Biomechanical Research of Three Parallel Cannulated Compression Screws in Oblique Triangle Configuration for Fixation of Femoral Neck Unstable Fractures

被引:1
|
作者
Zhang, Ru-Yi [1 ]
Zhang, Wu-Peng [2 ,3 ,4 ]
Yang, Guang-Min [5 ]
Wang, Dao-Feng [6 ]
Su, Peng [1 ]
Zhang, Yi [1 ]
Nie, Shao-Bo [2 ,3 ]
Li, Jia [2 ,3 ]
Zhao, Zhe [5 ,7 ]
Li, Jian-Tao [2 ,3 ,7 ]
Zhang, Li-Cheng [2 ,3 ,7 ]
Tang, Pei-Fu [2 ,3 ]
机构
[1] Capital Med Univ, Beijing Shijingshan Hosp, Shijingshan Teaching Hosp, Dept Orthopaed, Beijing, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Med Ctr 4, Dept Orthoped, Beijing, Peoples R China
[3] Natl Clin Res Ctr Orthoped Sports Med & Rehabil, Beijing, Peoples R China
[4] Nankai Univ, Sch Med, Tianjin, Peoples R China
[5] Tsinghua Univ, Sch Clin Med, Beijing Tsinghua Changgung Hosp, Dept Orthopaed, Beijing, Peoples R China
[6] Capital Med Univ, Beijing Jishuitan Hosp, Dept Sports Med, Sports Med Serv, Beijing, Peoples R China
[7] Chinese Peoples Liberat Army Gen Hosp, Med Ctr 4, Dept Orthoped, Beijing, Peoples R China
关键词
Biomechanics; Cannulated Compression Screw; Femoral Neck Fracture; Oblique Triangle; Spatial Configuration; INTERNAL-FIXATION; NAIL;
D O I
10.1111/os.14004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Surgical treatment with internal fixation, specifically percutaneous fixation with three cannulated compression screws (CCSs), is the preferred choice for young and middle-aged patients. The mechanical advantage of the optimal spatial configuration with three screws provides maximum dispersion and cortical support. We suspect that the spatial proportion of the oblique triangle configuration (OTC) in the cross-section of the femoral neck isthmus (FNI) may significantly improve shear and fatigue resistance of the fixed structure, thereby stabilizing the internal fixation system in femoral neck fracture (FNF). This study aims to explore the mechanical features of OTC and provide a mechanical basis for its clinical application. Methods: Twenty Sawbone femurs were prepared as Pauwels type III FNF models and divided equally into two fixation groups: OTC and inverted equilateral triangle configuration (IETC). Three 7.3 mm diameter cannulated compression screws (CCSs) were used for fixation. The specimens of FNF after screw internal fixation were subjected to static loading and cyclic loading tests, respectively, with five specimens for each test. Axial stiffness, 5 mm failure load, ultimate load, shear displacement, and frontal rotational angle of two fragments were evaluated. In the cyclic loading test, the load sizes were 700 N, 1400 N, and 2100 N, respectively, and the fracture end displacement was recorded. Results were presented as means +/- SD. Data with normal distributions were compared by the Student's t test. Results: In the static loading test, the axial stiffness, ultimate load, shear displacement, and frontal rotational angle of two fragments were (738.64 vs. 620.74) N/mm, (2957.61 vs. 2643.06) N, (4.67 vs. 5.39) mm, and (4.01 vs. 5.52)degrees (p < 0.05), respectively. Comparison between the femoral head displacement after 10,000 cycles of 700N cyclic loading and total displacement after 20,000 cycles of 700-1400N cyclic loading showed the OTC group was less than the IETC group (p < 0.05). A comparison of femoral head displacement after 10,000 cycles of 1400N and 2100N cycles and total displacement after 30,000 cycles of 700-2100N cycles showed the OTC group was less than another group, but the difference was not significant (p > 0.05). Conclusion: When three CCSs are inserted in parallel to fix FNF, the OTC of three screws has obvious biomechanical advantages, especially in shear resistance and early postoperative weight-bearing, which provides a mechanical basis for clinical selection of ideal spatial configuration for unstable FNF.
引用
收藏
页码:953 / 964
页数:12
相关论文
共 50 条
  • [21] Is stability of femoral neck fractures in the inverted triangle configuration related to the angle between the fracture line and the cannulated compression screws? A finite element analysis
    Niu, Zhipeng
    Wang, Qian
    Yuan, Baoming
    Cui, Yutao
    Ren, Guangkai
    Wu, Dankai
    Peng, Chuangang
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024,
  • [22] Comparison of Early Clinical Results for Femoral Neck System and Cannulated Screws in the Treatment of Unstable Femoral Neck Fractures
    Zhou, Xiao-Qiang
    Li, Zhi-Qiang
    Xu, Ren-Jie
    She, Yuan-Shi
    Zhang, Xiang-Xin
    Chen, Guang-Xiang
    Yu, Xiao
    ORTHOPAEDIC SURGERY, 2021, 13 (06) : 1802 - 1809
  • [23] Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study
    Selvan, VT
    Oakley, MJ
    Rangan, A
    Al-Lami, MK
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2004, 35 (02): : 136 - 141
  • [24] Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
    Zheng Wang
    Yong Yang
    Gangning Feng
    Haohui Guo
    Zhirong Chen
    Yaogeng Chen
    Qunhua Jin
    BioMedical Engineering OnLine, 21
  • [25] Biomechanical comparison of the femoral neck system versus InterTan nail and three cannulated screws for unstable Pauwels type III femoral neck fracture
    Wang, Zheng
    Yang, Yong
    Feng, Gangning
    Guo, Haohui
    Chen, Zhirong
    Chen, Yaogeng
    Jin, Qunhua
    BIOMEDICAL ENGINEERING ONLINE, 2022, 21 (01)
  • [26] A prediction model for osteonecrosis of femoral head after internal fixation with multiple cannulated compression screws for adult femoral neck fractures
    Jin, Zhang
    Chen, Luo
    Wang, Dafeng
    Ye, Yayun
    Fu, Jiaxing
    Yang, Zhifan
    He, Baoqiang
    JOINT DISEASES AND RELATED SURGERY, 2024, 35 (01): : 20 - 26
  • [27] Nonanatomical Reduction of Femoral Neck Fractures in Young Patients (≤65 Years Old) with Internal Fixation Using Three Parallel Cannulated Screws
    Zhao, Guanglei
    Liu, Changquan
    Chen, Kangming
    Lyu, Jinyang
    Chen, Jie
    Shi, Jingsheng
    Chen, Feiyan
    Wei, Yibing
    Wang, Siqun
    Xia, Jun
    Huang, Gangyong
    BIOMED RESEARCH INTERNATIONAL, 2021, 2021
  • [28] Use of fully threaded cannulated screws decreases femoral neck shortening after fixation of femoral neck fractures
    Yoram A. Weil
    Feras Qawasmi
    Meir Liebergall
    Rami Mosheiff
    Amal Khoury
    Archives of Orthopaedic and Trauma Surgery, 2018, 138 : 661 - 667
  • [29] Use of fully threaded cannulated screws decreases femoral neck shortening after fixation of femoral neck fractures
    Weil, Yoram A.
    Qawasmi, Feras
    Liebergall, Meir
    Mosheiff, Rami
    Khoury, Amal
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2018, 138 (05) : 661 - 667
  • [30] Biomechanical and clinical evaluation of interlocking hip screw in Pauwels III femoral neck fractures: A comparison with inverted triangle cannulated screws
    Zhang, Jian
    Jiang, Haozheng
    Dai, Wei
    Hersi, Salad Abdirahman
    Cheong, John Chun Tien Chui Wan
    Chu, Zhenchen
    Lou, Zhiyuan
    Zhang, Deqiang
    Liu, Changjian
    Tian, Kang
    Tang, Xin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10