Unilateral Spinous Process Noncovering Hook Type Patient-specific Drill Template for Thoracic Pedicle Screw Fixation

被引:14
|
作者
Kim, Sang B. [1 ]
Won, Yougun [1 ]
Yoo, Hyun Jin [1 ]
Sin, Lee J. [1 ]
Rhee, John M. [2 ]
Lee, Sang W. [3 ]
Lee, Gi-Soo [1 ]
机构
[1] Konyang Univ, Coll Med, Boston, MA USA
[2] Emory Univ, Dept Orthoped Surg, Atlanta, GA 30322 USA
[3] Konyang Univ, Sch Biomed Engn, Boston, MA USA
关键词
3D Printer; classification; drill template; rapid prototyping; thoracic pedicle screw; GUIDE TEMPLATE; INSERTION PROCEDURE; PERFORATION RISK; PLACEMENT; LUMBAR; ACCURACY; SYSTEM;
D O I
10.1097/BRS.0000000000002067
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Prospective clinical pilot study. Objective. To confirm the accuracy of thoracic pedicle screw placement by using the unilateral spinous noncovering hook type patient-specific drill template (PSDT) made through rapid prototyping (RP) and to analyze previously reported PSDT designs and their characteristics. Summary of Background Data. Pedicle screw fixation is the most common form of the posterior instrumentation of the thoracic and lumbar spine. Various techniques have been introduced to improve pedicle screw placement. Among them PSDT with a preplanned trajectory has been considered a promising solution; however, we don't have consensus on proper character of the template. Methods. Preoperative spiral three-dimensional (3D) computed tomography (CT) was performed on the thoracic spine. The images were stored in DICOM format and transferred to a workstation running MIMICS 17.0 software to generate a 3D reconstruction template for the desired thoracic vertebra. The accurate trajectory and screw diameter and length were calculated with UG Imageware 12.1. The guide template was sterilized and used intraoperatively to assist with the placement of thoracic pedicle screws. After all pedicle trajectory screws had been inserted. We reviewed 12 previous reports and classified them according to the shape and system of PSDT that met the inclusion criteria of the review. Results. Ten screws were placed by using the PSDT without violating the single laminar cortex. There was no violation of the spinal canal or the cortex of pedicle on postoperative CT scans. The results of 13 PSDT types included in the current study suggested that there is no significant difference in accuracy between the PSDTs. Conclusion. The unilateral spinous process noncovering hook type PSDT made through RP provided an accurate trajectory for the thoracic vertebra, and the classification of PSDT in this study could be helpful for further studies.
引用
收藏
页码:E1050 / E1057
页数:8
相关论文
共 43 条
  • [21] Design and Application of a Novel Patient-Specific Three-Dimensional Printed Drill Navigational Guiding in Atlantoaxial Pedicle Screw Placement
    Pu, Xinwei
    Yin, Mengchen
    Ma, Junming
    Liu, Yujie
    Chen, Guanghui
    Huang, Quan
    Zhao, Guoquan
    Lu, Tingsheng
    Yao, Shudan
    Chen, Qinlin
    Luo, Chunshan
    [J]. WORLD NEUROSURGERY, 2018, 114 : E1 - E10
  • [22] A novel Patient-Specific Three-Dimensional Printing Template Based on External Fixation for Pelvic Screw Insertion
    Zhou, Kaihua
    Tao, Xingguang
    Pan, Fugen
    Luo, Congfeng
    Yang, Huilin
    [J]. JOURNAL OF INVESTIGATIVE SURGERY, 2022, 35 (02) : 459 - 466
  • [23] Patient-specific template-guided versus free-hand lumbar pedicle screw implantation - a randomized controlled trial
    Spirig, Jose
    Golshani, Shayan
    Betz, Michael
    Farshad, Nadja
    Farshad, Mazda
    [J]. SWISS MEDICAL WEEKLY, 2019, 149 : 52S - 52S
  • [24] Construction and Accuracy Assessment of Patient-Specific Biocompatible Drill Template for Cervical Anterior Transpedicular Screw (ATPS) Insertion: An In Vitro Study
    Fu, Maoqing
    Lin, Lijun
    Kong, Xiangxue
    Zhao, Weidong
    Tang, Lei
    Li, Jianyi
    Ouyang, Jun
    [J]. PLOS ONE, 2013, 8 (01):
  • [25] Patient- Specific Drill Guide Template for Pedicle Screw Insertion into the Atlantoaxial Cervical Spine Using Stereolithographic Modeling: An In Vitro Study
    Bundoc, Rafael Cruz
    Obenieta, Harjoland Lim
    Dizon, Dave Anthony Gatchalian
    [J]. ASIAN SPINE JOURNAL, 2023, 17 (01) : 8 - 16
  • [26] Additive-manufactured patient-specific titanium templates for thoracic pedicle screw placement: novel design with reduced contact area
    Mitsuru Takemoto
    Shunsuke Fujibayashi
    Eigo Ota
    Bungo Otsuki
    Hiroaki Kimura
    Takeshi Sakamoto
    Toshiyuki Kawai
    Tohru Futami
    Kiyoyuki Sasaki
    Tomiharu Matsushita
    Takashi Nakamura
    Masashi Neo
    Shuich Matsuda
    [J]. European Spine Journal, 2016, 25 : 1698 - 1705
  • [27] Additive-manufactured patient-specific titanium templates for thoracic pedicle screw placement: novel design with reduced contact area
    Takemoto, Mitsuru
    Fujibayashi, Shunsuke
    Ota, Eigo
    Otsuki, Bungo
    Kimura, Hiroaki
    Sakamoto, Takeshi
    Kawai, Toshiyuki
    Futami, Tohru
    Sasaki, Kiyoyuki
    Matsushita, Tomiharu
    Nakamura, Takashi
    Neo, Masashi
    Matsuda, Shuich
    [J]. EUROPEAN SPINE JOURNAL, 2016, 25 (06) : 1698 - 1705
  • [28] Multistep pedicle screw insertion procedure with patient-specific lamina fit-and-lock templates for the thoracic spine Clinical article
    Sugawara, Taku
    Higashiyama, Naoki
    Kaneyama, Shuichi
    Takabatake, Masato
    Watanabe, Naoko
    Uchida, Fujio
    Sumi, Masatoshi
    Mizoi, Kazuo
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2013, 19 (02) : 185 - 190
  • [29] Accuracy of patient-specific template-guided vs. free-hand fluoroscopically controlled pedicle screw placement in the thoracic and lumbar spine: a randomized cadaveric study
    Farshad, Mazda
    Betz, Michael
    Farshad-Amacker, Nadja A.
    Moser, Manuel
    [J]. EUROPEAN SPINE JOURNAL, 2017, 26 (03) : 738 - 749
  • [30] Accuracy of patient-specific template-guided vs. free-hand fluoroscopically controlled pedicle screw placement in the thoracic and lumbar spine: a randomized cadaveric study
    Mazda Farshad
    Michael Betz
    Nadja A. Farshad-Amacker
    Manuel Moser
    [J]. European Spine Journal, 2017, 26 : 738 - 749