Commissioning and clinical implementation of an independent dose calculation system for scanned proton beams

被引:0
|
作者
Dreindl, Ralf [1 ,6 ]
Bolsa-Ferruz, Marta [1 ]
Fayos-Sola, Rosa [1 ,2 ]
Cabal, Fatima Padilla [1 ,3 ]
Scheuchenpflug, Lukas [1 ,4 ]
Elia, Alessio [1 ]
Amico, Antonio [1 ,5 ]
Carlino, Antonio [1 ]
Stock, Markus [1 ]
Grevillot, Loic [1 ]
机构
[1] MedAustron Ion Therapy Ctr, Wiener Neustadt, Austria
[2] Hosp Univ La Princesa, Dept Med Phys & Radiat Protect, Madrid, Spain
[3] Med Univ Vienna, AKH Wien, Dept Radiat Oncol, Div Med Radiat Phys, Vienna, Austria
[4] Univ Vienna, Fac Phys, Dept Isotope Phys, Vienna, Austria
[5] IRCCS, Med Phys Dept, Veneto Inst Oncol IOV, Padua, Italy
[6] MedAustron Ion Therapy Ctr, Marie Curie Str 5, A-2700 Wiener Neustadt, Austria
来源
关键词
beam time reduction; clinical implementation; commissioning; independent dose calculation; pencil beam scanning; proton therapy; GAMMA-EVALUATION; THERAPY; VALIDATION; INDEX;
D O I
10.1002/acm2.14328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeExperimental patient-specific QA (PSQA) is a time and resource-intensive process, with a poor sensitivity in detecting errors. Radiation therapy facilities aim to substitute it by means of independent dose calculation (IDC) in combination with a comprehensive beam delivery QA program. This paper reports on the commissioning of the IDC software tool myQA iON (IBA Dosimetry) for proton therapy and its clinical implementation at the MedAustron Ion Therapy Center.MethodsThe IDC commissioning work included the validation of the beam model, the implementation and validation of clinical CT protocols, and the evaluation of patient treatment data. Dose difference maps, gamma index distributions, and pass rates (GPR) have been reviewed. The performance of the IDC tool has been assessed and clinical workflows, simulation settings, and GPR tolerances have been defined.ResultsBeam model validation showed agreement of ranges within +/- 0.2 mm, Bragg-Peak widths within +/- 0.1 mm, and spot sizes at various air gaps within +/- 5% compared to physical measurements. Simulated dose in 2D reference fields deviated by -0.3% +/- 0.5%, while 3D dose distributions differed by 1.8% on average to measurements. Validation of the CT calibration resulted in systematic differences of 2.0% between IDC and experimental data for tissue like samples. GPRs of 99.4 +/- 0.6% were found for head, head and neck, and pediatric CT protocols on a 2%/2 mm gamma criterion. GPRs for the adult abdomen protocol were at 98.9% on average with 3%/3 mm. Root causes of GPR outliers, for example, implants were identified and evaluated.ConclusionIDC has been successfully commissioned and integrated into the MedAustron clinical workflow for protons in 2021. IDC has been stepwise and safely substituting experimental PSQA since February 2021. The initial reduction of proton experimental PSQA was about 25% and reached up to 90% after 1 year.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Commissioning and clinical implementation of myQAiON for proton independent dose calculation (IDC)
    Grevillot, L.
    Dreindl, R.
    Capilla, R. Fayos-Sola
    Elia, A.
    Bolsa-Ferruz, M.
    Gora, J.
    Amico, A.
    Padilla-Cabal, F.
    Carlino, A.
    Stock, M.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S461 - S462
  • [2] Clinical Implementation of Gating and Dose Verification with Scanned Ion Beams at HIT
    Chaudhri, N.
    Richter, D.
    Haertig, M.
    Ecker, S.
    Ackermann, B.
    Naumann, J.
    Haberer, T.
    Bert, C.
    Habermehl, D.
    Herfarth, K.
    Ellerbrock, M.
    Jaekel, O.
    MEDICAL PHYSICS, 2012, 39 (06) : 3780 - 3781
  • [3] Commissioning an Independent Dose Calculation System for the Unity MR-Linac
    Goodwin, E.
    Nill, S.
    Oelfke, U.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S978 - S979
  • [4] Clinical implementation of proton Monte Carlo dose calculation
    Paganetti, H
    Jiang, H
    Kollipara, S
    MEDICAL PHYSICS, 2005, 32 (06) : 2028 - 2028
  • [5] Development of a commissioning tool to implement a proton independent dose calculation software at MedAustron (MA)
    Scheuchenpflug, L.
    Carlino, A.
    Elia, A.
    Boersma, D. J.
    Fuchs, H.
    Stock, M.
    Grevillot, L.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (03) : 312 - 312
  • [6] Validation and Clinical Implementation of a Full Monte Carlo Code for Scanned Proton Pencil Beams
    Huang, S.
    Kang, M.
    Souris, K.
    Ainsley, C.
    Solberg, T.
    McDonough, J.
    Simone, C.
    Lin, L.
    MEDICAL PHYSICS, 2017, 44 (06) : 3180 - 3181
  • [7] Commissioning of IDEAL/GATE-RTion for Proton and Carbon ion Independent Dose Calculation (IDC)
    Grevillot, L.
    Boersma, D.
    Gleyzes, R. Gonzalo
    Scheuchenpflug, L.
    Carlino, A.
    Elia, A.
    Fuchs, H.
    Stock, M.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S109 - S110
  • [8] Out-of-field Dose for Scanned and Scattered Proton Beams
    Moyers, Michael F.
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY, 2009, 1099 : 456 - 459
  • [9] Clinical implementation and commissioning of the MedAustron Particle Therapy Accelerator for non-isocentric scanned proton beam treatments
    Grevillot, Loic
    Osorio Moreno, Jhonnatan
    Letellier, Virgile
    Dreindl, Ralf
    Elia, Alessio
    Fuchs, Hermann
    Carlino, Antonio
    Kragl, Gabriele
    Palmans, Hugo
    Vatnitsky, Stanislav
    Stock, Markus
    MEDICAL PHYSICS, 2020, 47 (02) : 380 - 392
  • [10] 4D DOSE CALCULATION FOR SCANNED PROTON BEAMS USING A DEFORMING DOSE GRID AND SIMULATED 4DCT
    Boye, D.
    Lomax, T.
    Knopf, A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S151 - S152