Cross-calibration of areal bone mineral densities and body composition between DMS Stratos and Hologic Horizon A dual- energy X-ray absorptiometers: The effect of body mass index

被引:0
|
作者
Maimoun, Laurent [1 ,2 ]
Alonso, Sandrine [3 ]
Mahadea, Krishna Kunal [4 ]
Dubois, Julien [5 ,6 ]
Paunet, Tom [1 ]
Kucharczak, Florentin [1 ,3 ]
Nande, Lisa Maimoun [4 ]
Boudousq, Vincent [4 ,7 ]
Mura, Thibault [3 ,4 ]
Mariano-Goulart, Denis [1 ,2 ]
机构
[1] CHU Montpellier, Hop Lapeyronie, Serv Med Nucl, Montpellier, France
[2] Univ Montpellier UM, Physiol & Med Expt Coeur & Muscles PhyMedEx, INSERM, CNRS, Montpellier, France
[3] CHU Nimes, Serv BESPIM Biostat Epidemiol Clin Sante Publ & In, Nimes, France
[4] Univ Montpellier, Montpellier, France
[5] Montpellier Univ, Montpellier Univ Hosp, Dept Radiopharm, Montpellier, France
[6] Univ Montpellier, Canc Res Inst Montpellier IRCM, INSERM U1194, Montpellier, France
[7] CHU Nimes, Hop Caremeau, Serv Med Nucl, Nimes, France
关键词
Cross-calibration; Dual X-ray absorptiometry (DXA); Stratos; Horizon A; Areal bone mineral density (aBMD); Body composition; Body mass index; PENCIL-BEAM; TISSUE MEASUREMENTS; PRODIGY; IDXA; PRECISION; DENSITOMETERS; PARAMETERS; SPINE; FAT; HIP;
D O I
10.1016/j.jocd.2024.101553
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: The aim of this study was to investigate the correlations between areal bone mineral density (aBMD) and body composition measured by two dual-energy X-ray absorptiometers (DXA), the DMS Stratos (R) (STR) and the Hologic Horizon A (R) (HRZ), and then generate cross-calibration equations between the two scanners. Methods: Repeat scans were obtained from 251 adults (85 % female), 36 +/- 14 years old with mean body mass index (BMI) of 28.7 +/- 11.1 kg/m2, using HRZ (fan-beam technology) and STR (pencil-beam technology). aBMD was measured at whole body [WB], femoral neck [FN], total hip [TH], lumbar spine [LS] and radius, while fat mass [FM] and lean tissue mass [LTM] were determined at whole body and at android and gynoid subregions. Results: Compared to HRZ, STR underestimated both aBMD at WB and radius and LTM at WB and android and gynoid regions. Conversely, STR overestimated aBMD at FN, TH, LS and FM at WB android and gynoid regions. Except for WB bone mineral content (r = 0.87) and WB aBMD (r = 0.84), there were strong correlations of aBMD and body composition between the two DXAs (r > 0.91; p < 0.0001). Several of the parameters that required the determination of specific cross-calibration equations because of the significant bias between the two DXAs were found to be influenced by BMI. Conclusions: Although the data from the STR and the HRZ were highly correlated for aBMD and body composition parameters, a systematic measurement bias between two DXAs was observed. The development of cross-calibration equations fully corrected these differences and they may thus be useful for multicenter studies when scans are performed with STR and HRZ.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Longitudinal reproducibility for measurement of body composition and bone mineral by a fan-beam dual energy x-ray absorptiometry
    Wang, J
    Thornton, J
    Ioannidou, E
    Gallagher, D
    Horlick, M
    Heymsfield, S
    Pierson, R
    Allen, L
    OBESITY RESEARCH, 2004, 12 : A227 - A227
  • [32] Does paracentesis of ascites influence measurements of bone mineral or body composition by dual-energy x-ray absorptiometry?
    Haderslev, KV
    Svendsen, OL
    Staun, M
    METABOLISM-CLINICAL AND EXPERIMENTAL, 1999, 48 (03): : 373 - 377
  • [33] Comparisons between a pencil beam and two fan beam dual energy X-ray absorptiometers used for measuring total body bone and soft tissue
    Tothill, P
    Hannan, WJ
    Wilkinson, S
    BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (878): : 166 - 176
  • [34] The Evaluation of Consistency Between Body Composition Assessments in Pediatric Population Using Pencil Beam and Fan Beam Dual-Energy X-Ray Absorptiometers
    Pludowski, Pawel
    Jaworski, Maciej
    Matusik, Halina
    Kobylinska, Maria
    Klimek, Paulina
    Lorenc, Roman S.
    JOURNAL OF CLINICAL DENSITOMETRY, 2010, 13 (01) : 84 - 95
  • [35] Cross-calibration of fat and lean measurements by dual-energy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range
    Pintauro, SJ
    Nagy, TR
    Duthie, CM
    Goran, MI
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1996, 63 (03): : 293 - 298
  • [36] THE ASSOCIATION OF BODY MASS INDEX AND BONE MINERAL DENSITY IN A MALE POPULATION REFERRED FOR DUAL-ENERGY X-RAY ABSORPTIOMETRY SCAN IN ISFAHAN, IRAN
    Salamat, Mohammad Reza
    Salamat, Amir Hossein
    Kompani, Khadijeh
    OSTEOPOROSIS INTERNATIONAL, 2012, 23 : S94 - S95
  • [37] Effect of 18F-FDG Administration on Measurements of Bone Mineral Density and Body Composition by Dual-Energy X-ray Absorptiometry
    Kim, Dae-Weung
    Kim, Woo Hyoung
    Kim, Myoung Hyoun
    Kim, Seong Su
    Mo, Eun Hee
    Lee, Chun Ho
    Kim, Chang Guhn
    JOURNAL OF CLINICAL DENSITOMETRY, 2013, 16 (03) : 283 - 286
  • [38] Validation of bone mass and body composition measurements in small subjects with pencil beam dual energy X-ray absorptiometry
    Koo, WWK
    Hammami, M
    Hockman, EM
    JOURNAL OF THE AMERICAN COLLEGE OF NUTRITION, 2004, 23 (01) : 79 - 84
  • [39] Quantitative Comparison of 2 Dual-Energy X-ray Absorptiometry Systems in Assessing Body Composition and Bone Mineral Measurements
    Xu, Wenhua
    Chafi, Hatim
    Guo, Beibei
    Heymsfield, Steven B.
    Murray, Kori B.
    Zheng, Jolene
    Jia, Guang
    JOURNAL OF CLINICAL DENSITOMETRY, 2016, 19 (03) : 298 - 304
  • [40] ANALYSIS OF CHANGES IN BODY COMPOSITION AND BONE MINERAL DENSITY IN CHILDREN RECEIVING CHEMOTHERAPY ANTIBLASTIC BY X-RAY ABSORPTIOMETRY IN DUAL ENERGY
    Verlicchi, L.
    Masetti, K.
    Gobbi, E.
    Martoni, A.
    Biagi, C.
    Gaspenni, P.
    Bazzocchi, A.
    Rondelli, R.
    Melchionda, F.
    Klemschmidt, K.
    Prete, A.
    Canini, R.
    Pession, A.
    HAEMATOLOGICA-THE HEMATOLOGY JOURNAL, 2010, 95 (07): : S88 - S89