Using Capillary Whole Blood to Quantitatively Measure Ferritin: A Validation Trial of a Point-of-Care System

被引:2
|
作者
Fiddler, Joanna L. [1 ,2 ]
McBurney, Michael I. [3 ,4 ]
Haas, Jere D. [1 ]
机构
[1] Cornell Univ, Div Nutr Sci, Ithaca, NY 14853 USA
[2] Clemson Univ, Dept Food Nutr & Packaging Sci, Clemson, SC 29634 USA
[3] Univ Guelph, Dept Human Hlth & Nutr Sci, Guelph, ON N1G 2W1, Canada
[4] Tufts Univ, Friedman Sch Nutr Policy & Sci, Boston, MA 02111 USA
关键词
ferritin; diagnostic device; finger stick; blood; capillary; venous; nutritional deficiency; IRON-DEFICIENCY; DONORS; PREVALENCE;
D O I
10.3390/nu15061305
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Iron deficiency is a public health problem with devastating health, developmental and behavioral effects which often exacerbated due to affordability and access to screening and diagnosis. Using IronScan (TM) a portable, point-of-care diagnostic system capable of quantitatively measuring ferritin in blood, we validated IronScan (TM) ferritin measurements using whole blood and serum with a lab-based, regulator-approved analytical device for measuring ferritin in venous serum. Capillary (finger stick) and venous whole blood samples were obtained from 44 male and female volunteers. Venous serum (vSer) ferritin concentrations were measured on Immulite 2000 Xpi (gold standard). Capillary whole blood (cWB), venous whole blood (vWB), and vSer ferritin levels were measured by IronScan (TM). cWB ferritin concentrations from IronScan (TM) were significantly correlated (R-2 = 0.86) with vSer measured with the FDA-approved Immulite system. The results from the multiple regression analysis indicate that 10% of the variability was due to the method of blood collection (venous vs. capillary) and 6% was due to the form of blood analysis (whole blood vs. serum). The sensitivity of diagnosing iron deficiency using the WHO cutoff of <30 ng/mL is 90%, with a specificity of 96%. In conclusion, IronScan (TM) is a rapid viable option for measuring ferritin as a point-of-care system.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Validation and method comparison for a point-of-care lateral flow assay measuring equine whole blood insulin concentrations
    Berryhill, Emily H.
    Urbina, Naomi S.
    Marton, Sam
    Vernau, William
    Alonso, Flavio H.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2023, 35 (02) : 124 - 131
  • [22] Impact of specimen mixing methods on presepsin point-of-care test results using whole blood
    Ham, Ji Yeon
    Song, Kyung Eun
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (05) : E151 - E154
  • [23] AReal Point-of-Care System for Complete Blood Counting
    Ekberg, Bjorn A.
    Larsen, Ulrik D.
    Fogh-Andersen, Niels
    POINT OF CARE, 2005, 4 (01): : 64 - +
  • [24] Validation of the bayer DCA2000™ as a point-of-care measure of glycosylated haemoglobin
    Harris, S
    Tokmakejian, S
    Edmonds, M
    Webster-Bogaert, S
    Mahon, J
    Porter, S
    McDonald, T
    DIABETES, 2000, 49 : A110 - A110
  • [25] Point-of-care, portable microfluidic blood analyzer system
    Maleki, Teimour
    Fricke, Todd
    Quesenberry, J. T.
    Todd, Paul W.
    Leary, James F.
    MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS X, 2012, 8251
  • [26] Introduction of an electrochemical point-of-care assay for quantitative determination of paracetamol in finger-prick capillary whole blood samples
    Kujala, Johanna
    Wester, Niklas
    Lohela, Terhi J.
    Kurkela, Mika
    Backman, Janne T.
    Mikladal, Bjorn
    Laurila, Tomi
    Koskinen, Jari
    Lilius, Tuomas O.
    Kalso, Eija A.
    BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 2023, 89 (09) : 2933 - 2938
  • [27] Accuracy of the HumaSensplus point-of-care uric acid meter using capillary blood obtained by fingertip puncture
    Fabre, Stephanie
    Clerson, Pierre
    Launay, Jean-Marie
    Gautier, Jean-Francois
    Vidal-Trecan, Tiphaine
    Riveline, Jean-Pierre
    Platt, Adam
    Abrahamsson, Anna
    Miner, Jeffrey N.
    Hughes, Glen
    Richette, Pascal
    Bardin, Thomas
    ARTHRITIS RESEARCH & THERAPY, 2018, 20
  • [28] Comparison of a Point-of-Care Whole Blood Beta-Hydroxybutyrate System with a Clinical Laboratory Reference Method
    Chin, A. C.
    Koch, D. D.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A104 - A104
  • [29] Accuracy of the HumaSensplus point-of-care uric acid meter using capillary blood obtained by fingertip puncture
    Stéphanie Fabre
    Pierre Clerson
    Jean-Marie Launay
    Jean-François Gautier
    Tiphaine Vidal-Trecan
    Jean-Pierre Riveline
    Adam Platt
    Anna Abrahamsson
    Jeffrey N. Miner
    Glen Hughes
    Pascal Richette
    Thomas Bardin
    Arthritis Research & Therapy, 20
  • [30] Comparison of a new point-of-care whole blood prothrombin time system to a laboratory reference instrument.
    Spink, BJ
    Schreiner, K
    Buschmann, DJ
    Taborski, U
    BLOOD, 2000, 96 (11) : 102B - 102B