Towards 50 years of platelet function analyser (PFA) testing

被引:11
|
作者
Favaloro, Emmanuel J. [1 ,2 ,3 ]
Pasalic, Leonardo [1 ,2 ,4 ]
Lippi, Giuseppe [5 ,6 ]
机构
[1] NSW Hlth Pathol, Inst Clin Pathol & Med Res ICPMR, Haematol, Westmead, NSW 2145, Australia
[2] Sydney Ctr Thrombosis & Haemostasis, Westmead, NSW, Australia
[3] Charles Sturt Univ, Fac Sci & Hlth, Wagga Wagga, NSW, Australia
[4] Univ Sydney, Westmead Clin Sch, Westmead, NSW, Australia
[5] Univ Verona, Sect Clin Biochem, Verona, Italy
[6] Univ Verona, Sch Med, Verona, Italy
关键词
in vitro bleeding time; PFA-100; PFA-200; platelet function analyser; LIGHT TRANSMITTANCE AGGREGOMETRY; VON-WILLEBRAND-DISORDER; BLEEDING-TIME PFA-100; ABO BLOOD-GROUP; IN-VITRO; LABORATORY DIAGNOSIS; CLINICAL UTILITY; CLOSURE TIMES; COAGULATION; PARAMETERS;
D O I
10.1515/cclm-2022-0666
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The platelet function analyser (PFA) is a prevalent platelet function screening instrument, and comes in two models-the original PFA-100 and the contemporary PFA-200. The instruments have 'identical' output, being a 'closure time' (CT). Moreover, normal reference ranges provided by the manufacturer, for the specific test cartridges, are the same for both models. There are three different types of test cartridge: collagen/epinephrine (C/Epi), collagen/adenosine diphosphate (C/ADP), and "Innovance PFA P2Y" (only available in certain geographical locations). The PFA-100 was released in the mid 1990s, and so is approaching 50 years of age. The PFA-200, released in some locations in the mid 2010s, is destined to eventually replace the PFA-100, but is not yet available in the USA. The test system is highly sensitive to von Willebrand disease (VWD; C/Epi and C/ADP) and to aspirin therapy (C/Epi only), but only has moderate sensitivity to defects in platelet function and/or deficiencies in platelet number. Accordingly, recommendations for use for screening platelet function vary according to user experience. Some workers have alternatively used the PFA to assess thrombosis risk or pre-operative bleeding risk. In this review, we provide an overview of the history of PFA, and summarise its current clinical utility.
引用
收藏
页码:851 / 860
页数:10
相关论文
共 50 条
  • [1] Influence of platelet clumps on platelet function analyser (PFA)-200® testing
    Mulliez, S. M. N.
    Hallaert, G.
    Van Roost, D.
    Vantilborgh, A.
    Devreese, K. M. J.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (05) : E103 - E105
  • [2] Performance of the platelet function analyser PFA-100® in testing abnormalities of primary haemostasis
    Harrison, P
    Robinson, MSC
    Mackie, IJ
    Joseph, J
    McDonald, SJ
    Liesner, R
    Savidge, GF
    Pasi, J
    Machin, SJ
    BLOOD COAGULATION & FIBRINOLYSIS, 1999, 10 (01) : 25 - 31
  • [3] Use of the platelet function analyser (PFA-100) in pregnancy
    Davies, JR
    Fernando, R
    Hallworth, SP
    BRITISH JOURNAL OF ANAESTHESIA, 2001, 86 (02) : 310P - 310P
  • [4] Diurnal variation in platelet aggregation with the PFA-100 platelet function analyser
    Dalby, MCD
    Davidson, SJ
    Burman, JF
    Davies, SW
    PLATELETS, 2000, 11 (06) : 320 - 324
  • [5] More on:: platelet function analyser (PFA)-100® closure time in the evaluation of platelet disorders and platelet function
    Favaloro, E. J.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2006, 4 (09) : 2099 - 2100
  • [6] Monitoring platelet function during thrombolytic therapy by PFA-100 platelet function analyser
    Kerényi, A
    Soltész, P
    Muszbek, L
    THROMBOSIS AND HAEMOSTASIS, 1999, : 455 - 455
  • [7] Influence of a low dosage of clopidogrel on platelet function in cats as measured by the platelet function analyser PFA-100 and the multiplate analyser
    Teuber, M.
    Mischke, R.
    RESEARCH IN VETERINARY SCIENCE, 2016, 109 : 149 - 156
  • [8] Use of the platelet function analyser PFA-100™ in juvenile pigs
    Constanze Leuser
    Sibylle Schlottmann
    Ralf Siekmann
    Martin Heidt
    Andreas Moritz
    Natali Bauer
    Comparative Clinical Pathology, 2012, 21 (5) : 761 - 767
  • [9] The effects of ropivacaine hydrochloride on platelet function: an assessment using the platelet function analyser (PFA-100)
    Porter, JM
    Crowe, B
    Cahill, M
    Shorten, GD
    ANAESTHESIA, 2001, 56 (01) : 15 - 18
  • [10] The Platelet Function Analyser (PFA)-100 and von Willebrand disease: a story well over 16 years in the making
    Favaloro, E. J.
    HAEMOPHILIA, 2015, 21 (05) : 642 - 645