Interferometric metrology probe for highly sloped freeform optics

被引:2
|
作者
Wisniewski, Hayden J. [1 ]
Stalcup, Thomas E. [1 ]
Smythe, Robert A. [1 ]
Olszak, Artur G. [1 ]
机构
[1] Apre Instruments Inc, 2440 W Ruthrauff Rd,Suite 100, Tucson, AZ 85705 USA
基金
美国国家航空航天局;
关键词
SELF-CALIBRATION; COORDINATE; COMPENSATION; ALGORITHM;
D O I
10.1364/AO.500553
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A wide variety of systems are employed to measure the surface profile of aspheric and freeform optical surfaces. Freeform metrology systems must accurately characterize the full surface under test, which can be difficult with steep surface slopes. Here we present an interferometric surface metrology probe for highly sloped aspheric and freeform optical surfaces. The optical design of this probe allows the measurement of surface slopes up to 50 deg without tilting the probe, which simplifies stage design and increases the accuracy of the system. A spectrally controlled light source is used to create a virtual ball in front of the probe tip to measure the surface distance and angle. This system produces a cloud of points, to which Zernike polynomials are fit and used to reconstruct the surface. We will show sensitivity tests and accuracy results.(c) 2023 Optica Publishing Group
引用
收藏
页码:8482 / 8490
页数:9
相关论文
共 50 条
  • [41] Characterizing the shape of freeform optics
    Forbes, G. W.
    OPTICS EXPRESS, 2012, 20 (03): : 2483 - 2499
  • [42] Importance of Fiducials on Freeform Optics
    Brunelle, Matt
    Yuan, Joe
    Medicus, Kate
    Nelson, Jessica DeGroote
    OPTIFAB 2015, 2015, 9633
  • [43] Freeform optics for dynamic illumination
    Shadalou, Shohreh
    Suleski, Thomas J.
    NONIMAGING OPTICS: EFFICIENT DESIGN FOR ILLUMINATION AND SOLAR CONCENTRATION XVII, 2020, 11495
  • [44] Design of Freeform Illumination Optics
    Wu, Rengmao
    Feng, Zexin
    Zheng, Zhenrong
    Liang, Rongguang
    Benitez, Pablo
    Minano, Juan C.
    Duerr, Fabian
    LASER & PHOTONICS REVIEWS, 2018, 12 (07)
  • [45] Interferometric methods for surface metrology
    Pfeifer, Tilo
    Tutsch, Rainer
    Broermann, Ernst
    SME Technical Paper (Series) MS, 1989, : 89 - 480
  • [46] Traceability in Interferometric Form Metrology
    Schulz, M.
    Blobel, G.
    Fortmeier, I.
    Stavridis, M.
    Elster, C.
    OPTICAL MEASUREMENT SYSTEMS FOR INDUSTRIAL INSPECTION IX, 2015, 9525
  • [47] Interferometric metrology of conformal domes
    Lerner, SA
    Sasian, JM
    Greivenkamp, JE
    Gappinger, RO
    Clark, SR
    WINDOW AND DOME TECHNOLOGIES AND MATERIALS VI, 1999, 3705 : 221 - 226
  • [48] Vibration insensitive fiber optic spectral interferometric probe for metrology of topography and thickness of multilayer samples
    Walecki, Wojtek J.
    INTERFEROMETRY XXI, 2022, 12223
  • [49] Current developments on optical asphere and freeform metrology
    Muehlig, S.
    Siepmann, J.
    Lotz, M.
    Wiegmann, A.
    Blobel, G.
    Mika, S.
    Beutler, A.
    Nehse, U.
    OPTICAL SYSTEMS DESIGN 2015: OPTICAL FABRICATION, TESTING, AND METROLOGY V, 2015, 9628
  • [50] APPLICATION OF RULED SURFACES IN FREEFORM AND GEAR METROLOGY
    Linkeova, Ivana
    Zeleny, Vit
    ACTA POLYTECHNICA, 2021, 61 : 99 - 109