Stroke Volume Measurements by Echocardiography and Ultrasonic Cardiac Output Monitor in Children

被引:0
|
作者
Fremuth, Jiri [1 ,2 ]
Huml, Michal [1 ]
Pomahacova, Tereza [1 ]
Kobr, Jiri [1 ]
Kormunda, Stanislav [1 ]
Sykora, Josef [1 ]
机构
[1] Charles Univ Prague, Fac Med Pilsen, Fac Hosp Pilsen, Dept Pediat, Prague, Czech Republic
[2] Charles Univ Prague, Fac Hosp Pilsen, Fac Med Pilsen, Dept Pediat, Alej Svobody 80, Plzen 30460, Czech Republic
关键词
ultrasonic cardiac output monitor; echocardiography; cardiac output; METAANALYSIS; PRECISION; INDEX;
D O I
10.1097/PEC.0000000000003018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
ObjectivesStroke volume (SV) and cardiac output monitoring is a cornerstone of hemodynamic assessment. Noninvasive technologies are increasingly used in children. This study compared SV measurements obtained by transcutaneous Doppler ultrasound techniques (ultrasonic cardiac output monitor [USCOM]), transthoracic echocardiography jugular (TTE-J), and parasternal (TTE-P) views performed by pediatric intensivists (OP-As) with limited training in cardiac sonography (20 previous examinations) and pediatric cardiologists (OP-Bs) with limited training in USCOM (30 previous examinations) in spontaneously ventilating children.MethodsA single-center study was conducted in 37 children. Each operator obtained 3 sets of USCOM SV measurements within a period of 3 to 5 minutes, followed with TTE measurements from both apical and jugular views. The investigators were blinded to each other's results to prevent visual and auditory bias.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ResultsBoth USCOM and TTE methods were applicable in 89% of patients. The intraobserver variability of USCOM, TTE-J, and TTE-P were less than 10% in both investigators. The SV measurements by OP-As using USCOM, TTE-J, and TTE-P were 46.15 (25.48) mL, 39.45 (20.65) mL, and 33.42 (16.69) mL, respectively. The SV measurements by OP-Bs using USCOM, TTE-J, and TTE-P were 43.99 (25.24) mL, 38.91 (19.98) mL, and 37.58 (19.81) mL, respectively.The percentage error in SV with USCOM relative to TTE-J was 36% in OP-As and 37% in OP-Bs. The percentage error in SV with TTE-P was 33% relative to TTE-J in OP-As and 21% in OP-Bs.ConclusionsOur findings show that the methods are not interchangeable because SV values by USCOM are higher in comparison with the SV values obtained by TTE. Both methods have low level of intraobserver variability. The SV measurements obtained by TTE-P were significantly lower compared with the TTE-J for the operator with limited training in echocardiography. The TTE-P requires longer practice compared with the TTE-J; therefore, we recommend to prefer TTE-J to TTE-P for inexperienced operators.
引用
收藏
页码:680 / 684
页数:5
相关论文
共 50 条
  • [41] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Wonshill Koh
    Kristin A. Schneider
    Huaiyu Zang
    Sarosh P. Batlivala
    Matthew P. Monteleone
    Alexis L. Benscoter
    Meghan M. Chlebowski
    Ilias D. Iliopoulos
    David S. Cooper
    Pediatric Cardiology, 2022, 43 : 1205 - 1213
  • [42] Bioimpedence: A noninvasive measure of stroke volume and cardiac output
    Maslow, Andrew
    JOURNAL OF CLINICAL ANESTHESIA, 2017, 40 : 107 - 109
  • [43] EMBRYONIC STROKE VOLUME AND CARDIAC-OUTPUT IN CHICK
    FABER, JJ
    GREEN, TJ
    THORNBURG, KL
    DEVELOPMENTAL BIOLOGY, 1974, 41 (01) : 14 - 21
  • [44] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Koh, Wonshill
    Schneider, Kristin A.
    Zang, Huaiyu
    Batlivala, Sarosh P.
    Monteleone, Matthew P.
    Benscoter, Alexis L.
    Chlebowski, Meghan M.
    Iliopoulos, Ilias D.
    Cooper, David S.
    PEDIATRIC CARDIOLOGY, 2022, 43 (06) : 1205 - 1213
  • [45] Validation of Noninvasive Measurements of Cardiac Output in Mice Using Echocardiography
    Tournoux, Francois
    Petersen, Bodil
    Thibault, Helene
    Zou, Lin
    Raher, Michael J.
    Kurtz, Baptiste
    Halpern, Elkan F.
    Chaput, Miguel
    Chao, Wei
    Picard, Michael H.
    Scherrer-Crosbie, Marielle
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2011, 24 (04) : 465 - 470
  • [46] Adolescent Children Born Preterm Have An Inhibited Stroke Volume And Cardiac Output Response To Exercise
    Haraldsdottir, Kristin
    Watson, Andrew
    Beshish, Arij
    Tetri, Laura
    Goss, Kara
    Palta, Mari
    Eldridge, Marlowe
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05): : 730 - 731
  • [47] Burn Injury Reduces Cardiac Output and Stroke Volume During Submaximal Aerobic Exercise in Children
    Rivas, Eric
    Bores, Julianna M.
    Herndon, David N.
    Kinsky, Michael
    Suman, Oscar E.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2017, 49 (05): : 905 - 906
  • [48] Electrical velocimetry for noninvasive cardiac output and stroke volume variation measurements in dogs undergoing cardiovascular surgery
    Sasaki, Kazumasu
    Mutoh, Tatsushi
    Mutoh, Tomoko
    Kawashima, Ryuta
    Tsubonez, Hirokazu
    VETERINARY ANAESTHESIA AND ANALGESIA, 2017, 44 (01) : 7 - 16
  • [49] Interchangeability of cardiac output measurements between oesophageal Doppler and pulse contour analysis is dependent on stroke volume
    Lorne, Emmanuel
    Fischer, Marc-Olivier
    ANAESTHESIA CRITICAL CARE & PAIN MEDICINE, 2019, 38 (02) : 113 - 114
  • [50] Cardiac Output Measurements in Young Children: Comment
    Morey, Timothy E.