Single energy metal artifact reduction performs better than virtual monoenergetic dual-energy reconstruction in CT of the equine proximal phalanx

被引:2
|
作者
Faulkner, Josephine E. [1 ]
Broeckx, Bart J. G. [2 ]
Martens, Ann [3 ]
Raes, Els [1 ]
Haardt, Hanna [3 ]
Vanderperren, Katrien [1 ]
机构
[1] Univ Ghent, Fac Vet Med, Dept Morphol Imaging Orthoped Rehabil & Nutr, Salisburylaan 133, B-9820 Merelbeke, Belgium
[2] Univ Ghent, Fac Vet Med, Dept Vet & Biosci, Merelbeke, Belgium
[3] Univ Ghent, Fac Vet Med, Dept Large Anim Surg Anesthesia & Orthoped, Ghent, Belgium
关键词
blooming artifact; distal limb; horse; image quality; metal implant; COMPUTED-TOMOGRAPHY; IMAGING FINDINGS; SAGITTAL GROOVE; HORSES; FRACTURES; CONFIGURATION;
D O I
10.1111/vru.13258
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Metal artifacts in CT negatively impact the evaluation of surgical implants and the surrounding tissues. The aim of this prospective experimental study was to evaluate the ability of a single energy metal artifact reduction (SEMAR (TM), Canon) algorithm and virtual monoenergetic (VM) dual-energy CT (DECT) scanning techniques to reduce metal artifacts from stainless steel screws surgically inserted into the equine proximal phalanx. Seven acquisitions of 18 cadaver limbs were performed on a Canon Aquilion One Vision CT scanner (Helical +SEMAR, Volume +SEMAR, Standard Helical, Standard Volume and VM DECT at 135, 120, and 105 keV) and reconstructed in a bone kernel. Blinded subjective evaluation performed by three observers indicated a significant effect of acquisition in both adjacent tissues (P < 0.001) and distant tissues (P < 0.001) and the best metal artifact reduction was seen with Helical +SEMAR and Volume +SEMAR. The subjective overall preference of CT acquisition type was (1) Helical +SEMAR, (2) Volume +SEMAR, (3) VM DECT 135 keV, (4) VM DECT 120 keV, (5) VM DECT 105 keV, (6) Standard Helical, (7) Standard Volume (P < 0.001). Unblinded objective evaluation performed by one observer showed that VM DECT 120 keV, Helical +SEMAR, and Volume +SEMAR performed similarly and were objectively the best at reducing blooming artifact. Overall, the best metal artifact reduction was obtained with SEMAR, followed by VM DECT. However, VM DECT performance varies with energy level and was associated with decreased image quality in distant tissues and artifactual overcorrection of metal artifacts at high energy levels.
引用
收藏
页码:677 / 685
页数:9
相关论文
共 50 条
  • [41] Evaluation of metal artefacts reduction by application of monoenergetic extrapolation of dual-energy CT: A phantom study with different metal implants
    Saidun, H. A.
    Shuaib, I. L.
    Daud, N. M.
    Sobri, N. F. M.
    Osman, N. D.
    18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [42] Benefits of iterative metal artifact reduction and dual-energy CT towards mitigating artifact in the setting of total shoulder prostheses
    Mohammadinejad, Payam
    Baffour, Francis, I
    Adkins, Mark C.
    Yu, Lifeng
    McCollough, Cynthia H.
    Fletcher, Joel G.
    Glazebrook, Katrina N.
    SKELETAL RADIOLOGY, 2021, 50 (01) : 51 - 58
  • [43] Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair
    Boos, Johannes
    Fang, Jieming
    Heidinger, Benedikt H.
    Raptopoulos, Vassilios
    Brook, Olga R.
    ABDOMINAL RADIOLOGY, 2017, 42 (03) : 749 - 758
  • [44] Quantitative and qualitative assessments of deep learning image reconstruction in low-keV virtual monoenergetic dual-energy CT
    Jack Junchi Xu
    Lars Lönn
    Esben Budtz-Jørgensen
    Kristoffer L. Hansen
    Peter S. Ulriksen
    European Radiology, 2022, 32 : 7098 - 7107
  • [45] Quantitative and qualitative assessments of deep learning image reconstruction in low-keV virtual monoenergetic dual-energy CT
    Xu, Jack Junchi
    Lonn, Lars
    Budtz-Jorgensen, Esben
    Hansen, Kristoffer L.
    Ulriksen, Peter S.
    EUROPEAN RADIOLOGY, 2022, 32 (10) : 7098 - 7107
  • [46] Reduction of Metal Artifacts Caused by Titanium Peduncular Screws in the Spine by Means of Monoenergetic Images and the Metal Artifact Reduction Software in Dual-Energy Computed Tomography
    Ceccarelli, Luca
    Vara, Giulio
    Ponti, Federico
    Miceli, Marco
    Golfieri, Rita
    Facchini, Giancarlo
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (02) : 152 - 158
  • [47] CT Number Stability Across Patient Sizes Using Virtual-Monoenergetic Dual-Energy CT
    Michalak, G.
    Grimes, J.
    Halaweish, A.
    Fletcher, J.
    McCollough, C.
    MEDICAL PHYSICS, 2014, 41 (06) : 477 - 477
  • [48] Impact of Phantom Size on Low-Energy Virtual Monoenergetic Images of Three Dual-Energy CT Platforms
    Greffier, Joel
    Van Ngoc Ty, Claire
    Fitton, Isabelle
    Frandon, Julien
    Beregi, Jean-Paul
    Dabli, Djamel
    DIAGNOSTICS, 2023, 13 (19)
  • [49] Optimal Energy of Virtual Monoenergetic Imaging From Dual-Energy CT for Target Delineation and Radiation Response Assessment
    Noid, G.
    Schott, D.
    Schmidt, T.
    Tai, A.
    Li, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E677 - E677
  • [50] Complementary contrast media for metal artifact reduction in dual-energy computed tomography
    Lambert, Jack W.
    Edic, Peter M.
    FitzGerald, Paul F.
    Torres, Andrew S.
    Yeh, Benjamin M.
    JOURNAL OF MEDICAL IMAGING, 2015, 2 (03)