摘要
Background: Lag screw osteosynthesis for odontoid fractures has a high rate of pseudoarthrosis, especially in elderly patients. Besides biomechanical properties of the different screw types, insufficient fragment compression or unnoticed screw stripping may be the main causing factors for this adverse event. The aim of the study was to compare two screws in clinical use with different design principles in terms of compression force and stability against screw stripping. Methods: Twelve human cadaveric C2 vertebral bodies were considered. Bone density was determined. The specimens were matched according to bone density and randomly assigned to two experimental groups. An odontoid fracture was induced, which were fixed either with a 3.5 mm standard compression screw or with a 5 mm sleeve nut screw. Both screws are certified for the treatment of odontoid fractures. The bone samples were fixed in a measuring device. The screwdriver was driven mechanically. The tests were analyzed for peak interfragmentary compression and screw-in torque with a frequency of 20 HzFindings: The maximum fragment compression was significantly higher with screw with sleeve nut at 346.13(SD +/- 72.35) N compared with classic compression screw at 162.68(SD +/- 114.13) N (p = 0.025). Screw stripping occurred significantly earlier in classic compression screw at 255.5(SD +/- 192.0)degrees rotation after reaching maximum compression than in screw with sleeve nut at 1005.2(SD +/- 341.1)degrees (p = 0.0039). Interpretation: Screw with sleeve nut achieves greater fragment compression and is more robust to screw stripping compared to classic compression screw. Whether the better biomechanical properties lead to a reduction of pseudoarthrosis has to be proven in clinical studies.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Biomechanical comparison of three different compression screws for treatment of odontoid fractures evaluation of a new screw design
    Mueller, Jan-Uwe
    Mueller, Jonas
    Marx, Sascha
    Matthes, Marc
    Nowak, Stephan
    Schroeder, Henry Werner Siegfried
    Pillich, Dirk Thomas
    CLINICAL BIOMECHANICS, 2020, 77
  • [2] DOUBLE COMPRESSION BONE SCREW OSTEOSYNTHESIS IN TREATMENT FOR FRESH FRACTURES AND PSEUDOARTHROSES OF ODONTOID PROCESS (DENS AXIS)
    BARTH, H
    LANG, G
    ZENTRALBLATT FUR CHIRURGIE, 1989, 114 (04): : 263 - 270
  • [3] A comparison of two headless compression screws for operative treatment of scaphoid fractures
    Ruby Grewal
    Joseph Assini
    David Sauder
    Louis Ferreira
    Jim Johnson
    Kenneth Faber
    Journal of Orthopaedic Surgery and Research, 6
  • [4] A comparison of two headless compression screws for operative treatment of scaphoid fractures
    Grewal, Ruby
    Assini, Joseph
    Sauder, David
    Ferreira, Louis
    Johnson, Jim
    Faber, Kenneth
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2011, 6
  • [5] Osteosynthesis of Phalangeal Fractures: Biomechanical Comparison of Kirschner Wires, Plates, and Compression Screws
    Rausch, Valentin
    Harbrecht, Andreas
    Kahmann, Stephanie L.
    Fenten, Thomas
    Jovanovic, Nebojsa
    Hackl, Michael
    Mueller, Lars P.
    Staat, Manfred
    Wegmann, Kilian
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2020, 45 (10): : 987.e1 - 987.e8
  • [6] Biomechanical Comparison of Cortical Lag Screws and Cortical Position Screws for Their Generation of Interfragmentary Compression and Area of Compression in Simulated Lateral Humeral Condylar Fractures
    Fuchter, Lucy M.
    Wills, Dan
    Walsh, William R.
    Newman, Mark
    Tan, Chris
    VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2023, 36 (05) : 250 - 256
  • [7] Biomechanical evaluation of four different posterior screw and rod fixation techniques for the treatment of the odontoid fractures
    Li, Lei
    Liu, Wen-Fei
    Jiang, Hong-Kun
    Li, Yun-Peng
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (06): : 8571 - 8580
  • [8] Biomechanical comparison of headless compression screws versus independent locking screw for intra-articular fractures
    Meghana Mandala
    Shalin Shaunak
    Paul Kreitmair
    Joideep Phadnis
    Enis Guryel
    European Journal of Orthopaedic Surgery & Traumatology, 2024, 34 : 1319 - 1325
  • [9] Biomechanical comparison of headless compression screws versus independent locking screw for intra-articular fractures
    Mandala, Meghana
    Shaunak, Shalin
    Kreitmair, Paul
    Phadnis, Joideep
    Guryel, Enis
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2024, 34 (03): : 1319 - 1325
  • [10] Treatment of Odontoid Type II Fractures in Octogenarians: Balancing Two Different Treatment Strategies
    Gembruch, Oliver
    Lemonas, Elias
    Ahmadipour, Yahya
    Sure, Ulrich
    El Hindy, Nicolai
    Dodel, Richard
    Mueller, Oliver
    NEUROSPINE, 2019, 16 (02) : 360 - 367