Effects of Non-invasive Keratograph Break-Up Time on the Repeatability of Keratometry Measurements

被引:1
|
作者
Shin, Haeeun [1 ]
Jung, Soyeon [1 ]
Jung, Ji Won [1 ,2 ]
机构
[1] Inha Univ, Sch Med, Dept Ophthalmol, Incheon, South Korea
[2] Inha Univ, Sch Med, Dept Ophthalmol, 27 Inhang RO, Incheon 22332, South Korea
来源
关键词
Dry eye syndrome; Keratometry; Keratograph; 5M; Noninvasive keratograph break-up time; INTRAOCULAR-LENS POWER; FILM;
D O I
10.3341/jkos.2023.64.2.99
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the effects of the non-invasive keratograph break-up time on the repeatability of keratometric measure-ments derived using the Keratograph 5M in patients with dry eye syndrome. Methods: Thirty-six patients with dry eye and 30 controls were enrolled. We measured keratometric values twice in all subjects and explored whether the two measurements differed by more than 0.5 diopter (D) (the absolute value). We also evaluated the relationship between the absolute difference and the non-invasive keratograph break-up time (NIKBUT). Results: The intraclass correlation coefficients for astigmatism, flat keratometry (K), steep K, and mean K were all higher than 0.95 in patients, thus lower than in controls. The proportion of eyes with absolute differences over 0.5 D between the two kerato-metric values differed significantly between the dry eye and control groups in terms of the flat K and Kmax values (p = 0.033, 0.037). The average NIKBUT was negatively correlated with the absolute differences in Kmax and astigmatism (p = 0.030, R2 = 0.134 and p = 0.017, R2 = 0.160). Conclusions: In patients with dry eye syndrome, keratometric measurements are reliably repeatable, but slightly less so than in normal controls. The proportions of eyes exhibiting absolute keratometric differences over 0.5 D (two measurements) differed in the dry eye and control groups. The lower the non-invasive keratograph break-up time, the poorer the repeatability of kerato-metric measurements in the dry eye group. Therefore, care is required when obtaining keratometric values for such patients. J Korean Ophthalmol Soc 2023;64(2):99-107
引用
收藏
页码:99 / 107
页数:9
相关论文
共 50 条
  • [21] Comparative analysis of tear film stability Automated non-invasive and fluorescein tear break-up time
    Zeri, F.
    Rizzo, G. C.
    Ponzini, E.
    Borghesi, A.
    Tavazzi, S.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2024, 47 (05):
  • [22] Ability of non-invasive tear break-up time to determine tear instability in contact lens wearers
    Ersin Muhafiz
    Mehmet Siraç Demir
    International Ophthalmology, 2022, 42 : 959 - 968
  • [23] Diagnostic accuracy of non-invasive tear film break-up time assessed by the simple manual interferometric device
    Pauk, Sania Vidas
    Petricek, Igor
    Tomic, Martina
    Bulum, Tomislav
    Jandrokovic, Sonja
    Gulic, Maja Pauk
    Kalauz, Miro
    Gacina, Dina Lesin
    CONTACT LENS & ANTERIOR EYE, 2023, 46 (02):
  • [24] Reliability and agreement of subjective and objective non-invasive break-up time measurements in contact lens wearers (vol 44, pg 124, 2024)
    Valencia-Nieto, L.
    Lopez-de la Rosa, A.
    Gonzalez-Garcia, M. J.
    Lopez-Miguel, A.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2025, 45 (03) : 895 - 895
  • [25] Comparing automated and manual assessments of tear break-up time using different non-invasive devices and a fluorescein procedure
    Fabrizio Zeri
    Giulia Carlotta Rizzo
    Erika Ponzini
    Silvia Tavazzi
    Scientific Reports, 14
  • [26] Assessment of effect of artificial tears on Tear Meniscus Height and Non-invasive Keratographic Break-Up Time measures in rabbits
    Galenchik-Chan, Andre
    Langman, Lauren
    Reed, Ken
    Huang, Liqun
    Khwaja, Ishan Amin
    Kaplowitz, Kevin
    Tourmouzis, Konstantinos
    Chernoff, Diane
    Li, Deborah
    Rigas, Basil
    Honkanen, Robert A.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2024, 65 (07)
  • [27] Comparing automated and manual assessments of tear break-up time using different non-invasive devices and a fluorescein procedure
    Zeri, Fabrizio
    Rizzo, Giulia Carlotta
    Ponzini, Erika
    Tavazzi, Silvia
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [28] A pre-test post-test assessment of non-invasive keratograph break up time and corneal epithelial thickness after vaping
    Munsamy, Alvin
    Bhanprakash, Bhavna
    Sirkhot, Amina
    Mlambo, Lufun
    Dlamuka, Samukelisiwe
    Mhlongo, Ndumiso
    Naidoo, Ronelle
    AFRICAN HEALTH SCIENCES, 2019, 19 (04) : 2926 - 2933
  • [29] THE RELATION BETWEEN INVASIVE AND NONINVASIVE TEAR BREAK-UP TIME
    CHO, P
    DOUTHWAITE, W
    OPTOMETRY AND VISION SCIENCE, 1995, 72 (01) : 17 - 22
  • [30] Relation between invasive and noninvasive tear break-up time
    Cho, Pauline
    Douthwaite, William
    Optometry and Vision Science, 1995, 72 (01):