An inversion decomposition method for better energy resolution of NaI(Tl) scintillation detectors based on a Gaussian response matrix

被引:1
|
作者
Jian-Feng He [1 ,2 ]
Yao-Zong Yang [3 ]
Jin-Hui Qu [1 ]
Qi-Fan Wu [2 ]
Hai-Ling Xiao [1 ]
Cong-Cong Yu [4 ]
机构
[1] Fundamental Science on Radioactive Geology and Exploration Technology Laboratory,East China Institute of Technology
[2] Department of Engineering Physics,Tsinghua University
[3] Jiujiang Power Supply Subsidiary Company of State Grid Jianxi Electronic Power Company
[4] Sichuan Integrative Medicine Hospital Sub-health Testing and Evaluation Laboratory
基金
中国国家自然科学基金;
关键词
Na I(Tl) scintillation detector; Gamma-ray energy spectrum; Gaussian response matrix; Inversion decomposition;
D O I
暂无
中图分类号
TL812 [闪烁探测技术和仪器];
学科分类号
082704 ;
摘要
NaI(T1) scintillation detectors have been widely applied for gamma-ray spectrum measurements owing to advantages such as high detection efficiency and low price.However,the mitigation of the limited energy resolution of these detectors,which detracts from an accurate analysis of the instrument spectra obtained,remains a crucial need.Based on the physical properties and spectrum formation processes of NaI(T1) scintillation detectors,the detector response to gamma photons with different energies is represented by photopeaks that are approximately Gaussian in shape with unique full-width-at-half-maximum(FWHM) values.The FWHM is established as a detector parameter based on resolution calibrations and is used in the construction of a general Gaussian response matrix,which is employed for the inverse decomposition of gamma spectra obtained from the detector.The Gold and Boosted Gold iterative algorithms are employed to accelerate the decomposition of the measured spectrum.Tests of the inverse decomposition method on multiple simulated overlapping peaks and on experimentally obtained U and Th radionuclide series spectra verify the practicability of the method,particularly in the low-energy region of the spectrum,providing for the accurate qualitative and quantitative analysis of radionuclides.
引用
收藏
页码:111 / 120
页数:10
相关论文
共 48 条
  • [11] THE PROMPT RESPONSE OF BISMUTH-GERMANATE AND NAI(TL) SCINTILLATION DETECTORS TO FAST-NEUTRONS
    HAUSSER, O
    LONE, MA
    ALEXANDER, TK
    KUSHNERIUK, SA
    GASCON, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 213 (2-3): : 301 - 309
  • [12] CONTRIBUTION OF PHOTOCATHODE NONUNIFORMITY TO ENERGY RESOLUTION IN NAI(T1) SCINTILLATION DETECTORS
    FRASERMITCHELL, J
    WRIGHT, AG
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 288 (2-3): : 429 - 438
  • [13] Comparison of two spectrum-dose conversion methods based on NaI(Tl) scintillation detectors
    Yudong, W.
    Xiaobing, L.
    Chi, Z.
    Rong, Z.
    Chaowen, Y.
    JOURNAL OF INSTRUMENTATION, 2018, 13
  • [14] Model-based pulse deconvolution method for NaI(Tl) detectors
    Xiao, Wuyun
    Farsoni, Abi T.
    Yang, Haori
    Hamby, David M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 769 : 5 - 8
  • [15] Calculation of spectrum to dose conversion factors for a NaI(Tl) scintillation detector using the response matrix
    Jun, JS
    Yi, CY
    Chai, HS
    Cho, H
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 1995, 28 (06) : 716 - 726
  • [16] Nonproportional scintillation response of NaI:Tl to low energy x-ray photons and electrons
    Khodyuk, I. V.
    Rodnyi, P. A.
    Dorenbos, P.
    JOURNAL OF APPLIED PHYSICS, 2010, 107 (11)
  • [17] A New Temperature Correction Method for NaI(Tl) Detectors Based on Pulse Deconvolution
    Xie, Jianming
    Yang, Liu
    Li, Jinglun
    Qi, Sheng
    Chen, Wenzhuo
    Xu, Hang
    Xiao, Wuyun
    SENSORS, 2023, 23 (11)
  • [18] Multilayer passive shielding of scintillation detectors based on BGO, NaI(Tl), and stilbene crystals operating in intense neutron fields with an energy of 14.1 MeV
    Bystritsky V.M.
    Valkovic V.
    Grozdanov D.N.
    Zontikov A.O.
    Ivanov I.Z.
    Kopatch Y.N.
    Krylov A.R.
    Rogov Y.N.
    Ruskov I.N.
    Sapozhnikov M.G.
    Skoy V.R.
    Shvetsov V.N.
    Physics of Particles and Nuclei Letters, 2015, 12 (2) : 325 - 335
  • [19] Inversion Correction Method for NaI(Tl) Gamma Spectra on Arbitrary Energy Scale Based on the Function Theory of Random Variables
    Chongjie Wang
    Yuxin Miao
    Conger Wang
    Qiuyan Zhang
    Yue Sun
    Min Zhang
    Instruments and Experimental Techniques, 2022, 65 : 362 - 369
  • [20] Inversion Correction Method for NaI(Tl) Gamma Spectra on Arbitrary Energy Scale Based on the Function Theory of Random Variables
    Wang, Chongjie
    Miao, Yuxin
    Wang, Conger
    Zhang, Qiuyan
    Sun, Yue
    Zhang, Min
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2022, 65 (02) : 362 - 369