Patient- Specific Drill Guide Template for Pedicle Screw Insertion into the Atlantoaxial Cervical Spine Using Stereolithographic Modeling: An In Vitro Study

被引:2
|
作者
Bundoc, Rafael Cruz [1 ]
Obenieta, Harjoland Lim [1 ]
Dizon, Dave Anthony Gatchalian [1 ]
机构
[1] Univ Philippines, Dept Orthopaed, Sect Spine Surg, Philippine Gen Hosp, Manila, Philippines
关键词
Axial cervical spine; Cervical pedicle screw fixation; Three-dimensional printing; Stereolithography; Patient-specific surgical template; FIXATION; PLACEMENT; ACCURACY;
D O I
10.31616/asj.2021.0362
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Study Design: Cadaveric study. Purpose: This study aimed to assess the accuracy and feasibility of cervical pedicle screw (CPS) insertion into the atlantoaxial cervical spine using a patient-specific drill guide template constructed from a stereolithographic model. Overview of Literature: CPS fixation is a widely accepted procedure for posterior cervical fixation because of its biomechanical advantages, particularly in the subaxial cervical region. The extremely narrow corridors of the atlantoaxial spine make CPS insertion more difficult, requiring the development of new tools to ensure accurate placement. Methods: Fifteen atlantoaxial cervical vertebra specimens from 15 cadavers were scanned into thin slices using computed tomography. Images of the cadaver spine were digitally processed and rendered stl files so that they could be printed to scale as three-dimensional (3D) plastic models. Manually molded dental acrylic drill guide templates with pins inserted in the pedicles of the plastic cervical models were placed over the 3D printed models. The drill guide templates were used for precise placement of the drill holes in the pedicles of cadaveric specimens for pedicle screw fixation. The accuracy of screw placement was evaluated by an independent evaluator. Results: A total of 60 pedicles (combined C1 and C2) from 15 cadaveric axial cervical vertebrae were evaluated. The total acceptable accuracy for pedicle screw insertion in the atlantoaxial cervical vertebrae is 95%. An accuracy rate of 100% was achieved for C1 while an acceptable accuracy rate of 90% was achieved for C2. Conclusions: The use of a patient-specific drill guide constructed using stereolithography improved the accuracy of CPS placement in a cadaveric model.
引用
下载
收藏
页码:8 / 16
页数:9
相关论文
共 50 条
  • [31] A multi-level rapid prototyping drill guide template reduces the perforation risk of pedicle screw placement in the lumbar and sacral spine
    Matjaz Merc
    Igor Drstvensek
    Matjaz Vogrin
    Tomaz Brajlih
    Gregor Recnik
    Archives of Orthopaedic and Trauma Surgery, 2013, 133 : 893 - 899
  • [32] Safety and Feasibility of Cervical Pedicle Screw Insertion in Pediatric Subaxial Cervical Spine Without Navigation: A Retrospective Cohort Study
    Garg, Bhavuk
    Mehta, Nishank
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2021, 41 (02) : 119 - 126
  • [33] Unilateral Spinous Process Noncovering Hook Type Patient-specific Drill Template for Thoracic Pedicle Screw Fixation
    Kim, Sang B.
    Won, Yougun
    Yoo, Hyun Jin
    Sin, Lee J.
    Rhee, John M.
    Lee, Sang W.
    Lee, Gi-Soo
    SPINE, 2017, 42 (18) : E1050 - E1057
  • [34] Design and Application of a Novel Patient-Specific Three-Dimensional Printed Drill Navigational Guiding in Atlantoaxial Pedicle Screw Placement
    Pu, Xinwei
    Yin, Mengchen
    Ma, Junming
    Liu, Yujie
    Chen, Guanghui
    Huang, Quan
    Zhao, Guoquan
    Lu, Tingsheng
    Yao, Shudan
    Chen, Qinlin
    Luo, Chunshan
    WORLD NEUROSURGERY, 2018, 114 : E1 - E10
  • [35] Is a patient-specific drill template via a cortical bone trajectory safe in cervical anterior transpedicular insertion?
    Peng, Peng
    Xu, Yafei
    Zhang, Xintao
    Zhu, Meisong
    Du, Bingran
    Li, Wenrui
    Huang, Wenhua
    Song, Jun
    Li, Jianyi
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2018, 13
  • [36] Is a patient-specific drill template via a cortical bone trajectory safe in cervical anterior transpedicular insertion?
    Peng Peng
    Yafei Xu
    Xintao Zhang
    Meisong Zhu
    Bingran Du
    Wenrui Li
    Wenhua Huang
    Jun Song
    Jianyi Li
    Journal of Orthopaedic Surgery and Research, 13
  • [37] A novel patient-specific three-dimensional-printed external template to guide iliosacral screw insertion: a retrospective study
    Yang, Fan
    Yao, Sheng
    Chen, Kai-fang
    Zhu, Feng-zhao
    Xiong, Ze-kang
    Ji, Yan-hui
    Sun, Ting-fang
    Guo, Xiao-dong
    BMC MUSCULOSKELETAL DISORDERS, 2018, 19
  • [38] A novel patient-specific three-dimensional-printed external template to guide iliosacral screw insertion: a retrospective study
    Fan Yang
    Sheng Yao
    Kai-fang Chen
    Feng-zhao Zhu
    Ze-kang Xiong
    Yan-hui Ji
    Ting-fang Sun
    Xiao-dong Guo
    BMC Musculoskeletal Disorders, 19
  • [39] Accuracy of Cervical Pedicle Screw Insertion With and Without a Navigation-Linked High-Speed Drill: A Retrospective Clinical Study
    Ono, Kosei
    Murata, Sohei
    Matsushita, Mutsumi
    Shimizu, Yu
    Nakamura, Yusuke
    Yabe, Taisuke
    Ito, Hiromu
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (09)
  • [40] Accuracy of Patient-Specific 3D-Printed Guides for Pedicle Screw Insertion in Spine Revision Surgery: Results of a Retrospective Study
    Faldini, Cesare
    Barile, Francesca
    Cerasoli, Tosca
    Ialuna, Marco
    Viroli, Giovanni
    Manzetti, Marco
    Traversari, Matteo
    Mazzotti, Antonio
    Ruffilli, Alberto
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2022, 41