Color Stability of Bulk-Fill Flowable Resin Composites After Artificial Aging

被引:0
|
作者
Floriani, Franciele [1 ]
Jurado, Carlos A. [2 ]
Madhu, Nikkita [1 ]
Lackey, Mark A. [2 ]
Azpiazu-Flores, Francisco X. [3 ]
Lopes, Guilherme Carpena [4 ]
机构
[1] Univ Iowa, Coll Dent, Dept Prosthodont, Iowa City, IA 52241 USA
[2] Univ Tennessee Hlth Sci Ctr, Coll Dent, Dept Gen Dent, Memphis, TN 38103 USA
[3] Ohio State Univ, Coll Dent, Div Restorat & Prosthet Dent, Columbus, OH 43210 USA
[4] Univ Fed Santa Catarina, Dept Operat Dent, Florianopolis, SC, Brazil
关键词
resin composite; bulk-fill; color; flowable composite; shade; CLINICAL-PERFORMANCE; RESTORATIONS; VISCOSITY; PROFILE; TRIAL;
D O I
10.3390/dj12110350
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: This study aimed to evaluate the color stability of bulk-fill flowable resin composites with 2 difference shades at baseline and after artificial aging. Methods: Disk-shaped specimens (& Oslash;10 x 4 mm) were fabricated from three bulk-fill flowable resin resin composites (Filtek Bulk-Fill Flow, Venus Bulk-Fill Flow, and Estelite Bulk-Fill Flow). The specimens in each bulk-fill resin composite group were divided into two subgroups (n = 10 per subgroup) with two different shades, A1 (N = 30) and A3 (N = 30), and were polymerized with light curing (800 mW/cm(2)/Valo LED Unit, Ultradent) and polished. The color difference between bulk-fill resin composites was evaluated at baseline and after artificial aging using a spectrophotometer (CM-700d, Konica Minolta, Tokyo, Japan) under D65 illumination. Color coordinates were measured with CIEDE2000, and color differences (triangle E-00) and relative translucency parameter (RTP) values were calculated. Subsequently, the comparison of color changes (triangle E-00) before and after thermocycling was performed using the t-test for paired samples. Results: The bulk-fill flow resin composites evaluated in the present study were capable of mimicking important optical properties such as light transmission. All the resin composites provided acceptable color stability at baseline and after thermocycling when the color A1 was used. On the other hand, whenever the shade A3 was used, the Venus Bulk-Fill Flow demonstrated the best optical properties. There was no statistically significant difference when comparing baseline and after thermocycling in bulk-fill flowable resin composites (p > 0.05). After thermocycling, A1 bulk-fill flowable resin composites provided acceptable color stability, and all A3 bulk-fill flowable resin composites provided visible color change, except for the Venus Bulk-Fill Flow (triangle E-00 = 2.35). Conclusions: Estelite Bulk-Fill Flow displayed the best color stability (triangle E-00 = 2.22) between all the combinations evaluated.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Effects of flowable liners on the shrinkage vectors of bulk-fill composites
    Kaisarly, Dalia
    Meierhofer, D.
    El Gezawi, M.
    Roesch, P.
    Kunzelmann, K. H.
    CLINICAL ORAL INVESTIGATIONS, 2021, 25 (08) : 4927 - 4940
  • [22] Effects of Fibers on Color and Translucency Changes of Bulk-Fill and Anterior Composites after Accelerated Aging
    Tuncdemir, Ali Riza
    Guven, Mehmet Esad
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [23] The effects of different repolishing procedures on the color change of bulk-fill resin composites
    Fidan, Muhammet
    EUROPEAN ORAL RESEARCH, 2024, 58 (01) : 14 - 21
  • [24] Effect of Polishing Systems on the Surface Roughness and Color Stability of Aged and Stained Bulk-Fill Resin Composites
    Aktu, Aysem
    Ulusoy, Nuran
    MATERIALS, 2024, 17 (14)
  • [25] Color of bulk-fill composite resin restorative materials
    Barutcigil, Cagatay
    Barutcigil, Kubilay
    Ozarslan, Mehmet Mustafa
    Dundar, Ayse
    Yilmaz, Burak
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2018, 30 (02) : E3 - E8
  • [26] Influence of resin matrix on the rheology, translucency, and curing potential of experimental flowable composites for bulk-fill applications
    Eiras Pereira, Leonardo Duarte
    Couto Neto, Mario Pereira
    Pereira, Roberto Guimaraes
    Schneider, Luis Felipe J.
    DENTAL MATERIALS, 2021, 37 (06) : 1046 - 1053
  • [27] Flowable and Regular Bulk-Fill Composites: A Comprehensive Report on Restorative Treatment
    Sebold, Maicon
    Esteves Lins, Rodrigo Barros
    Andre, Carolina Bosso
    Marcondes Martins, Luis Roberto
    Giannini, Marcelo
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2020, 40 (02) : 293 - 300
  • [28] Genotoxic potential of dental bulk-fill resin composites
    Tauboeck, Tobias T.
    Marovic, Danijela
    Zeljezic, Davor
    Steingruber, Andrea D.
    Attin, Thomas
    Tarle, Zrinka
    DENTAL MATERIALS, 2017, 33 (07) : 788 - 795
  • [29] Evaluation of Radiopacity of Bulk-fill Flowable Composites Using Digital Radiography
    Tarcin, B.
    Gumru, B.
    Peker, S.
    Ovecoglu, H. S.
    OPERATIVE DENTISTRY, 2016, 41 (04) : 424 - 431
  • [30] Varying the Polishing Protocol Influences the Color Stability and Surface Roughness of Bulk-Fill Resin-Based Composites
    Freitas, Filipa
    de Melo, Teresa Pinheiro
    Delgado, Antonio H. S.
    Monteiro, Paulo
    Rua, Joao
    Proenca, Luis
    Caldeira, Jorge
    Azul, Ana Mano
    Mendes, Jose Joao
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2021, 12 (01)