Unilateral Spinous Process Noncovering Hook Type Patient-specific Drill Template for Thoracic Pedicle Screw Fixation

被引:14
|
作者
Kim, Sang B. [1 ]
Won, Yougun [1 ]
Yoo, Hyun Jin [1 ]
Sin, Lee J. [1 ]
Rhee, John M. [2 ]
Lee, Sang W. [3 ]
Lee, Gi-Soo [1 ]
机构
[1] Konyang Univ, Coll Med, Boston, MA USA
[2] Emory Univ, Dept Orthoped Surg, Atlanta, GA 30322 USA
[3] Konyang Univ, Sch Biomed Engn, Boston, MA USA
关键词
3D Printer; classification; drill template; rapid prototyping; thoracic pedicle screw; GUIDE TEMPLATE; INSERTION PROCEDURE; PERFORATION RISK; PLACEMENT; LUMBAR; ACCURACY; SYSTEM;
D O I
10.1097/BRS.0000000000002067
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Prospective clinical pilot study. Objective. To confirm the accuracy of thoracic pedicle screw placement by using the unilateral spinous noncovering hook type patient-specific drill template (PSDT) made through rapid prototyping (RP) and to analyze previously reported PSDT designs and their characteristics. Summary of Background Data. Pedicle screw fixation is the most common form of the posterior instrumentation of the thoracic and lumbar spine. Various techniques have been introduced to improve pedicle screw placement. Among them PSDT with a preplanned trajectory has been considered a promising solution; however, we don't have consensus on proper character of the template. Methods. Preoperative spiral three-dimensional (3D) computed tomography (CT) was performed on the thoracic spine. The images were stored in DICOM format and transferred to a workstation running MIMICS 17.0 software to generate a 3D reconstruction template for the desired thoracic vertebra. The accurate trajectory and screw diameter and length were calculated with UG Imageware 12.1. The guide template was sterilized and used intraoperatively to assist with the placement of thoracic pedicle screws. After all pedicle trajectory screws had been inserted. We reviewed 12 previous reports and classified them according to the shape and system of PSDT that met the inclusion criteria of the review. Results. Ten screws were placed by using the PSDT without violating the single laminar cortex. There was no violation of the spinal canal or the cortex of pedicle on postoperative CT scans. The results of 13 PSDT types included in the current study suggested that there is no significant difference in accuracy between the PSDTs. Conclusion. The unilateral spinous process noncovering hook type PSDT made through RP provided an accurate trajectory for the thoracic vertebra, and the classification of PSDT in this study could be helpful for further studies.
引用
收藏
页码:E1050 / E1057
页数:8
相关论文
共 43 条
  • [1] Rapid prototype patient-specific drill template for cervical pedicle screw placement
    Owen, Brent D.
    Christensen, Gary E.
    Reinhardt, Joseph M.
    Ryken, Timothy C.
    [J]. COMPUTER AIDED SURGERY, 2007, 12 (05) : 303 - 308
  • [2] Accuracy and efficacy of thoracic pedicle screws in scoliosis with patient-specific drill template
    Sheng Lu
    Yuan Z. Zhang
    Zheng Wang
    Ji H. Shi
    Yu B. Chen
    Xing M. Xu
    Yong Q. Xu
    [J]. Medical & Biological Engineering & Computing, 2012, 50 : 751 - 758
  • [3] Accuracy and efficacy of thoracic pedicle screws in scoliosis with patient-specific drill template
    Lu, Sheng
    Zhang, Yuan Z.
    Wang, Zheng
    Shi, Ji H.
    Chen, Yu B.
    Xu, Xing M.
    Xu, Yong Q.
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2012, 50 (07) : 751 - 758
  • [4] A Novel Patient-Specific Navigational Template for Cervical Pedicle Screw Placement
    Lu, Sheng
    Xu, Yong Q.
    Lu, William W.
    Ni, Guo X.
    Li, Yan B.
    Shi, Ji H.
    Li, Dong P.
    Chen, Guo P.
    Chen, Yu B.
    Zhang, Yuan Z.
    [J]. SPINE, 2009, 34 (26) : E959 - E964
  • [5] Rapid prototype patient-specific drill templates for cervical pedicle screw placement
    Ryken, TC
    Owen, BD
    Reinhardt, JM
    Christensen, GE
    [J]. JOURNAL OF NEUROSURGERY, 2006, 104 (04) : A640 - A640
  • [6] Design, Fabrication, and Accuracy of a Novel Noncovering Lock-Mechanism Bilateral Patient-Specific Drill Guide Template for Nondeformed and Deformed Thoracic Spines
    Ashouri-Sanjani, Mehran
    Mohammadi-Moghadam, Shima
    Azimi, Parisa
    Arjmand, Navid
    [J]. HSS JOURNAL, 2021, 17 (02) : 213 - 222
  • [7] Accuracy of pedicle screw placement using patient-specific template guide system
    Fujita, Ryo
    Oda, Itaru
    Takeuchi, Hirohito
    Oshima, Shigeki
    Fujiya, Masanori
    Yahara, Yasuhito
    Kawaguchi, Yoshiharu
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2022, 27 (02) : 348 - 354
  • [8] Design and application of a novel patient-specific 3D printed drill navigational guiding template in percutaneous thoracolumbar pedicle screw fixation: A cadaveric study
    Li, Zhonghai
    Xu, Dingjun
    Li, FengNing
    Liu, Mozhen
    Xu, Gang
    Yang, Ming
    [J]. JOURNAL OF CLINICAL NEUROSCIENCE, 2020, 73 : 294 - 298
  • [9] Safe and Accurate Midcervical Pedicle Screw Insertion Procedure With the Patient-Specific Screw Guide Template System
    Kaneyama, Shuichi
    Sugawara, Taku
    Sumi, Masatoshi
    [J]. SPINE, 2015, 40 (06) : E341 - E348
  • [10] Accuracy of Patient-Specific 3D-Printed Drill Guides for Pedicle and Lateral Mass Screw Insertion An Analysis of 76 Cervical and Thoracic Screw Trajectories
    Pijpker, Peter A. J.
    Kraeima, Joep
    Witjes, Max J. H.
    Oterdoom, D. L. Marinus
    Vergeer, Rob A.
    Coppes, Maarten H.
    Groen, Rob J. M.
    Kuijlen, Jos M. A.
    [J]. SPINE, 2021, 46 (03) : 160 - 168