Influence of template design on the accuracy of static computer-assisted implant surgery

被引:0
|
作者
Chung, Jinmook [1 ]
Park, Junghyun [1 ]
Son, Hyeonseong [1 ]
Hong, Seoung-Jin [2 ]
Pae, Ahran [2 ]
机构
[1] Kyung Hee Univ, Grad Sch, Dept Dent, Seoul, South Korea
[2] Kyung Hee Univ, Coll Dent, Dept Prosthodont, 26 Kyungheedae Ro, Seoul 02447, South Korea
来源
JOURNAL OF ADVANCED PROSTHODONTICS | 2025年 / 17卷 / 01期
关键词
Guided surgery; Implant; Open sleeve; Static computer-assisted implant surgery (s-CAIS); Surgical template; STEREOLITHOGRAPHIC SURGICAL GUIDES; PLACEMENT; TOMOGRAPHY; TOLERANCE; PRECISION; IMAGES;
D O I
10.4047/jap.2025.17.1.22
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This study aimed to evaluate the accuracy of static computer- assisted implant surgery (s-CAIS) by comparing the planned and actual positions of implants placed in a partially edentulous model using surgical templates with different sleeve designs. MATERIALS AND METHODS. Forty-nine mandibular partially edentulous models were scanned using a model scanner to create standard tessellation language (STL) files. The models were divided into seven groups based on sleeve design: Group 1-4 (Open non-metal sleeves with varying heights (3 mm, 6 mm) and buccal widths (5 mm, 6 mm)), Group 5-6 (Closed non-metal sleeves with heights of 3 mm and 6 mm), and Group 7 (Closed metal sleeve). After implant placement at #45 and #47, STL files were generated and analyzed using 3D measurement software. Deviations were measured in terms of horizontal deviation (coronal and apical), vertical deviation, and angular deviation. Statistical significance was evaluated using the Wilcoxon rank sum test, Kruskal-Wallis test, and multiple generalized linear models. RESULTS. While height differences showed no significant deviations (P > .05), there was a tendency for larger apical deviation with increased sleeve height (P < .1). Closed metal and closed non-metal sleeves showed smaller deviations compared to open non-metal sleeves, except in vertical deviation. CONCLUSION. Closed sleeves demonstrated greater accuracy in s-CAIS compared to open sleeves. Higher sleeves may improve implant placement accuracy, but buccal insertion width did not significantly affect accuracy.
引用
收藏
页码:22 / 35
页数:14
相关论文
共 50 条
  • [1] Influence of implant macrodesign and insertion connection technology on the accuracy of static computer-assisted implant surgery
    El Kholy, Karim
    Ebenezer, Supriya
    Wittneben, Julia-Gabriela
    Lazarin, Rafael
    Rousson, Dominique
    Buser, Daniel
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (05) : 1073 - 1079
  • [2] Influence of surgical guide support and implant site location on accuracy of static Computer-Assisted Implant Surgery
    El Kholy, Karim
    Lazarin, Rafael
    Janner, Simone F. M.
    Faerber, Karin
    Buser, Ramona
    Buser, Daniel
    CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (11) : 1067 - 1075
  • [3] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Anna Takács
    Gyula Marada
    Kinga Turzó
    Ákos Nagy
    Orsolya Németh
    Eitan Mijiritsky
    Márton Kivovics
    Attila Mühl
    BMC Oral Health, 23
  • [4] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Takacs, Anna
    Marada, Gyula
    Turzo, Kinga
    Nagy, Akos
    Nemeth, Orsolya
    Mijiritsky, Eitan
    Kivovics, Marton
    Muhl, Attila
    BMC ORAL HEALTH, 2023, 23 (01)
  • [5] The Impact of Surgical Guide Fixation and Implant Location on Accuracy of Static Computer-Assisted Implant Surgery
    Pessoa, Roberto
    Siqueira, Rafael
    Li, Junying
    Saleh, Islam
    Meneghetti, Priscila
    Bezerra, Fabio
    Wang, Hom-Lay
    Mendonca, Gustavo
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (02): : 155 - 164
  • [6] In Vivo and In Vitro Accuracy Analysis of Static Computer-Assisted Implant Surgery in an Edentulous Patient
    Makarov, Nikolay
    Papi, Piero
    Santomauro, Vincenzo
    Di Carlo, Gabriele
    Polimeni, Antonella
    Di Murro, Bianca
    Saccucci, Matteo
    APPLIED SCIENCES-BASEL, 2023, 13 (02):
  • [7] Accuracy of Different Surgical Guide Designs for Static Computer-Assisted Implant Surgery: An In Vitro Study
    Wu, Yu Tsung
    Papaspyridakos, Panos
    Kang, Kiho
    Finkelman, Matthew
    Kudara, Yukio
    De Souza, Andre B.
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (05) : 351 - 357
  • [8] Accuracy of dental implant placement using static versus dynamic computer-assisted implant surgery: An in vitro study
    Otaghsara, Seyedeh Sahar Taheri
    Joda, Tim
    Thieringer, Florian Markus
    JOURNAL OF DENTISTRY, 2023, 132
  • [9] The influence of guided sleeve height, drilling distance, and drilling key length on the accuracy of static Computer-Assisted Implant Surgery
    El Kholy, Karim
    Janner, Simone F. M.
    Schimmel, Martin
    Buser, Daniel
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (01) : 101 - 107
  • [10] Influence of alveolar ridge morphology and guide-hole design on the accuracy of static Computer-Assisted Implant Surgery with two implant macro-designs: An in vitro study
    Dulla, Fabrice Alain
    Couso-Queiruga, Emilio
    Chappuis, Vivianne
    Yilmaz, Burak
    Abou-Ayash, Samir
    Raabe, Clemens
    JOURNAL OF DENTISTRY, 2023, 130