/MS)
    Buttler, Rahel M.
    Martens, Frans
    Kushnir, Mark M.
    Ackermans, Mariette T.
    Blankenstein, Marinus A.
    Heijboer, Annemieke C.
    CLINICA CHIMICA ACTA, 2015, 438 : 157 - 159
  • [38] Development of Triglyceride Certified Reference Materials in Human Frozen Serum Using Isotope Dilution-Liquid Chromatography-Tandem Mass Spectrometry
    Lee, Sun Young
    Kim, Kwonseong
    Kwon, Hyun Soo
    Kim, Young Eun
    Kim, Ki Young
    Jeong, Ji-Seon
    Oh, Han Bin
    Kim, Tae Young
    Hong, Jongki
    Kang, Dukjin
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2019, 40 (05) : 418 - 423
  • [39] Accurate Quantification of Saccharin Using Isotope Dilution Liquid Chromatography Mass Spectrometry (ID-LC/MS)
    Lee, Yunjung
    Kim, Byungjoo
    Kim, Jeongkwon
    Ahn, Seonghee
    MASS SPECTROMETRY LETTERS, 2011, 2 (02) : 37 - 40
  • [40] Isotope-Dilution Liquid Chromatography-Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum
    Botelho, Julianne Cook
    Shacklady, Christopher
    Cooper, Hans C.
    Tai, Susan S. -C.
    Van Uytfanghe, Katleen
    Thienpont, Linda M.
    Vesper, Hubert W.
    CLINICAL CHEMISTRY, 2013, 59 (02) : 372 - 380