Measurement of central corneal thickness and precorneal tear film thickness of rabbits using the Scheimpflug system

被引:0
|
作者
Jing Dong [1 ]
Qiang Wu [2 ]
Xiao-Gang Wang [2 ]
机构
[1] Department of Ophthalmology,the First Hospital of Shanxi Medical University
[2] Department of Ophthalmology,the Sixth People's Hospital Affiliated to Shanghai Jiaotong University
关键词
tear film; Scheimpflug; rabbits; central corneal thickness; fluorescein;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
AIM:To measure central corneal thickness(CCT) and pre-corneal tear film thickness using the Galilei dual Scheimpflug analyzer(GSA) in New Zealand White rabbits.METHODS:Ten normal New Zealand White rabbits(20 eyes) were included in this study.With the assistance of 0.1% fluorescein,the pre-corneal tear film can be well visualized.Both eyes of each rabbit were scanned once with the GSA pre- and post-instillation of 1μL 0.1% fluorescein.The difference between the two measurements of CCT(4-mm diameter) was recorded as the pachymetric values of the central tear film.RESULTS:The CCT of pre- and post-instillation was 388.8±9.5μm and 407.0±10.5μm,respectively.After a paired t-test analysis,the central pre-corneal tear film thickness of 4mm diameter was 18.2±5.31μm with a 95% confidence interval of(15.7,20.6)μm(P<0.001).CONCLUSION:GSA can be used to measure CCT and analyze central tear film thickness of rabbits with the help of fluorescein.
引用
收藏
页码:584 / 587
页数:4
相关论文
共 50 条
  • [41] IOP Measurement and Central Corneal Thickness Reply
    Park, Serena J. K.
    Ang, Ghee Soon
    Wells, Anthony P.
    [J]. OPHTHALMOLOGY, 2012, 119 (12) : 2648 - 2649
  • [42] IOP measurement and central corneal thickness - Reply
    Feltgen, N
    Funk, J
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2002, 86 (01) : 121 - 121
  • [43] Measurement of central corneal thickness in health and disease
    Chaudhry, Imtiaz A.
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2009, 23 (3-4) : 179 - 180
  • [44] Central corneal thickness measurement in clinical practice
    Shildkrot, Y
    Liebmann, JM
    Fabijanczyk, B
    Tello, CA
    Ritch, R
    [J]. JOURNAL OF GLAUCOMA, 2005, 14 (05) : 331 - 336
  • [45] Assessments of tear meniscus height, tear film thickness, and corneal epithelial thickness after deep anterior lamellar keratoplasty
    Xie, Wen-jia
    Xu, Ye-sheng
    Zhang, Xia
    Yao, Yu-feng
    [J]. JOURNAL OF ZHEJIANG UNIVERSITY-SCIENCE B, 2018, 19 (03): : 218 - 226
  • [46] Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography
    Wang, JH
    Fonn, D
    Simpson, TL
    Jones, L
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (06) : 2524 - 2528
  • [47] Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography
    Huang, Jinhai
    Ding, Xixia
    Savini, Giacomo
    Jiang, Zhengxuan
    Pan, Chao
    Hua, Yanjun
    Wu, Fan
    Feng, Yifan
    Yu, Ye
    Wang, Qinmei
    [J]. PLOS ONE, 2014, 9 (05):
  • [48] Agreement and reliability in measuring central corneal thickness with a rotating Scheimpflug-Placido system and ultrasound pachymetry
    Maresca, Nunzio
    Zeri, Fabrizio
    Palumbo, Paolo
    Calossi, Antonio
    [J]. CONTACT LENS & ANTERIOR EYE, 2014, 37 (06): : 442 - 446
  • [49] Central corneal thickness measurement using a pachometer: Mean or lowest values?
    Haun, AM
    Gunvant, P
    Baskaran, M
    Vijaya, L
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U131 - U131
  • [50] Film Thickness Measurement System using FPGA Technology
    Riano, Adriana Bonilla
    Bannwart, Antonio
    Velasco, Hugo Fernando
    Rodriguez, O. M. H.
    Prasser, Horst-Michael
    [J]. 2015 20TH SYMPOSIUM ON SIGNAL PROCESSING, IMAGES AND COMPUTER VISION (STSIVA), 2015,