Using Capillary Whole Blood to Quantitatively Measure Ferritin: A Validation Trial of a Point-of-Care System

被引:2
|
作者
Fiddler, Joanna L. [1 ,2 ]
McBurney, Michael I. [3 ,4 ]
Haas, Jere D. [1 ]
机构
[1] Cornell Univ, Div Nutr Sci, Ithaca, NY 14853 USA
[2] Clemson Univ, Dept Food Nutr & Packaging Sci, Clemson, SC 29634 USA
[3] Univ Guelph, Dept Human Hlth & Nutr Sci, Guelph, ON N1G 2W1, Canada
[4] Tufts Univ, Friedman Sch Nutr Policy & Sci, Boston, MA 02111 USA
关键词
ferritin; diagnostic device; finger stick; blood; capillary; venous; nutritional deficiency; IRON-DEFICIENCY; DONORS; PREVALENCE;
D O I
10.3390/nu15061305
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Iron deficiency is a public health problem with devastating health, developmental and behavioral effects which often exacerbated due to affordability and access to screening and diagnosis. Using IronScan (TM) a portable, point-of-care diagnostic system capable of quantitatively measuring ferritin in blood, we validated IronScan (TM) ferritin measurements using whole blood and serum with a lab-based, regulator-approved analytical device for measuring ferritin in venous serum. Capillary (finger stick) and venous whole blood samples were obtained from 44 male and female volunteers. Venous serum (vSer) ferritin concentrations were measured on Immulite 2000 Xpi (gold standard). Capillary whole blood (cWB), venous whole blood (vWB), and vSer ferritin levels were measured by IronScan (TM). cWB ferritin concentrations from IronScan (TM) were significantly correlated (R-2 = 0.86) with vSer measured with the FDA-approved Immulite system. The results from the multiple regression analysis indicate that 10% of the variability was due to the method of blood collection (venous vs. capillary) and 6% was due to the form of blood analysis (whole blood vs. serum). The sensitivity of diagnosing iron deficiency using the WHO cutoff of <30 ng/mL is 90%, with a specificity of 96%. In conclusion, IronScan (TM) is a rapid viable option for measuring ferritin as a point-of-care system.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Capillary blood for point-of-care testing
    Tang, Ruihua
    Yang, Hui
    Choi, Jane Ru
    Gong, Yan
    You, MinLi
    Wen, Ting
    Li, Ang
    Li, XiuJun
    Xu, Bo
    Zhang, Sufeng
    Mei, Qibing
    Xu, Feng
    CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2017, 54 (05) : 294 - 308
  • [2] Validation of 2 point-of-care meters for measuring triglycerides in chickens using whole blood and plasma
    Irvine, Katherine L.
    Mans, Christoph
    Friedrichs, Kristen R.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2018, 30 (02) : 197 - 204
  • [3] A novel point-of-care device accurately measures thyrotropin in whole blood, capillary blood and serum
    Kahaly, George J.
    Lotz, Johannes
    Walder, Sara
    Hammad, Cara
    Kraemer, Rebecca
    Frommer, Lara
    Koenig, Jochem
    Wolf, Jan
    Gottwald-Hostalek, Ulrike
    Urgatz, Bogumila
    Lackner, Karl J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2022, 60 (10) : 1607 - 1616
  • [4] Ferritin Determination in Blood Using a Point-of-Care Surface Plasmon Resonance Sensor
    Dubois, C.
    Robidoux, J.
    Masson, J.
    Brouard, D.
    TRANSFUSION, 2023, 63
  • [5] Whole-Blood Validation of a New Point-of-care Equine Serum Amyloid A Assay
    Karam, Bruno
    Hines, Siddra
    Skipper, Lauren
    Pusterla, Nicola
    JOURNAL OF EQUINE VETERINARY SCIENCE, 2020, 94
  • [6] A Novel Whole Blood Point-of-Care Coagulometer to Measure the Effect of Direct Oral Anticoagulants and Heparins
    Ansell, Jack
    Zappe, Stefan
    Jiang, Xuan
    Chen, Lirong
    Steiner, Solomon
    Laulicht, Bryan
    Bakhru, Sasha
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2019, 45 (03): : 259 - 263
  • [7] A lab-on-a-CD system for point-of-care testing of whole blood samples
    Jun Yang
    医用生物力学, 2013, 28(S1) (S1) : 79 - 79
  • [8] Using Sequence Analyses to Quantitatively Measure Oropharyngeal Swallowing Temporality in Point-of-Care Ultrasound Examinations: A Pilot Study
    Lam, Wilson Yiu Shun
    Kwong, Elaine
    Chan, Huberta Wai Tung
    Zheng, Yong-Ping
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (08)
  • [9] Accuracy and validation of a point-of-care blood glucose monitoring system for use in horses
    Velineni, Sridhar
    Schiltz, Paul
    Chang, Ko-Hsin
    Peng, Yi-Ming
    Cowles, Bobby
    FRONTIERS IN VETERINARY SCIENCE, 2024, 11
  • [10] Validation of a new point-of-care assay for determination of ß-carotene concentration in bovine whole blood and plasma
    Raila, Jens
    Enjalbert, Francis
    Mothes, Ralf
    Hurtienne, Andrea
    Schweigert, Florian J.
    VETERINARY CLINICAL PATHOLOGY, 2012, 41 (01) : 119 - 122