Biomechanical Analysis of Cortical Bone Trajectory Screw Versus Bone Cement Screw for Fixation in Porcine Spinal Low Bone Mass Model

被引:2
|
作者
Li, Yifan [1 ]
Wang, Silian [2 ]
Zhu, Zhenbiao [4 ]
Chen, Liwei [2 ]
Shi, Zhangpeng [3 ]
Ye, Xiaojian [1 ]
Xu, Wei [1 ]
Li, Zhikun [1 ,5 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Orthoped, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Tongren Hosp, Sch Med, Dept Imaging, Shanghai, Peoples R China
[3] Anhui Med Univ, Shanghai Clin Coll, Shanghai, Peoples R China
[4] Cent South Univ, Dept Orthopaed, Xiangya Sch Med, Affiliated Haikou Hosp, Haikou, Hainan, Peoples R China
[5] Shanghai Tongren Hosp, Shanghai, Peoples R China
来源
CLINICAL SPINE SURGERY | 2023年 / 36卷 / 04期
关键词
cortical bone trajectory screw; bone cement screw; biomechanical analysis; porcine spinal model; PEDICLE; AUGMENTATION; PLACEMENT; ADJACENT;
D O I
10.1097/BSD.0000000000001395
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: A prospective study of in vitro animal. Objection: To compare the biomechanics of cortical bone trajectory screw (CBT) and bone cement screw (BC) in an isolated porcine spinal low bone mass model. Summary of Background Data: The choice of spinal fixation in patients with osteoporosis remains controversial. Is CBT better than BC? Research on this issue is lacking. Methods:Ten porcine spines with 3 segments were treated with EDTA decalcification. After 8 weeks, all the models met the criteria of low bone mass. Ten specimens were randomly divided into groups, group was implanted with CBT screw (CBT group) and the other group was implanted with bone cement screw (BC group). The biomechanical material testing machine was used to compare the porcine spine activities of the two groups in flexion, extension, bending, and axial rotation, and then insertional torque, pull-out force, and anti-compression force of the 2 groups were compared. Independent sample t test was used for comparison between groups. Results: Ten 3 segments of porcine spine models with low bone mass were established, and the bone mineral density of all models was lower than 0.75 g/cm(2). There is no difference between the CBT and BC groups in flexion, extension, bending, and axial rotation angle, P>0.05. However, there were significant differences between the 2 groups and the control group, with P<0.01. The 2 groups significantly differed between the insertional torque (P=0.03) and the screw pull-out force (P=0.021). The anti-compression forces between the 2 groups have no significant difference between the two groups (P=0.946). Conclusions: The insertional torque and pull-out force of the CBT were higher than those of the BC in the isolated low bone porcine spine model. The range of motion and anti-compression ability of the model was similar between the 2 fixation methods.
引用
收藏
页码:E145 / E152
页数:8
相关论文
共 50 条
  • [1] A Biomechanical Study on Cortical Bone Trajectory Screw Fixation Augmented With Cement in Osteoporotic Spines
    Wang, Yuetian
    Yang, Lei
    Li, Chunde
    Sun, Haolin
    [J]. GLOBAL SPINE JOURNAL, 2023, 13 (08) : 2115 - 2123
  • [2] Biomechanical evaluation of bone screw fixation with a novel bone cement
    Tiina Juvonen
    Juha-Pekka Nuutinen
    Arto P Koistinen
    Heikki Kröger
    Reijo Lappalainen
    [J]. BioMedical Engineering OnLine, 14
  • [3] Biomechanical evaluation of bone screw fixation with a novel bone cement
    Juvonen, Tiina
    Nuutinen, Juha-Pekka
    Koistinen, Arto P.
    Kroger, Heikki
    Lappalainen, Reijo
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2015, 14
  • [4] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Hiroki Oshino
    Toshihiko Sakakibara
    Tadashi Inaba
    Takamasa Yoshikawa
    Takaya Kato
    Yuichi Kasai
    [J]. Journal of Orthopaedic Surgery and Research, 10
  • [5] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Oshino, Hiroki
    Sakakibara, Toshihiko
    Inaba, Tadashi
    Yoshikawa, Takamasa
    Kato, Takaya
    Kasai, Yuichi
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2015, 10
  • [6] Erratum to: A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Hiroki Oshino
    Toshihiko Sakakibara
    Tadashi Inaba
    Takamasa Yoshikawa
    Takaya Kato
    Yuichi Kasai
    [J]. Journal of Orthopaedic Surgery and Research, 10
  • [7] Biomechanical Changes of Adjacent and Fixed Segments Through Cortical Bone Trajectory Screw Fixation versus Traditional Trajectory Screw Fixation in the Lumbar Spine: A Finite Element Analysis
    Zhang, Lai
    Li, Hui-Min
    Zhang, Renjie
    Zhang, Huaqing
    Shen, Cai-Liang
    [J]. WORLD NEUROSURGERY, 2021, 151 : E447 - E456
  • [8] Enhanced bone screw fixation with biodegradable bone cement in osteoporotic bone model
    Juvonen, Tiina
    Koistinen, Arto
    Kroger, Heikki
    Lappalainen, Reijo
    [J]. JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2012, 10 (02): : 113 - 118
  • [9] Cortical bone trajectory screw technique
    Goel, Atul
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2018, 29 (01) : 121 - 122
  • [10] The Biomechanical Properties of Pedicle Screw Fixation Combined With Trajectory Bone Cement Augmentation in Osteoporotic Vertebrae
    Fan, Haitao T.
    Zhang, Renjie J.
    Shen, Cailiang L.
    Dong, Fulong L.
    Li, Yong
    Song, Peiwen W.
    Gong, Chen
    Wang, Yijin J.
    [J]. CLINICAL SPINE SURGERY, 2016, 29 (02): : 78 - 85