Biomechanical Research of Three Parallel Cannulated Compression Screws in Oblique Triangle Configuration for Fixation of Femoral Neck Unstable Fractures

被引:1
|
作者
Zhang, Ru-Yi [1 ]
Zhang, Wu-Peng [2 ,3 ,4 ]
Yang, Guang-Min [5 ]
Wang, Dao-Feng [6 ]
Su, Peng [1 ]
Zhang, Yi [1 ]
Nie, Shao-Bo [2 ,3 ]
Li, Jia [2 ,3 ]
Zhao, Zhe [5 ,7 ]
Li, Jian-Tao [2 ,3 ,7 ]
Zhang, Li-Cheng [2 ,3 ,7 ]
Tang, Pei-Fu [2 ,3 ]
机构
[1] Capital Med Univ, Beijing Shijingshan Hosp, Shijingshan Teaching Hosp, Dept Orthopaed, Beijing, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Med Ctr 4, Dept Orthoped, Beijing, Peoples R China
[3] Natl Clin Res Ctr Orthoped Sports Med & Rehabil, Beijing, Peoples R China
[4] Nankai Univ, Sch Med, Tianjin, Peoples R China
[5] Tsinghua Univ, Sch Clin Med, Beijing Tsinghua Changgung Hosp, Dept Orthopaed, Beijing, Peoples R China
[6] Capital Med Univ, Beijing Jishuitan Hosp, Dept Sports Med, Sports Med Serv, Beijing, Peoples R China
[7] Chinese Peoples Liberat Army Gen Hosp, Med Ctr 4, Dept Orthoped, Beijing, Peoples R China
关键词
Biomechanics; Cannulated Compression Screw; Femoral Neck Fracture; Oblique Triangle; Spatial Configuration; INTERNAL-FIXATION; NAIL;
D O I
10.1111/os.14004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Surgical treatment with internal fixation, specifically percutaneous fixation with three cannulated compression screws (CCSs), is the preferred choice for young and middle-aged patients. The mechanical advantage of the optimal spatial configuration with three screws provides maximum dispersion and cortical support. We suspect that the spatial proportion of the oblique triangle configuration (OTC) in the cross-section of the femoral neck isthmus (FNI) may significantly improve shear and fatigue resistance of the fixed structure, thereby stabilizing the internal fixation system in femoral neck fracture (FNF). This study aims to explore the mechanical features of OTC and provide a mechanical basis for its clinical application. Methods: Twenty Sawbone femurs were prepared as Pauwels type III FNF models and divided equally into two fixation groups: OTC and inverted equilateral triangle configuration (IETC). Three 7.3 mm diameter cannulated compression screws (CCSs) were used for fixation. The specimens of FNF after screw internal fixation were subjected to static loading and cyclic loading tests, respectively, with five specimens for each test. Axial stiffness, 5 mm failure load, ultimate load, shear displacement, and frontal rotational angle of two fragments were evaluated. In the cyclic loading test, the load sizes were 700 N, 1400 N, and 2100 N, respectively, and the fracture end displacement was recorded. Results were presented as means +/- SD. Data with normal distributions were compared by the Student's t test. Results: In the static loading test, the axial stiffness, ultimate load, shear displacement, and frontal rotational angle of two fragments were (738.64 vs. 620.74) N/mm, (2957.61 vs. 2643.06) N, (4.67 vs. 5.39) mm, and (4.01 vs. 5.52)degrees (p < 0.05), respectively. Comparison between the femoral head displacement after 10,000 cycles of 700N cyclic loading and total displacement after 20,000 cycles of 700-1400N cyclic loading showed the OTC group was less than the IETC group (p < 0.05). A comparison of femoral head displacement after 10,000 cycles of 1400N and 2100N cycles and total displacement after 30,000 cycles of 700-2100N cycles showed the OTC group was less than another group, but the difference was not significant (p > 0.05). Conclusion: When three CCSs are inserted in parallel to fix FNF, the OTC of three screws has obvious biomechanical advantages, especially in shear resistance and early postoperative weight-bearing, which provides a mechanical basis for clinical selection of ideal spatial configuration for unstable FNF.
引用
收藏
页码:953 / 964
页数:12
相关论文
共 50 条
  • [1] Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation
    Li, Jiantao
    Wang, Menglin
    Zhou, Jianfeng
    Han, Lin
    Zhang, Hao
    Li, Chen
    Li, Lianting
    Hao, Ming
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [2] Ordinary Cannulated Compression Screws or Headless Cannulated Compression Screws? A Synthetic Bone Biomechanical Research in the Internal Fixation of Vertical Femoral Neck Fracture
    Zhang, Baokun
    Liu, Jingwen
    Zhang, Wei
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [3] The Effect of Configuration of Rhombic Cannulated Screws on Internal Fixation of Femoral Neck Fractures
    Guo, Jialiang
    Dong, Weichong
    Yin, Yingchao
    Zhang, Ruipeng
    Hou, Zhiyong
    Zhang, Yingze
    ORTHOPEDICS, 2020, 43 (02) : E72 - E78
  • [4] Fixation of femoral neck fractures using divergent versus parallel cannulated screws
    Papanastassiou I.D.
    Mavrogenis A.F.
    Kokkalis Z.T.
    Nikolopoulos K.
    Skourtas K.
    Papagelopoulos P.J.
    Journal of Long-Term Effects of Medical Implants, 2011, 21 (01) : 63 - 69
  • [5] The oblique triangle configuration of three parallel screws for femoral neck fracture fixation using computer-aided design modules
    Ru Yi Zhang
    Jian Tao Li
    Jing Xin Zhao
    Zhe Zhao
    Li Cheng Zhang
    Cai Yun
    Xiu Yun Su
    Pei Fu Tang
    Scientific Reports, 12
  • [6] The oblique triangle configuration of three parallel screws for femoral neck fracture fixation using computer-aided design modules
    Zhang, Ru Yi
    Li, Jian Tao
    Zhao, Jing Xin
    Zhao, Zhe
    Zhang, Li Cheng
    Yun, Cai
    Su, Xiu Yun
    Tang, Pei Fu
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [7] A Comparative Biomechanical Analysis of Fixation Devices for Unstable Femoral Neck Fractures: The Intertan Versus Cannulated Screws or a Dynamic Hip Screw
    Rupprecht, Martin
    Grossterlinden, Lars
    Ruecker, Andreas H.
    de Oliveira, Alexander Novo
    Sellenschloh, Kay
    Nuechtern, Jakob
    Pueschel, Klaus
    Morlock, Michael
    Rueger, Johannes Maria
    Lehmann, Wolfgang
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 71 (03): : 625 - 634
  • [8] Biomechanical Evaluation of the Modified Cannulated Screws Fixation of Unstable Femoral Neck Fracture with Comminuted Posteromedial Cortex
    Liu, Jingwen
    Zhang, Baokun
    Yin, Bohao
    Chen, Hongchi
    Sun, Hui
    Zhang, Wei
    BIOMED RESEARCH INTERNATIONAL, 2019, 2019
  • [9] Outcomes of three cannulated screws in a modified triangular transverse configuration for fixation of intra-capsular femoral neck fractures
    Assi C.
    Boulos K.
    Yammine K.
    MUSCULOSKELETAL SURGERY, 2023, 107 (4) : 423 - 430
  • [10] Risk Factors for Nonunion in Patients with Intracapsular Femoral Neck Fractures Treated with Three Cannulated Screws Placed in Either a Triangle or an Inverted Triangle Configuration
    Yang, Jui-Jung
    Lin, Leou-Chyr
    Chao, Kuo-Hua
    Chuang, Shih-Youeng
    Wu, Chia-Chun
    Yeh, Tsu-Te
    Lian, Yu-Tung
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (01): : 61 - 69