Voltage Matrix Algorithm for Intraoperative Detection of Cochlear Implant Electrode Misplacement

被引:0
|
作者
Franke-Trieger, Annett [1 ]
Lailach, Susen [1 ]
Strahl, Stefan B. [2 ]
Murrmann, Katrin [1 ]
Schebsdat, Erik [3 ]
Neudert, Marcus [1 ]
Zahnert, Thomas [1 ]
机构
[1] Tech Univ Dresden, Dept Otorhinolaryngol Head & Neck Surg, Saxonian Cochlear Implant Ctr, Dresden, Germany
[2] MED EL Med Elect, Res & Dev, Innsbruck, Austria
[3] MED EL Elektromed Gerate Deutschland GmbH, Starnberg, Germany
关键词
Cochlear implant; Voltagematrix; Tip fold-over; Basal kinking; TIP FOLD-OVER; ARRAY; PANEL;
D O I
10.1159/000543264
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: To develop an algorithm, based on the voltage matrix, for detecting regular cochlear implant (CI) electrode position during the implantation procedure, tip fold-over or basal kinking for lateral-wall electrodes. The availability of an algorithm would be valuable in clinical routine, as incorrect positioning of the electrode array can potentially be recognized intraoperatively. Design: In this retrospective study, intraoperative voltage matrix and postoperative digital volume tomography of 525 CI recipients were analyzed. On the basis of these data an algorithm was developed for detecting various kinds of electrode misplacements. Results: Seven incorrect electrode positions, three tip fold (0.57%) and four basal kinking (0.76%) were detected. For detecting correct positioning, a sensitivity of 100%, a specificity of 83.3%, and a positive predictive value (PPV) of 99.8% were found. For detecting tip fold-over, a sensitivity of 100%, a specificity of 100%, and a PPV of 100% were found. For detecting basal kinking, a sensitivity of 66%, a specificity of 100%, and a PPV of 100% were determined. Conclusion: The algorithm was found to be an effective screening tool for detecting tip fold-over or basal kinking. (c) 2024 The Author(s). Published by S. Karger AG, Basel
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Cochlear implant electrode misplacement: Incidence, evaluation, and management
    Ying, Yu-Lan Mary
    Lin, Jerry W.
    Oghalai, John S.
    Williamson, Robert A.
    LARYNGOSCOPE, 2013, 123 (03): : 757 - 766
  • [2] Cochlear Implant Electrode Misplacement: A Case Series and Contemporary Review
    Shin, Timothy J.
    Totten, Douglas J.
    Tucker, Brady J.
    Nelson, Rick F.
    OTOLOGY & NEUROTOLOGY, 2022, 43 (05) : 547 - 558
  • [3] Cochlear implant electrode array misplacement: a cautionary case report
    Muzzi, E.
    Boscolo-Rizzo, P.
    Santarelli, R.
    Beltrame, M. A.
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2012, 128 (04): : 414 - 417
  • [4] Intraoperative Correction of Cochlear Implant Electrode Translocation
    Morrel, William G.
    Manzoor, Nauman F.
    Dawant, Benoit M.
    Noble, Jack H.
    Labadie, Robert F.
    AUDIOLOGY AND NEURO-OTOLOGY, 2022, 27 (02) : 104 - 108
  • [5] Cochlear implant electrode array tip-foldover detection by electrode voltage telemetry
    Leblans, Marc
    Zarowski, Andrzej
    Molisz, Andrzej
    van Dinther, Joost
    Dedeyne, Janne
    Lerut, Bob
    Kuhweide, Rudolf
    Offeciers, Erwin
    COCHLEAR IMPLANTS INTERNATIONAL, 2023, 24 (02) : 95 - 106
  • [6] A Rare Case of Cochlear Implant Electrode Array Misplacement Into the Posterior Semicircular Canal
    Zhao, Christopher H.
    Quimby, Alexandra E.
    Bigelow, Douglas C.
    LARYNGOSCOPE, 2023, 133 (01): : 175 - 177
  • [7] Evaluation of a Transimpedance Matrix Algorithm to Detect Anomalous Cochlear Implant Electrode Position
    Hoppe, Ulrich
    Brademann, Goetz
    Stover, Timo
    Ramos de Miguel, Angel
    Cowan, Robert
    Manrique, Manuel
    Falcon-Gonzalez, Juan Carlos
    Hey, Matthias
    Baumann, Uwe
    Huarte, Alicia
    Liebscher, Tim
    Bennett, Christopher
    English, Ruth
    Neben, Nicole
    Ramos Macias, Angel
    AUDIOLOGY AND NEURO-OTOLOGY, 2022, 27 (05) : 347 - 355
  • [8] Intraoperative Cochlear Implant Reinsertion Effects Evaluated by Electrode Impedance
    Toner, Fearghal M.
    Sanli, Halit
    Hall, Andrew C.
    Birman, Catherine S.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (06) : E695 - E699
  • [9] "Seeing" cochlear implant misplacement without imaging test
    Batuecas-Caletrio, Angel
    Ying, Yu-Lan Mary
    Gonzalez-Sanchez, Enrique
    Santacruz-Ruiz, Santiago
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2014, 35 (05) : 661 - 663
  • [10] COCHLEAR IMPLANT ELECTRODE
    CLARK, GM
    HALLWORTH, RJ
    ZDANIUS, K
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1975, 89 (08): : 787 - 792