A ceramic-GEM neutron detector with a wide neutron-flux measurement range for the beam monitoring at China spallation neutron source

被引:3
|
作者
Zhu, Lin [1 ,2 ,3 ]
Zhou, Jian-rong [1 ,2 ,3 ]
Zeng, Li-xin [1 ,2 ]
Xiao, Liang
Zhou, Xiao-juan [1 ,2 ]
Tan, Jin-hao [2 ,4 ]
Xia, Yuan-guang [1 ,2 ]
Xu, Hong
Jiang, Xing-fen [1 ,2 ]
Yang, Wen-qin [3 ]
Wang, Yan-feng
Yang, Gui-an [1 ,2 ]
Xie, Yu-guang
Teng, Hai-yun [1 ,2 ]
Li, Jia-jie [1 ,2 ]
Qiu, Yong-xiang
Shen, Pei-xun [1 ,2 ]
Wang, Song-lin
Zhuang, Jian [1 ,2 ,3 ]
Zhao, Yu-bin [1 ,2 ,3 ]
Zhang, Jun-rong [1 ,2 ,3 ]
Sun, Zhi-jia [1 ,2 ,3 ]
Chen, Yuan-bo [1 ,2 ,3 ]
机构
[1] Chinese Acad Sci, Inst High Energy Phys, State Key Lab Particle Detect & Elect, Beijing 100049, Peoples R China
[2] Spallat Neutron Source Sci Ctr, Dongguan 523803, Peoples R China
[3] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
[4] Harbin Engn Univ, Coll Nucl Sci & Technol, Harbin 150001, Peoples R China
基金
中国国家自然科学基金;
关键词
Electron multipliers (gas); Neutron detectors (cold; thermal; fast neutrons); CASCADE;
D O I
10.1088/1748-0221/18/01/P01033
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The increase of neutron flux in pulsed spallation neutron facilities imposes demands on the neutron detector for beam monitoring, including excellent neutron/gamma discrimination, wide dynamic neutron-flux measurement range, wavelength resolution, and long-term stability. In this paper, we developed a ceramic GEM-based (gas electron multiplier) neutron detector with an active area of 100 mm x 100 mm. It adopted a thinner conversion material and the stopping layer to lower the detection efficiency so as to extend the dynamic measurement range of the detector. The detection efficiency of this detector was investigated by the Monte Carlo (MC) tool FLUKA, showing that the low efficiency around 0.01% for 1 Aneutrons was reached by using a 0.1 mu m natB4C converter appended with an aluminum film of 2 mu m thickness. Its validation of the wavelength spectra measurement was verified by comparisons with that made with an LND monitor, and it could work at the condition of 7.1 x 109 n/s for 2.5 Aneutrons. In addition, it was demonstrated that this detector was able to measure the beam profile with a position resolution of better than 2.1 +/- 0.1 mm. The results of simulations and experiments show that this ceramic-GEM neutron detector can meet the requirements of the direct measurement of high-flux beam, and it will be a new neutron detector for the beam monitoring at the China spallation neutron source (CSNS).
引用
收藏
页数:15
相关论文
共 50 条
  • [1] A ceramic GEM-based neutron beam monitor for China Spallation Neutron Source
    Zhou, Jianrong
    Zhou, Xiaojuan
    Zhou, Jianjin
    Teng, Haiyun
    Yang, Jianqing
    Ma, Yichao
    Zhou, Ke
    Xia, Yuanguang
    Xiu, Qinglei
    Yang, Tao
    Jiang, Xingfen
    Zhu, Lin
    Yang, Wenqin
    Yang, Guian
    Xie, Yuguang
    Hu, Bitao
    Sun, Zhijia
    Chen, Yuanbo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 962 (962):
  • [2] A FISSION IONIZATION DETECTOR FOR NEUTRON-FLUX MEASUREMENTS AT A SPALLATION SOURCE
    WENDER, SA
    BALESTRINI, S
    BROWN, A
    HAIGHT, RC
    LAYMON, CM
    LEE, TM
    LISOWSKI, PW
    MCCORKLE, W
    NELSON, RO
    PARKER, W
    HILL, NW
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 336 (1-2): : 226 - 231
  • [3] SIMULTANEOUS MEASUREMENT OF NEUTRON-FLUX AND NEUTRON DETECTOR EFFICIENCY
    DUPONT, C
    LELEUX, P
    LIPNIK, P
    SINDANO, WK
    WAUTERS, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 256 (02): : 339 - 341
  • [4] Highly efficient GEM-based neutron detector for China Spallation Neutron Source
    Zhou, Jianrong
    Xiu, Qinglei
    Zhou, Xiaojuan
    Zhou, Jianjin
    Ma, Lingling
    Schmidt, Christian J.
    Klein, Martin
    Xia, Yuanguang
    Zhu, Lin
    Huang, Chaoqiang
    Sun, Guangai
    Hu, Bitao
    Sun, Zhijia
    Chen, Yuanbo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 953 (953):
  • [5] NEUTRON-FLUX MONITORING WITH A BLACK DETECTOR
    MEIER, MM
    WASSON, OA
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1978, 23 (01): : 72 - 72
  • [6] A CALORIMETER FOR NEUTRON-FLUX MEASUREMENT
    RICHARDSON, JM
    CHUPP, TE
    ROBERTSON, RGH
    WILKERSON, JF
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 306 (1-2): : 291 - 299
  • [7] MEASUREMENT OF UNDERGROUND NEUTRON-FLUX
    FLORKOWSKI, T
    KOSTKA, T
    KOTAS, M
    NUCLEAR GEOPHYSICS, 1992, 6 (02): : 243 - 248
  • [8] UNDERGROUND NEUTRON-FLUX MEASUREMENT
    RINDI, A
    CELANI, F
    LINDOZZI, M
    MIOZZI, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 272 (03): : 871 - 874
  • [9] Fission ionization detector for neutron flux measurements at a spallation source
    Wender, S.A.
    Balestrini, S.
    Brown, A.
    Haight, R.C.
    Laymon, C.M.
    Lee, T.M.
    Lisowski, P.W.
    McCorkle, W.
    Nelson, R.O.
    Parker, W.
    Hill, N.W.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993, A336 (1-2) : 226 - 231
  • [10] RESEARCH ON DOSE MONITORING IN BACKSCATTERING NEUTRON HALL IN CHINA SPALLATION NEUTRON SOURCE
    Wu, Qingbiao
    Zhuang, Sixuan
    Liu, Qiongyao
    Jing, Hantao
    Ye, Rong
    Li, Lun
    Wang, Yufei
    Wang, Qingbin
    RADIATION PROTECTION DOSIMETRY, 2020, 189 (02) : 253 - 269