A comparative study on the accuracy of an automated multi-parameter high-dose-rate brachytherapy quality assurance system

被引:0
|
作者
Zhang, Jing [1 ]
Jia, Mengyu [2 ]
Yuan, Zhiyong [1 ]
Qian, Shaowen [3 ]
Zhang, Daguang [1 ]
Wang, Wei [1 ]
机构
[1] Tianjin Med Univ Canc Inst & Hosp, Key Lab Canc Prevent & Therapy, Natl Clin Res Ctr Canc, Dept Radiat Oncol,Tianjins Clin Res Ctr Canc, Tianjin 300060, Peoples R China
[2] Tianjin Univ, Sch Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
[3] 960th Hosp Joint Logist Support Force PLA, Dept Med Imaging, Jinan, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
High-dose-rate brachytherapy; Quality assurance; Afterloader; RADIATION-THERAPY; CANCER; AAPM;
D O I
10.1016/j.radphyschem.2023.111472
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This study aims to explore the measuring accuracy of an automated high-dose-rate brachytherapy (HDR BT) quality assurance system (hereinafter referred to as OriQA system) that could provide simultaneous measurements of dwell position, dwell time and source strength in a single test. Elekta-Flexitron was selected as the HDR BT afterloader, and measured the farthest transmission distance by source position simulator. Within the effective distance, we set dwell positions in two groups, specifically, plan_1 with an interval of 2.00 cm: 106.00 cm, 108.00 cm, 110.00 cm, 102.00 cm, 104.00 cm, 106.00 cm, 108.00 cm, and plan_2 with an interval of 0.20 cm: 111.40 cm, 111.60 cm, 111.80 cm, 112.00 cm, 112.20 cm, 112.40 cm, 112.60 cm. The measured values collected from OriQA device were compared with the data from traditional HDR BT QA tool (source position check ruler). Three groups of dwell time were set, specifically, plan_1: with fixed dwell time 2.00 s, plan_3 with interval 0.10 s: 3.70 s, 3.80 s, 3.90 s, 4.00 s, 4.10 s, 4.20 s, 4.30 s, and plan_4: fixed dwell time 5.00 s. Source strength was also measured continuously for a period of 10 weeks, and compared with the data collected from the well-chamber. The OriQA system showed reliable performance in the evaluation of measuring accuracy. The adjusted farthest transmission distance was 119.33 cm. And according to the time-position graph, the maximum deviation of dwell position between the measured and planned data was 0.34 mm, which was smaller than that between the ruler and planed data (0.75 mm). The maximum deviation of dwell time between the measured and planned data was 0.05 s. Fitting analysis revealed linear relationship between source strength Sk and the intensity detector response, with R2 = 0.9877, and intensity deviation fluctuated between 1.14% and 2.19%. Compared with the traditional QA tool, the OriQA device provided improved measuring accuracy and efficiency QA procedure for HDR BT.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] An Automated Treatment Plan Quality Assurance Program for High-Dose Rate (HDR) Brachytherapy with a VaginalCylinder Applicator
    Zhou, Y.
    Tan, J.
    Jiang, S.
    Albuquerque, K.
    Jia, X.
    MEDICAL PHYSICS, 2016, 43 (06) : 3622 - 3623
  • [32] Simulation study of high-dose-rate brachytherapy for early glottic cancer
    Hoffman, Matthew R.
    McCulloch, Timothy M.
    Mohindra, Pranshu
    Das, Rupak
    Geurts, Mark
    Harari, Paul M.
    BRACHYTHERAPY, 2016, 15 (01) : 94 - 101
  • [33] A Study of Computed Radiography for High Dose-Rate Brachytherapy Source Position Quality Assurance
    Piriyasang, D.
    Tangboonduangjit, P.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [34] Automated construction of an intraoperative high-dose-rate treatment plan library for the Varian brachytherapy treatment planning system
    Deufel, Christopher L.
    Furutani, Keith M.
    Dahl, Robert A.
    Haddock, Michael G.
    BRACHYTHERAPY, 2016, 15 (04) : 531 - 536
  • [35] AutoBrachy: A Streamlined Automated Treatment Planning, Quality Assurance, and Documentation System for Tandem-And-Ovoid High Dose-Rate Brachytherapy
    Tan, J.
    Klages, P.
    Shen, C.
    Yang, M.
    Albuquerque, K.
    Jia, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E658 - E658
  • [36] 3D CT-based high-dose-rate breast brachytherapy implants: Treatment planning and quality assurance
    Department of Human Oncology, University of Wisconsin, Madison, WI, United States
    不详
    1600, 1224-1228 (July 15, 2004):
  • [37] 3D CT-based high-dose-rate breast brachytherapy implants: Treatment planning and quality assurance
    Das, RK
    Patel, R
    Shah, H
    Odau, H
    Kuske, RR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 1224 - 1228
  • [38] High-Dose-Rate 192Ir Brachytherapy Dose Verification: A Phantom Study
    Nikoofar, Alireza
    Hoseinpour, Zohreh
    Mahdavi, Seied Rabi
    Hasanzadeh, Hadi
    Tavirani, Mostafa Rezaei
    IRANIAN JOURNAL OF CANCER PREVENTION, 2015, 8 (03)
  • [39] Dosimetric study of Varian Universal Multi-Channel Cylinder System for High-Dose-Rate 192Ir Brachytherapy
    Tien, Christopher J.
    Damast, Shari
    Chen, Zhe
    BRACHYTHERAPY, 2022, 21 (02) : 244 - 254
  • [40] Low-Dose-Rate Versus High-Dose-Rate Versus Pulsed-Dose-Rate Intracavitary Brachytherapy in Cervical Carcinoma: A Monoinstitutional Comparative Study
    Sharma, D. N.
    Rath, G. K.
    Gandhi, A. K.
    Kumar, P.
    Haresh, K.
    Gupta, S.
    Pandey, R.
    Giridhar, P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E278 - E278