Measurement of Occupational Eye and Thyroid Radiation Doses in Pediatric Interventional Cardiologists at a Tertiary Hospital

被引:3
|
作者
Khafaji, Mawya [1 ]
Ashour, Tariq M. [2 ]
Mozahim, Naif F. [2 ]
Tashkandi, Abdullah [2 ]
Alqarni, Basil [2 ]
Malaikah, Abdulrahman A. [2 ]
Bagabas, Abdulaziz K. [2 ]
Alwasabi, Abdullah A. [2 ]
Albahiti, Sarah K. [1 ]
机构
[1] King Abdulaziz Univ, Fac Med, Radiol Dept, Jeddah, Saudi Arabia
[2] King Abdulaziz Univ, Fac Med, Jeddah, Saudi Arabia
关键词
thyroid dose; eye lens dose; staff exposure; intervention pediatric cardiology; occupational radiation dose; cardiac catheterization; STAFF; EXTREMITIES; EXPOSURE; SCATTER;
D O I
10.7759/cureus.44894
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Advances in imaging techniques have led to increased utilization of imaging devices in catheterization laboratories. Invasive surgical procedures for cardiac disorders have been largely replaced by fluoroscopic cardiac catheterization. With this increase, concerns and risks associated with exposure to ionizing radiation among interventional cardiologists are growing. This study aims to measure and compare the occupational doses to the eye lens and thyroid of pediatric interventional cardiologists during different procedures in the catheterization laboratory and its significance. Methodology In this study, cardiologists wore bandanas with attached dosimeters to measure the absorbed doses to the eyes and thyroid gland. The dosimeters were collected for reading. The procedure types were also collected. In addition, the total fluoroscopy time and tube voltage of the biplane machine were measured. SPSS version 23 (IBM Corp., Armonk, NY, USA) was used to analyze the data. The characteristics of the study sample were described using simple counts and percentages, whereas means and standard deviations were used for continuous variables. Statistical significance was set at p-values <0.05. Results A total of 93 procedures were evaluated. The mean absorbed doses for all 93 procedures in both eyes and thyroid were 0.09 mGy and 0.08 mGy, respectively. A significant difference was found between the left and right eye measurements (p = 0.034), with higher doses administered to the left eye. However, no significant difference was observed between the right and left thyroid doses (p = 0.281). Significant correlations were found between the eye and thyroid doses and the procedure type (p = 0.02 and p = 0.009, respectively). Conclusions A significant amount of radiation was measured in the measurements of both organs. In addition, radiation dose measurements varied between different types of procedures. Our current results indicate the importance and necessity of applying the radiation protection concept of dose optimization.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Estimation of eye lens doses received by pediatric interventional cardiologists
    Alejo, L.
    Koren, C.
    Ferrer, C.
    Corredoira, E.
    Serrada, A.
    APPLIED RADIATION AND ISOTOPES, 2015, 103 : 43 - 47
  • [2] EVALUATION OF EYE LENS DOSES OF INTERVENTIONAL CARDIOLOGISTS
    Yokoyama, Sumi
    Suzuki, Shoichi
    Toyama, Hiroshi
    Arakawa, Shinji
    Inoue, Satoshi
    Kinomura, Yutaka
    Kobayashi, Ikuo
    RADIATION PROTECTION DOSIMETRY, 2017, 173 (1-3) : 218 - 222
  • [3] Eye lens radiation exposure to interventional cardiologists: a retrospective assessment of cumulative doses
    Jacob, Sophie
    Donadille, Laurent
    Maccia, Carlo
    Bar, Olivier
    Boveda, Serge
    Laurier, Dominique
    Bernier, Marie-Odile
    RADIATION PROTECTION DOSIMETRY, 2013, 153 (03) : 282 - 293
  • [4] Assessment of radiation doses to cardiologists during interventional examinations
    Bor, D.
    Olgar, T.
    Onal, E.
    Caglan, A.
    Toklu, T.
    MEDICAL PHYSICS, 2009, 36 (08) : 3730 - 3736
  • [5] Occupational doses to the eye lens in pediatric and adult noncardiac interventional radiology procedures
    Morcillo, Ana Belen
    Alejo, Luis
    Huerga, Carlos
    Bayon, Jose
    Marin, Alberto
    Corredoira, Eva
    Novo, Joan Ricardo
    Hernandez, Teresa
    Ponce, Maria Dolores
    Garzon, Gonzalo
    Vano, Eliseo
    Guibelalde, Eduardo
    MEDICAL PHYSICS, 2021, 48 (04) : 1956 - 1966
  • [6] Occupational dose constraints for the lens of the eye for interventional radiologists and interventional cardiologists in the UK
    Mairs, William D. A.
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1062):
  • [7] Measurement of occupational doses of ionising radiation to the lens of the eyes of interventional radiologists
    Higgins, Anthony
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (01) : 74 - 92
  • [8] Assessment of occupational radiation doses to eye lens during interventional radiology procedures
    Ali, Amir
    Al-Nuaimi, Meshari
    Khan, Banaras
    Burki, Shamsullah
    Nishtar, Tahira
    Yaseen, Muhammad
    Shah, Syed Sher Azam
    Irfan, Muhammad
    RADIATION PHYSICS AND CHEMISTRY, 2024, 225
  • [9] Estimation of cardiologists radiation doses received during interventional examinations
    Bor, D.
    Onal, E.
    Olgar, T.
    Caglan, A.
    Toklu, T.
    MEDICAL PHYSICS, 2006, 33 (06) : 2211 - 2212
  • [10] OCCUPATIONAL EXPOSURE TO IONIZING RADIATION AND LENS OPACITY IN INTERVENTIONAL CARDIOLOGISTS
    Domienik-Andrzejewska, Joanna
    Kaluzny, Pawel
    Piernik, Gabriela
    Jurewicz, Joanna
    INTERNATIONAL JOURNAL OF OCCUPATIONAL MEDICINE AND ENVIRONMENTAL HEALTH, 2019, 32 (05) : 663 - 675