Using Capillary Whole Blood to Quantitatively Measure Ferritin: A Validation Trial of a Point-of-Care System

被引:2
|
作者
Fiddler, Joanna L. [1 ,2 ]
McBurney, Michael I. [3 ,4 ]
Haas, Jere D. [1 ]
机构
[1] Cornell Univ, Div Nutr Sci, Ithaca, NY 14853 USA
[2] Clemson Univ, Dept Food Nutr & Packaging Sci, Clemson, SC 29634 USA
[3] Univ Guelph, Dept Human Hlth & Nutr Sci, Guelph, ON N1G 2W1, Canada
[4] Tufts Univ, Friedman Sch Nutr Policy & Sci, Boston, MA 02111 USA
关键词
ferritin; diagnostic device; finger stick; blood; capillary; venous; nutritional deficiency; IRON-DEFICIENCY; DONORS; PREVALENCE;
D O I
10.3390/nu15061305
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Iron deficiency is a public health problem with devastating health, developmental and behavioral effects which often exacerbated due to affordability and access to screening and diagnosis. Using IronScan (TM) a portable, point-of-care diagnostic system capable of quantitatively measuring ferritin in blood, we validated IronScan (TM) ferritin measurements using whole blood and serum with a lab-based, regulator-approved analytical device for measuring ferritin in venous serum. Capillary (finger stick) and venous whole blood samples were obtained from 44 male and female volunteers. Venous serum (vSer) ferritin concentrations were measured on Immulite 2000 Xpi (gold standard). Capillary whole blood (cWB), venous whole blood (vWB), and vSer ferritin levels were measured by IronScan (TM). cWB ferritin concentrations from IronScan (TM) were significantly correlated (R-2 = 0.86) with vSer measured with the FDA-approved Immulite system. The results from the multiple regression analysis indicate that 10% of the variability was due to the method of blood collection (venous vs. capillary) and 6% was due to the form of blood analysis (whole blood vs. serum). The sensitivity of diagnosing iron deficiency using the WHO cutoff of <30 ng/mL is 90%, with a specificity of 96%. In conclusion, IronScan (TM) is a rapid viable option for measuring ferritin as a point-of-care system.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Validation of Point-of-Care Ultrasound to Measure Perioperative Edema in Infants With Congenital Heart Disease
    Persson, Jessica N.
    Holstein, Jacqueline
    Silveira, Lori
    Irons, Aimee
    Rajab, Taufiek Konrad
    Jaggers, James
    Twite, Mark D.
    Scahill, Carly
    Kohn, Mary
    Gold, Christine
    Davidson, Jesse A.
    FRONTIERS IN PEDIATRICS, 2021, 9
  • [42] Performance evaluation of the epoc® point-of-care blood analysis system
    Chen, Yu
    Gorman, Monique
    O'Reilly, Bill
    CLINICAL BIOCHEMISTRY, 2015, 48 (15) : 1013 - 1013
  • [43] Hemolysis Rates in Whole Blood Samples for Blood Gas/Electrolyte Analysis by Point-of-Care Testing
    Wilson, Amanda
    Sweeney, Maranna
    Lynch, P. L. Mark
    O'Kane, Maurice J.
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2018, 3 (01): : 144 - 145
  • [44] Using Computable Phenotypes in Point-of-Care Clinical Trial Recruitment
    Chapman, Martin
    Dominguez, Jesus
    Fairweather, Elliot
    Delaney, Brendan C.
    Curcin, Vasa
    PUBLIC HEALTH AND INFORMATICS, PROCEEDINGS OF MIE 2021, 2021, 281 : 560 - 564
  • [45] Emergency medical technician-performed point-of-care blood analysis using the capillary blood obtained from skin puncture
    Kim, Changsun
    Kim, Hansol
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2018, 36 (07): : 1215 - 1221
  • [46] Method validation of GEM® 4000 blood gas analyzer for point-of-care testing
    Martinez Sanchez, L. M.
    Lafuente Muro, S.
    Castillo Ribelles, L.
    Martinez Casademont, M.
    CLINICA CHIMICA ACTA, 2019, 493 : S765 - S765
  • [47] Development of a point-of-care assay system for high-sensitivity C-reactive protein in whole blood
    Ahn, JS
    Choi, S
    Jang, SH
    Chang, HJ
    Kim, JH
    Nahm, KB
    Oh, SW
    Choi, EY
    CLINICA CHIMICA ACTA, 2003, 332 (1-2) : 51 - 59
  • [48] AVOIDING PHLEBOTOMY ENTIRELY: VALIDATION OF A POINT-OF-CARE MEASURE OF APRI & FIB4 FROM FINGER-PRICK BLOOD
    Smookler, David
    Tran, Sarah
    Barber, Bethany
    Saunthar, Ahreni
    Kuczynski, Magdalena
    Mendlowitz, Andrew
    Vanderhoff, Aaron
    Biondi, Mia
    Capraru, Camelia
    Feld, Jordan
    HEPATOLOGY, 2024, 80 : S363 - S364
  • [49] Evaluation of point-of-care capillary and venous blood glucose concentrations in hospitalized neonatal foals
    Moreira, Ana Sofia Dias
    Weng, Hsin-Yi
    Hostnik, Laura D.
    Beasley, Erin M.
    Peek, Simon F.
    Munsterman, Amelia S.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2024,
  • [50] Point-of-care technologies for molecular diagnostics using a drop of blood
    Song, Yujun
    Huang, Yu-Yen
    Liu, Xuewu
    Zhang, Xiaojing
    Ferrari, Mauro
    Qin, Lidong
    TRENDS IN BIOTECHNOLOGY, 2014, 32 (03) : 132 - 139