Cross-calibration of areal bone mineral densities and body composition between DMS Stratos and Hologic Horizon A dual- energy X-ray absorptiometers: The effect of body mass index

被引:0
|
作者
Maimoun, Laurent [1 ,2 ]
Alonso, Sandrine [3 ]
Mahadea, Krishna Kunal [4 ]
Dubois, Julien [5 ,6 ]
Paunet, Tom [1 ]
Kucharczak, Florentin [1 ,3 ]
Nande, Lisa Maimoun [4 ]
Boudousq, Vincent [4 ,7 ]
Mura, Thibault [3 ,4 ]
Mariano-Goulart, Denis [1 ,2 ]
机构
[1] CHU Montpellier, Hop Lapeyronie, Serv Med Nucl, Montpellier, France
[2] Univ Montpellier UM, Physiol & Med Expt Coeur & Muscles PhyMedEx, INSERM, CNRS, Montpellier, France
[3] CHU Nimes, Serv BESPIM Biostat Epidemiol Clin Sante Publ & In, Nimes, France
[4] Univ Montpellier, Montpellier, France
[5] Montpellier Univ, Montpellier Univ Hosp, Dept Radiopharm, Montpellier, France
[6] Univ Montpellier, Canc Res Inst Montpellier IRCM, INSERM U1194, Montpellier, France
[7] CHU Nimes, Hop Caremeau, Serv Med Nucl, Nimes, France
关键词
Cross-calibration; Dual X-ray absorptiometry (DXA); Stratos; Horizon A; Areal bone mineral density (aBMD); Body composition; Body mass index; PENCIL-BEAM; TISSUE MEASUREMENTS; PRODIGY; IDXA; PRECISION; DENSITOMETERS; PARAMETERS; SPINE; FAT; HIP;
D O I
10.1016/j.jocd.2024.101553
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: The aim of this study was to investigate the correlations between areal bone mineral density (aBMD) and body composition measured by two dual-energy X-ray absorptiometers (DXA), the DMS Stratos (R) (STR) and the Hologic Horizon A (R) (HRZ), and then generate cross-calibration equations between the two scanners. Methods: Repeat scans were obtained from 251 adults (85 % female), 36 +/- 14 years old with mean body mass index (BMI) of 28.7 +/- 11.1 kg/m2, using HRZ (fan-beam technology) and STR (pencil-beam technology). aBMD was measured at whole body [WB], femoral neck [FN], total hip [TH], lumbar spine [LS] and radius, while fat mass [FM] and lean tissue mass [LTM] were determined at whole body and at android and gynoid subregions. Results: Compared to HRZ, STR underestimated both aBMD at WB and radius and LTM at WB and android and gynoid regions. Conversely, STR overestimated aBMD at FN, TH, LS and FM at WB android and gynoid regions. Except for WB bone mineral content (r = 0.87) and WB aBMD (r = 0.84), there were strong correlations of aBMD and body composition between the two DXAs (r > 0.91; p < 0.0001). Several of the parameters that required the determination of specific cross-calibration equations because of the significant bias between the two DXAs were found to be influenced by BMI. Conclusions: Although the data from the STR and the HRZ were highly correlated for aBMD and body composition parameters, a systematic measurement bias between two DXAs was observed. The development of cross-calibration equations fully corrected these differences and they may thus be useful for multicenter studies when scans are performed with STR and HRZ.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] The precision study of dual energy X-ray absorptiometry for bone mineral density and body composition measurements in female cynomolgus monkeys
    Guo, Bin
    Cai, Qijun
    Mai, Jinci
    Hou, Lu
    Zeng, Chunyuan
    Gan, Jiefeng
    Tan, Zhiqiang
    Li, Yuefeng
    Cheng, Yong
    Shang, Jingjie
    Tang, Yongjin
    Ling, Xueying
    Gong, Jian
    Wang, Lu
    Xu, Hao
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (03) : 2051 - 2057
  • [42] DUAL ENERGY X-RAY ABSORPTIOMETRY ASSESSMENT OF TOTAL BONE-MINERAL CONTENT AND BODY-COMPOSITION IN NORMAL NEWBORNS
    BRAILLON, PM
    SALLE, BL
    LAPILLONNE, A
    CHAMBON, M
    CAVERO, E
    DELMAS, PD
    MEUNIER, PJ
    JOURNAL OF BONE AND MINERAL RESEARCH, 1992, 7 : S264 - S264
  • [43] Anomalies in dual energy X-ray absorptiometry measurements of total-body bone mineral during weight change using Lunar, Hologic and Norland instruments
    Tothill, P
    Laskey, MA
    Orphanidou, CI
    van Wijk, M
    BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (859): : 661 - 669
  • [44] Changes in bone mineral, lean body mass and eat content as measured by dual energy X-ray absorptiometry: A longitudinal study
    Karlsson, MK
    Obrant, KJ
    Nilsson, BE
    Johnell, O
    CALCIFIED TISSUE INTERNATIONAL, 2000, 66 (02) : 97 - 99
  • [45] Changes in Bone Mineral, Lean Body Mass and Fat Content as Measured by Dual Energy X-ray Absorptiometry: A Longitudinal Study
    M. K. Karlsson
    K. J. Obrant
    B. E. Nilsson
    O. Johnell
    Calcified Tissue International, 2000, 66 : 97 - 99
  • [46] Correlation of Body Mass Index to Dual-Energy X-Ray Absorptiometry in Assessment of Body Composition during Therapy for Childhood Acute Lymphoblastic Leukemia
    Orgel, Etan
    Mueske, Nicole
    Freyer, David Robert
    Mittelman, Steven D.
    BLOOD, 2014, 124 (21)
  • [47] VALIDATION AND APPLICATION OF DUAL-ENERGY X-RAY ABSORPTIOMETRY TO MEASURE BONE MASS AND BODY-COMPOSITION IN SMALL INFANTS
    BRUNTON, JA
    BAYLEY, HS
    ATKINSON, SA
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1993, 58 (06): : 839 - 845
  • [48] A study on bone mass density using dual energy X-ray absorptiometry: Does high body mass index have protective effect on bone density in obese patients?
    Shayganfar, Azin
    Ebrahimian, Shadi
    Masjedi, Mahsa
    Daryaei, Sadaf
    JOURNAL OF RESEARCH IN MEDICAL SCIENCES, 2020, 25
  • [49] Bone mineral density and body composition in boys with distal. forearm fractures: A dual-energy x-ray absorptiometry study
    Goulding, A
    Jones, IE
    Taylor, RW
    Williams, SM
    Manning, PJ
    JOURNAL OF PEDIATRICS, 2001, 139 (04): : 509 - 515
  • [50] DUAL-ENERGY X-RAY ABSORPTIOMETRY FOR TOTAL-BODY AND REGIONAL BONE-MINERAL AND SOFT-TISSUE COMPOSITION
    MAZESS, RB
    BARDEN, HS
    BISEK, JP
    HANSON, J
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1990, 51 (06): : 1106 - 1112