A comparative study on the accuracy of an automated multi-parameter high-dose-rate brachytherapy quality assurance system

被引:0
|
作者
Zhang, Jing [1 ]
Jia, Mengyu [2 ]
Yuan, Zhiyong [1 ]
Qian, Shaowen [3 ]
Zhang, Daguang [1 ]
Wang, Wei [1 ]
机构
[1] Tianjin Med Univ Canc Inst & Hosp, Key Lab Canc Prevent & Therapy, Natl Clin Res Ctr Canc, Dept Radiat Oncol,Tianjins Clin Res Ctr Canc, Tianjin 300060, Peoples R China
[2] Tianjin Univ, Sch Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
[3] 960th Hosp Joint Logist Support Force PLA, Dept Med Imaging, Jinan, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
High-dose-rate brachytherapy; Quality assurance; Afterloader; RADIATION-THERAPY; CANCER; AAPM;
D O I
10.1016/j.radphyschem.2023.111472
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This study aims to explore the measuring accuracy of an automated high-dose-rate brachytherapy (HDR BT) quality assurance system (hereinafter referred to as OriQA system) that could provide simultaneous measurements of dwell position, dwell time and source strength in a single test. Elekta-Flexitron was selected as the HDR BT afterloader, and measured the farthest transmission distance by source position simulator. Within the effective distance, we set dwell positions in two groups, specifically, plan_1 with an interval of 2.00 cm: 106.00 cm, 108.00 cm, 110.00 cm, 102.00 cm, 104.00 cm, 106.00 cm, 108.00 cm, and plan_2 with an interval of 0.20 cm: 111.40 cm, 111.60 cm, 111.80 cm, 112.00 cm, 112.20 cm, 112.40 cm, 112.60 cm. The measured values collected from OriQA device were compared with the data from traditional HDR BT QA tool (source position check ruler). Three groups of dwell time were set, specifically, plan_1: with fixed dwell time 2.00 s, plan_3 with interval 0.10 s: 3.70 s, 3.80 s, 3.90 s, 4.00 s, 4.10 s, 4.20 s, 4.30 s, and plan_4: fixed dwell time 5.00 s. Source strength was also measured continuously for a period of 10 weeks, and compared with the data collected from the well-chamber. The OriQA system showed reliable performance in the evaluation of measuring accuracy. The adjusted farthest transmission distance was 119.33 cm. And according to the time-position graph, the maximum deviation of dwell position between the measured and planned data was 0.34 mm, which was smaller than that between the ruler and planed data (0.75 mm). The maximum deviation of dwell time between the measured and planned data was 0.05 s. Fitting analysis revealed linear relationship between source strength Sk and the intensity detector response, with R2 = 0.9877, and intensity deviation fluctuated between 1.14% and 2.19%. Compared with the traditional QA tool, the OriQA device provided improved measuring accuracy and efficiency QA procedure for HDR BT.
引用
收藏
页数:6
相关论文
共 50 条
  • [11] Development of Dosimetric Verification System for Patient-Specific Quality Assurance of High-Dose-Rate Brachytherapy
    Kang, Sang-Won
    Chung, Jin-Beom
    Kim, Kyeong-Hyeon
    Choi, Chang Heon
    Kang, Seonghee
    Shin, Dong-Seok
    Cho, Woong
    Eom, Kuen-Yong
    Park, Hae-Jin
    Kim, Jin-Young
    Song, Changhoon
    Kim, In Ah
    Kim, Jae-Sung
    Suh, Tae Suk
    Park, Justin C.
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [12] Feasibility Study of Patient-specific Quality Assurance System for High-dose-rate Brachytherapy in Patients with Cervical Cancer
    Lee, Boram
    Ahn, Sung Hwan
    Kim, Hyeyoung
    Han, Youngyih
    Huh, Seung Jae
    Kim, Jin Sung
    Kim, Dong Wook
    Sim, Jina
    Yoon, Myonggeun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 68 (08) : L1029 - L1036
  • [13] Feasibility study of patient-specific quality assurance system for high-dose-rate brachytherapy in patients with cervical cancer
    Boram Lee
    Sung Hwan Ahn
    Hyeyoung Kim
    Youngyih Han
    Seung Jae Huh
    Jin Sung Kim
    Dong Wook Kim
    Jina Sim
    Myonggeun Yoon
    Journal of the Korean Physical Society, 2016, 68 : 1029 - 1036
  • [15] Automated Quality Assurance for Remote-Afterloading High Dose Rate Brachytherapy
    Kim, Anthony
    Ravi, Ananth
    MEDICAL PHYSICS, 2014, 41 (08) : 15 - 15
  • [16] A quality assurance device for measuring afterloader performance and transit dose for nasobiliary high-dose-rate brachytherapy
    Deufel, Christopher L.
    Mullins, John P.
    Zakhary, Mark J.
    BRACHYTHERAPY, 2018, 17 (04) : 726 - 731
  • [17] "Magic phantom" - a quality assurance system for high dose rate brachytherapy
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Corde, S.
    Bucci, J.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S139 - S140
  • [18] A Quality Assurance System with a Web Camera for High Dose Rate Brachytherapy
    Ueda, Y.
    Hirose, A.
    Oohira, S.
    Isono, M.
    Tsujii, K.
    Miyazaki, M.
    Kawaguchi, Y.
    Konishi, K.
    Teshima, T.
    MEDICAL PHYSICS, 2015, 42 (06) : 3346 - 3346
  • [19] Standardization and automation of quality assurance for high-dose-rate brachytherapy planning with application programming interface
    Cai, Bin
    Altman, Michael B.
    Reynoso, Francisco
    Garcia-Ramirez, Jose
    He, Angell
    Edward, Sharbacha S.
    Zoberi, Imran
    Thomas, Maria A.
    Gay, Hiram
    Mutic, Sasa
    Zoberi, Jacqueline E.
    BRACHYTHERAPY, 2019, 18 (01) : 108 - 114
  • [20] Radiochromic film-based quality assurance for CT-based high-dose-rate brachytherapy
    Asgharizadeh, Saeid
    Bekerat, Hamed
    Syme, Alasdair
    Aldelaijan, Saad
    DeBlois, Francois
    Te Vuong
    Evans, Michael
    Seuntjens, Jan
    Devic, Slobodan
    BRACHYTHERAPY, 2015, 14 (04) : 578 - 585