Cross-calibration of areal bone mineral densities and body composition between DMS Stratos and Hologic Horizon A dual- energy X-ray absorptiometers: The effect of body mass index

被引:0
|
作者
Maimoun, Laurent [1 ,2 ]
Alonso, Sandrine [3 ]
Mahadea, Krishna Kunal [4 ]
Dubois, Julien [5 ,6 ]
Paunet, Tom [1 ]
Kucharczak, Florentin [1 ,3 ]
Nande, Lisa Maimoun [4 ]
Boudousq, Vincent [4 ,7 ]
Mura, Thibault [3 ,4 ]
Mariano-Goulart, Denis [1 ,2 ]
机构
[1] CHU Montpellier, Hop Lapeyronie, Serv Med Nucl, Montpellier, France
[2] Univ Montpellier UM, Physiol & Med Expt Coeur & Muscles PhyMedEx, INSERM, CNRS, Montpellier, France
[3] CHU Nimes, Serv BESPIM Biostat Epidemiol Clin Sante Publ & In, Nimes, France
[4] Univ Montpellier, Montpellier, France
[5] Montpellier Univ, Montpellier Univ Hosp, Dept Radiopharm, Montpellier, France
[6] Univ Montpellier, Canc Res Inst Montpellier IRCM, INSERM U1194, Montpellier, France
[7] CHU Nimes, Hop Caremeau, Serv Med Nucl, Nimes, France
关键词
Cross-calibration; Dual X-ray absorptiometry (DXA); Stratos; Horizon A; Areal bone mineral density (aBMD); Body composition; Body mass index; PENCIL-BEAM; TISSUE MEASUREMENTS; PRODIGY; IDXA; PRECISION; DENSITOMETERS; PARAMETERS; SPINE; FAT; HIP;
D O I
10.1016/j.jocd.2024.101553
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: The aim of this study was to investigate the correlations between areal bone mineral density (aBMD) and body composition measured by two dual-energy X-ray absorptiometers (DXA), the DMS Stratos (R) (STR) and the Hologic Horizon A (R) (HRZ), and then generate cross-calibration equations between the two scanners. Methods: Repeat scans were obtained from 251 adults (85 % female), 36 +/- 14 years old with mean body mass index (BMI) of 28.7 +/- 11.1 kg/m2, using HRZ (fan-beam technology) and STR (pencil-beam technology). aBMD was measured at whole body [WB], femoral neck [FN], total hip [TH], lumbar spine [LS] and radius, while fat mass [FM] and lean tissue mass [LTM] were determined at whole body and at android and gynoid subregions. Results: Compared to HRZ, STR underestimated both aBMD at WB and radius and LTM at WB and android and gynoid regions. Conversely, STR overestimated aBMD at FN, TH, LS and FM at WB android and gynoid regions. Except for WB bone mineral content (r = 0.87) and WB aBMD (r = 0.84), there were strong correlations of aBMD and body composition between the two DXAs (r > 0.91; p < 0.0001). Several of the parameters that required the determination of specific cross-calibration equations because of the significant bias between the two DXAs were found to be influenced by BMI. Conclusions: Although the data from the STR and the HRZ were highly correlated for aBMD and body composition parameters, a systematic measurement bias between two DXAs was observed. The development of cross-calibration equations fully corrected these differences and they may thus be useful for multicenter studies when scans are performed with STR and HRZ.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Body composition and bone mineral density by Dual Energy X-ray Absorptiometry: Reference values for young children
    van Beijsterveldt, Inge A. L. P.
    van der Steen, Manouk
    de Fluiter, Kirsten S.
    Spaans, Sander A. M. J.
    Hokken-Koelega, Anita C. S.
    CLINICAL NUTRITION, 2022, 41 (01) : 71 - 79
  • [22] Comparisons of regional body composition and bone mineral from whole body dual energy X-ray absorptiometry across four scanners
    Soriano, JMP
    Ioannidou, E
    Wang, J
    Thornton, JC
    Horlick, M
    Gallagher, D
    Pierson, RN
    Heymsfield, SB
    FASEB JOURNAL, 2004, 18 (04): : A551 - A551
  • [23] COMPARISONS BETWEEN HOLOGIC, LUNAR AND NORLAND DUAL-ENERGY X-RAY ABSORPTIOMETERS AND OTHER TECHNIQUES USED FOR WHOLE-BODY SOFT-TISSUE MEASUREMENTS
    TOTHILL, P
    AVENELL, A
    LOVE, J
    REID, DM
    EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1994, 48 (11) : 781 - 794
  • [24] Relationship between Weight, Body Mass Index, and Bone Mineral Density in Men Referred for Dual-Energy X-Ray Absorptiometry Scan in Isfahan, Iran
    Salamat, Mohammad Reza
    Salamat, Amir Hossein
    Abedi, Iraj
    Janghorbani, Mohsen
    JOURNAL OF OSTEOPOROSIS, 2013, 2013
  • [25] Dual-energy x-ray absorptiometry derived body composition trajectories across adulthood: Reference values and associations with body roundness index and body mass index
    Pratt, Jedd
    Narici, Marco
    Boreham, Colin
    De Vito, Giuseppe
    CLINICAL NUTRITION, 2025, 46 : 137 - 146
  • [26] The effect of Tc-99m on Dual-energy X-ray absorptiometry measurement of body composition and bone mineral density
    Fosbol, M.
    Dupont, A.
    Zerahn, B.
    Alslev, L.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S231 - S231
  • [27] Effects of metal implants on whole-body dual-energy x-ray absorptiometry measurements of bone mineral content and body composition
    Giangregorio, LM
    Webber, CE
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2003, 54 (05): : 305 - 309
  • [28] Whole body-dual energy X-ray absorptiometry to test body composition and bone mineral density in children with mucopolysaccharidosis type IV
    Rigante, D
    Segni, G
    Caradonna, P
    BONE, 2001, 28 (05) : S123 - S124
  • [29] The use of dual-energy X-ray absorptiometry in evaluating bone mineral density and body composition of preterm neonates
    Ng, SCY
    Liew, WK
    Lee, YS
    Joseph, R
    Low, SL
    Ooi, PL
    Lee, EH
    PEDIATRIC RESEARCH, 2002, 52 (05) : 785 - 785
  • [30] Dual-energy X-ray absorptiometry assessment of the bone mineral content and body composition changes in overweight youths
    Braillon, PM
    Serban, A
    RADIOLOGY, 2000, 217 : 341 - 341