Dosimetric response of Gafchromic™ EBT-XD film to therapeutic protons

被引:10
|
作者
Guan, Fada [1 ,2 ]
Wang, Xiaochun [1 ]
Yang, Ming [1 ]
Draeger, Emily [2 ]
Han, Dae [2 ]
Iga, Kiminori [1 ,3 ]
Guo, Fanqing [2 ]
Perles, Luis [1 ]
Li, Yuting [1 ]
Sahoo, Narayan [1 ]
Mohan, Radhe [1 ]
Chen, Zhe [2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, 1515 Holcombe Blvd, Houston, TX 77030 USA
[2] Yale Univ, Sch Med, Dept Therapeut Radiol, 333 Cedar St, New Haven, CT 06510 USA
[3] Hitachi Amer Ltd, Particle Therapy Div, Santa Clara, CA USA
来源
PRECISION RADIATION ONCOLOGY | 2023年 / 7卷 / 01期
关键词
EBT-XD film; FLASH dose rate; linear energy transfer; proton therapy; DEPENDENCE; MODEL;
D O I
10.1002/pro6.1187
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The EBT-XD model of Gafchromic (TM) films has a broader optimal dynamic dose range, up to 40 Gy, compared with its predecessor models. This characteristic has made EBT-XD films suitable for high-dose applications, such as stereotactic body radiotherapy and stereotactic radiosurgery, as well as ultra-high dose rate FLASH radiotherapy. The purpose of the current study was to characterize the dependence of EBT-XD film response on linear energy transfer (LET) and dose rate of therapeutic protons from a synchrotron. A clinical spot-scanning proton beam was used to study LET dependence at three dose-averaged LET values of 1.0 keV/mu m, 3.6 keV/mu m, and 7.6 keV/mu m. A research proton beamline was used to study dose rate dependence at 150 Gy/s in the FLASH mode and 0.3 Gy/s in the non-FLASH mode. Film response data from dose-averaged LET values of 0.9 keV/mu m and 9.0 keV/mu m of the proton FLASH beam were also compared. Film response data from a clinical 6-MV photon beam were used as a reference. Both the gray value method and optical density (OD) method were used in film calibration. Calibration results using a specific OD calculation method and a generic OD calculation method were compared. The four-parameter NIH Rodbard function and three-parameter rational function were compared in fitting the calibration curves. Experimental results showed that the response of EBT-XD film is proton LET dependent, but independent of dose rate. Goodness-of-fit analysis showed that using the NIH Rodbard function is superior for both protons and photons. Using the "specific OD + NIH Rodbard function" method for EBT-XD film calibration is recommended. The results of dose-rate independence of EBT-XD film to protons are shown in Figure G1. The FLASH dose rate is 150 Gy/s and the non-FLASH dose rate is 0.3 Gy/s. The linear energy transfer dependence of EBT-XD film to proton beams is shown in Figure G2. Film calibration curves from three different dose-averaged linear energy transfer values (1.0 keV/mu m, 3.6 keV/mu m, and 7.6 keV/mu m) using clinical 87.2 MeV protons are compared in Figure G2(A). Film calibration curves from two dose-averaged linear energy transfer values (0.9 keV/mu m and 9.0 keV/mu m) using FLASH 87.2 MeV protons are compared in Figure G2(B). image
引用
收藏
页码:15 / 26
页数:12
相关论文
共 50 条
  • [31] Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry
    Rebecca Moylan
    Trent Aland
    Tanya Kairn
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 331 - 337
  • [32] Monte Carlo Calculated Stopping-Power of EBT3 and EBT-XD Film for Pristine Proton Beams
    Shi, C.
    Chen, C.
    Mah, D.
    Chan, M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [33] Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance
    Fuss, Martina
    Sturtewagen, Eva
    De Wagter, Carlos
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14): : 4211 - 4225
  • [34] Neutron response of GafChromic® EBT2 film
    Hsiao, Ming-Chen
    Liu, Yuan-Hao
    Chen, Wei-Lin
    Jiang, Shiang-Huei
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05): : 1391 - 1413
  • [35] Technical note: Proton beam dosimetry at ultra-high dose rates (FLASH): Evaluation of GAFchromic™ (EBT3, EBT-XD) and OrthoChromic (OC-1) film performances
    Institut de Cancérologie de l'Ouest, Saint-Herblain, France
    不详
    不详
    Med. Phys., 4 (2732-2745):
  • [36] Temporal Growth of the Polymerization in Irradiated EBT3 and EBT-XD Radiochromic Films
    Hao, Y.
    Momin, S.
    Williamson, J.
    Li, H.
    Khan, R.
    Darafsheh, A.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [37] Dosimetric Verification of IMRT Fields Using GafChromic® EBT-2 Film
    McLawhorn, R.
    Ter-Antonyan, R.
    Benedict, S.
    Wijesooriya, K.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [38] Spectral characterization and comparison of EBT3, EBT4, and EBT-XD radiochromic films
    Pecic, S.
    Devic, S.
    Belca, I
    Mosic, M.
    Kurij, Lj
    Nidzovic, B.
    Stojadinovic, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (01):
  • [39] An Evaluation of the Use of EBT-XD Film for SRS/SBRT Commissioning of Unity MR-Linac System
    Lim, S.
    Tyagi, N.
    Subashi, E.
    Liang, J.
    Chan, M.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [40] AcurosXB dose verification of ultra-small lung lesions with EBT-XD film in a heterogeneous phantom
    Ollers, M.
    Swinnen, A.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S479 - S479