Influence of template design on the accuracy of static computer-assisted implant surgery

被引:0
|
作者
Chung, Jinmook [1 ]
Park, Junghyun [1 ]
Son, Hyeonseong [1 ]
Hong, Seoung-Jin [2 ]
Pae, Ahran [2 ]
机构
[1] Kyung Hee Univ, Grad Sch, Dept Dent, Seoul, South Korea
[2] Kyung Hee Univ, Coll Dent, Dept Prosthodont, 26 Kyungheedae Ro, Seoul 02447, South Korea
来源
JOURNAL OF ADVANCED PROSTHODONTICS | 2025年 / 17卷 / 01期
关键词
Guided surgery; Implant; Open sleeve; Static computer-assisted implant surgery (s-CAIS); Surgical template; STEREOLITHOGRAPHIC SURGICAL GUIDES; PLACEMENT; TOMOGRAPHY; TOLERANCE; PRECISION; IMAGES;
D O I
10.4047/jap.2025.17.1.22
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This study aimed to evaluate the accuracy of static computer- assisted implant surgery (s-CAIS) by comparing the planned and actual positions of implants placed in a partially edentulous model using surgical templates with different sleeve designs. MATERIALS AND METHODS. Forty-nine mandibular partially edentulous models were scanned using a model scanner to create standard tessellation language (STL) files. The models were divided into seven groups based on sleeve design: Group 1-4 (Open non-metal sleeves with varying heights (3 mm, 6 mm) and buccal widths (5 mm, 6 mm)), Group 5-6 (Closed non-metal sleeves with heights of 3 mm and 6 mm), and Group 7 (Closed metal sleeve). After implant placement at #45 and #47, STL files were generated and analyzed using 3D measurement software. Deviations were measured in terms of horizontal deviation (coronal and apical), vertical deviation, and angular deviation. Statistical significance was evaluated using the Wilcoxon rank sum test, Kruskal-Wallis test, and multiple generalized linear models. RESULTS. While height differences showed no significant deviations (P > .05), there was a tendency for larger apical deviation with increased sleeve height (P < .1). Closed metal and closed non-metal sleeves showed smaller deviations compared to open non-metal sleeves, except in vertical deviation. CONCLUSION. Closed sleeves demonstrated greater accuracy in s-CAIS compared to open sleeves. Higher sleeves may improve implant placement accuracy, but buccal insertion width did not significantly affect accuracy.
引用
收藏
页码:22 / 35
页数:14
相关论文
共 50 条
  • [31] Accuracy of dynamic computer-assisted implant surgery in fully edentulous patients: An in vitro study
    Ruiz-Romero, Victor
    Jorba-Garcia, Adria
    Camps-Font, Octavi
    Figueiredo, Rui
    Valmaseda-Castellon, Eduard
    JOURNAL OF DENTISTRY, 2024, 149
  • [32] Registration accuracy in computer-assisted pelvic surgery
    Docquier, P. L.
    Paul, L.
    Cartiaux, O.
    Banse, X.
    COMPUTER AIDED SURGERY, 2009, 14 (1-3) : 37 - 44
  • [33] Comparison of Dental Surface Image Registration and Fiducial Marker Registration: An In Vivo Accuracy Study of Static Computer-Assisted Implant Surgery
    Han, Yen-Ting
    Lin, Wei-Chun
    Fan, Fang-Yu
    Chen, Chih-Long
    Lin, Chia-Cheng
    Cheng, Hsin-Chung
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (18)
  • [34] Accuracy and patient-centered results of static and dynamic computer-assisted implant surgery in edentulous jaws: a retrospective cohort study
    Fu, Shuang
    Sun, Wei
    Zhu, Jingxian
    Huang, Bin
    Ji, Wei
    Shi, Bin
    CLINICAL ORAL INVESTIGATIONS, 2023, 27 (09) : 5427 - 5438
  • [35] Accuracy and patient-centered results of static and dynamic computer-assisted implant surgery in edentulous jaws: a retrospective cohort study
    Shuang Fu
    Wei Sun
    Jingxian Zhu
    Bin Huang
    Wei Ji
    Bin Shi
    Clinical Oral Investigations, 2023, 27 : 5427 - 5438
  • [36] Bead-anchored surgical templates for static computer-assisted implant surgery: A dental technique
    Orgev, Ahmet
    Pellerito, John
    Polido, Waldemar
    Morton, Dean
    Lin, Wei-Shao
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (08): : 722 - 727
  • [37] Dynamic and static computer-assisted implant surgery for completely edentulous patients. A proof of a concept
    Pomares-Puig, Carmen
    Sanchez-Garces, M. Angeles
    Jorba-Garcia, Adria
    JOURNAL OF DENTISTRY, 2023, 130
  • [38] Seating accuracy of implant immediate provisional prostheses fabricated by digital workflow prior to implant placement by fully guided static computer-assisted implant surgery: An in vitro study
    Abduo, Jaafar
    Lau, Douglas
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (05) : 608 - 618
  • [39] Accuracy of computer-assisted, template-guided implant placement compared with conventional implant placement by hand-An in vitro study
    Schneider, David
    Sax, Caroline
    Sancho-Puchades, Manuel
    Hammerle, Christoph H. F.
    Jung, Ronald Ernst
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (09) : 1052 - 1060
  • [40] Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems: an in vitro study
    Zhou, Miao
    Zhou, Hui
    Li, Shu-Yi
    Zhu, Yi-Bo
    Geng, Yuan-Ming
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2021, 122 (04) : 343 - 348