Blue-Light Filtering Monofocal Intraocular Lenses: A Study on Optical Function and Tolerance to Misalignment

被引:3
|
作者
Yan, Weijia [1 ]
Auffarth, Gerd U. [1 ]
Khoramnia, Ramin [1 ]
Labuz, Grzegorz [1 ]
机构
[1] Univ Hosp Heidelberg, David J Apple Ctr Vis Res, Dept Ophthalmol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
关键词
SPHERICAL-ABERRATION; CATARACT-SURGERY; VISUAL-ACUITY; IMAGE QUALITY; MODEL; DECENTRATION; FOCUS; PERFORMANCE; IMPACT; DEPTH;
D O I
10.3928/1081597X-20240112-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To investigate the optical performance and toler- ance to misalignment of blue -light filtering monofocal intra- ocular lenses (IOLs). METHODS: The optical properties of two monofocal IOLs featuring yellow chromophores, CT Lucia 621 PY (Carl Zeiss Meditec AG) and Clareon CNA0T0 (Alcon Laboratories, Inc), were assessed in monochromatic and polychromatic light while introducing spherical aberration (SA). Optical quality metrics derived from the modulation transfer function were assessed after optimal IOL centration at 3- and 4.5 -mm pu- pils. In addition, each IOL's tolerance to misalignment was ex- amined by inducing up to 1 mm of decentration and the effect of tilting it by 5 degrees at 3 mm. RESULTS: The IOLs' resolution and contrast, while tested us- ing a 3 -mm aperture and an SA -neutral corneal model, indi- cated the CT Lucia 621 PY had a slightly higher modulation transfer function (MTF) at 50 lp/mm than the CNA0T0 under monochromatic conditions (0.77 vs 0.69). On introducing SA with (0.49 vs 0.40) and without (0.75 vs. 0.70) chromatic ab- erration, the CT Lucia 621 PY maintained its minimally bet- ter performance. When assessed with a 4.5 -mm aperture in monochromatic light, the CT Lucia 621 PY displayed improved MTF with aberration -free cornea (0.71 vs 0.40) but performed worse after introducing SA (0.44 vs 0.62). However, both lenses achieved comparable MTF values under spherical and chromatic aberrations (0.28 vs 0.27). The IOL misalignment test revealed a better tolerance to tilt and decentration of the CT Lucia 621 PY across all conditions. CONCLUSIONS: The CT Lucia 621 PY and CNA0T0 showed similar optical quality in different situations, with equal sim- ulated distance visual acuity for both models. However, the CT Lucia 621 PY's aspheric design offers an advantage when dealing with often imperfect physiological conditions, display- ing a more robust performance under tilt and decentration. [J Refract Surg. 2024;40(2):e79 -e88.]
引用
收藏
页码:e79 / e88
页数:10
相关论文
共 50 条
  • [1] Do Blue-Light Filtering Intraocular Lenses Affect Visual Function?
    Lavric, Alenka
    Pompe, Manca Tekavcic
    OPTOMETRY AND VISION SCIENCE, 2014, 91 (11) : 1348 - 1354
  • [2] The Effect of Blue-Light Filtering Intraocular Lenses on the Development and Progression of Glaucoma
    Hecht, Idan
    Kanclerz, Piotr
    Achiron, Asaf
    Elbaz, Uri
    Tuuminen, Raimo
    JOURNAL OF GLAUCOMA, 2023, 32 (06) : 451 - 457
  • [3] Blue-light filtering intraocular lenses (IOLs) for protecting macular health
    Downie, Laura E.
    Busija, Ljoudmila
    Keller, Peter R.
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2018, (05):
  • [4] Blue-Light Filtering Spectacle Lenses: Optical and Clinical Performances
    Leung, Tsz Wing
    Li, Roger Wing-hong
    Kee, Chea-su
    PLOS ONE, 2017, 12 (01):
  • [5] Blue-light filtering ophthalmic lenses: to prescribe, or not to prescribe?
    Downie, Laura E.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2017, 37 (06) : 640 - 643
  • [6] Drusen quantification in post-mortem eyes implanted with or without blue-light filtering intraocular lenses
    Nassrallah, Emmanuel Issa Bassam
    Youhnovska, Emma Nikolaeva
    Abdouh, Mohamed
    Marcotte, Emily
    Mastromonaco, Christina
    Burnier, Miguel N.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [7] The Effects of Blue-Light Filtering Intraocular Implants on Glare Geometry
    Hammond, Billy R.
    Gardner, Colin R.
    Renzi-Hammond, Lisa
    CURRENT EYE RESEARCH, 2023, 48 (07) : 639 - 644
  • [8] The Effects of Blue Light-Filtering Intraocular Lenses on the Protection and Function of the Visual System
    Hammond, Billy R.
    Sreenivasan, Vidhyapriya
    Suryakumar, Rajaraman
    CLINICAL OPHTHALMOLOGY, 2019, 13 : 2427 - 2438
  • [9] In vivo evaluation of blue-light attenuation with tinted and untinted intraocular lenses
    Kontadakis, Georgios A.
    Plainis, Sotiris
    Moschandreas, Joanna
    Tsika, Chrisanthi
    Pallikaris, Ioannis G.
    Tsilimbaris, Miltiadis K.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (06): : 1031 - 1037
  • [10] Determination of the action spectrum of the blue-light hazard for different intraocular lenses
    Pons, Alicia
    Delgado, Diego
    Campos, Joaquin
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2007, 24 (06) : 1545 - 1550