Dosimetric response of Gafchromic™ EBT-XD film to therapeutic protons

被引:10
|
作者
Guan, Fada [1 ,2 ]
Wang, Xiaochun [1 ]
Yang, Ming [1 ]
Draeger, Emily [2 ]
Han, Dae [2 ]
Iga, Kiminori [1 ,3 ]
Guo, Fanqing [2 ]
Perles, Luis [1 ]
Li, Yuting [1 ]
Sahoo, Narayan [1 ]
Mohan, Radhe [1 ]
Chen, Zhe [2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, 1515 Holcombe Blvd, Houston, TX 77030 USA
[2] Yale Univ, Sch Med, Dept Therapeut Radiol, 333 Cedar St, New Haven, CT 06510 USA
[3] Hitachi Amer Ltd, Particle Therapy Div, Santa Clara, CA USA
来源
PRECISION RADIATION ONCOLOGY | 2023年 / 7卷 / 01期
关键词
EBT-XD film; FLASH dose rate; linear energy transfer; proton therapy; DEPENDENCE; MODEL;
D O I
10.1002/pro6.1187
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The EBT-XD model of Gafchromic (TM) films has a broader optimal dynamic dose range, up to 40 Gy, compared with its predecessor models. This characteristic has made EBT-XD films suitable for high-dose applications, such as stereotactic body radiotherapy and stereotactic radiosurgery, as well as ultra-high dose rate FLASH radiotherapy. The purpose of the current study was to characterize the dependence of EBT-XD film response on linear energy transfer (LET) and dose rate of therapeutic protons from a synchrotron. A clinical spot-scanning proton beam was used to study LET dependence at three dose-averaged LET values of 1.0 keV/mu m, 3.6 keV/mu m, and 7.6 keV/mu m. A research proton beamline was used to study dose rate dependence at 150 Gy/s in the FLASH mode and 0.3 Gy/s in the non-FLASH mode. Film response data from dose-averaged LET values of 0.9 keV/mu m and 9.0 keV/mu m of the proton FLASH beam were also compared. Film response data from a clinical 6-MV photon beam were used as a reference. Both the gray value method and optical density (OD) method were used in film calibration. Calibration results using a specific OD calculation method and a generic OD calculation method were compared. The four-parameter NIH Rodbard function and three-parameter rational function were compared in fitting the calibration curves. Experimental results showed that the response of EBT-XD film is proton LET dependent, but independent of dose rate. Goodness-of-fit analysis showed that using the NIH Rodbard function is superior for both protons and photons. Using the "specific OD + NIH Rodbard function" method for EBT-XD film calibration is recommended. The results of dose-rate independence of EBT-XD film to protons are shown in Figure G1. The FLASH dose rate is 150 Gy/s and the non-FLASH dose rate is 0.3 Gy/s. The linear energy transfer dependence of EBT-XD film to proton beams is shown in Figure G2. Film calibration curves from three different dose-averaged linear energy transfer values (1.0 keV/mu m, 3.6 keV/mu m, and 7.6 keV/mu m) using clinical 87.2 MeV protons are compared in Figure G2(A). Film calibration curves from two dose-averaged linear energy transfer values (0.9 keV/mu m and 9.0 keV/mu m) using FLASH 87.2 MeV protons are compared in Figure G2(B). image
引用
收藏
页码:15 / 26
页数:12
相关论文
共 50 条
  • [41] Evaluation of Dosimetric Characteristics of Electron Beam Arc Therapy Using Gafchromic EBT Film
    Swamidas, J.
    Upreti, R.
    Sharma, P.
    Sharma, S.
    Pai, R.
    Deshpande, D.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [42] Dosimetry characteristics of GAFCHROMIC® EBT film responding to therapeutic electron beams
    Su, Fan-Chi
    Liu, Yaxi
    Stathakis, Sotirlos
    Shi, Chengyu
    Esquivel, Carlos
    Papanikolaou, Niko
    APPLIED RADIATION AND ISOTOPES, 2007, 65 (10) : 1187 - 1192
  • [43] Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
    Borca, Valeria Casanova
    Pasquino, Massimo
    Russo, Giuliana
    Grosso, Pierangelo
    Cante, Domenico
    Sciacero, Piera
    Girelli, Giuseppe
    La Porta, Maria Rosa
    Tofani, Santi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 158 - 171
  • [44] Dosimetric Verification of the Valencia Skin Applicator Using Gafchromic EBT3 Film
    Lindsey, A.
    Lamberto, M.
    Chen, H.
    Sarkar, A.
    Mourtada, F.
    MEDICAL PHYSICS, 2015, 42 (06) : 3535 - 3535
  • [45] Dosimetric characteristics of gafchromic EBT3 film on small field electron beam
    Ulya, S.
    Wibowo, W. E.
    Nuruddin, N.
    Pawiro, S. A.
    10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [46] Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: A preliminary study
    Bakar, K. S. A. K.
    Noor, N. M.
    Hassan, M. F.
    Idris, F.
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [47] Effect of glass compression plate on EBT-XD film dosimetry for pretreatment quality assurance of stereotactic body radiotherapy
    Raj, Sathiya
    Shree, Nithya
    Ganesh, K. M.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2024, 29 (03) : 357 - 361
  • [48] Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films
    Schoenfeld, Andreas A.
    Wieker, Soeren
    Harder, Dietrich
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14): : 5426 - 5442
  • [49] Comparison of the Absorption Spectra of the EBT-XD and EBT3 Films Irradiated with Clinical Radiotherapy Beams
    Leon, Y.
    Mulrow, D.
    Khan, R.
    Darafsheh, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E621 - E621
  • [50] An evaluation of the use of EBT-XD film for SRS/SBRT commissioning of a 1.5 Tesla MR-Linac system
    Lim, Seng Boh
    Tyagi, Neelam
    Subashi, Ergys
    Liang, Jiayi
    Chan, Maria
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 96 : 9 - 17