Accuracy validation and user performance analysis of two new self-monitoring blood glucose systems

被引:0
|
作者
Huang, Tun-Yu [1 ]
Chen, Yu-Hsi [2 ]
Li, Yu-Ching [3 ]
Chiu, Yen-Hsin [4 ]
Kang, Hui-Jun [4 ]
Chen, Chih-Yu [5 ,6 ]
机构
[1] Show Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
[2] Show Chwan Mem Hosp, Dept Orthoped, Changhua, Taiwan
[3] Show Chwan Mem Hosp, Dept Family Med, Changhua, Taiwan
[4] China Med Univ, Dept Hlth Serv Adm, 100 Sect 1,Jingmao Rd, Taichung 406040, Taiwan
[5] Taipei Med Univ, Shuang Ho Hosp, Dept Orthoped, Taipei, Taiwan
[6] Taipei Med Univ, Coll Biomed Engn, Int Ph D Program Biomed Engn, Taipei, Taiwan
来源
ENDOCRINOLOGY DIABETES AND METABOLISM CASE REPORTS | 2025年 / 2025卷 / 01期
关键词
diabetes; self-glucose detection (monitoring) systems; user performance;
D O I
10.1530/EDM-24-0134
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with diabetes require regular blood glucose monitoring. We evaluated the precision and user performance of two blood glucose monitoring systems, the accuracy results of GlucoTeq BGM200 and DiaRite BGM300, adhering to the EN_ISO 15197:2015 standards, which require blood glucose >= 100 mg/dL, >95% of the data within +/- 15% difference, and when blood glucose <100 mg/dL, >95% of the data within +/- 15 mg/dL. This study assessed their conformity with the YSI analyzer and other leading systems. Participants (n = 101), 18 years of age were included, covering a diverse demographic. Accuracy was determined using the YSI analyzer as the standard, employing linear regression, a consensus error grid and Bland-Altman analyses. The performance of both system users showed 100% conformity within zone A of the consensus error grid, high linear regression coefficients and minimal bias in the Bland-Altman analysis. Subjective satisfaction analyses showed average scores of 4.59 and 4.62 for BGM200 and BGM300, respectively. Comparison with other systems revealed high regression coefficients (CONTOUR (R) PLUS; R-2 = 0.9960, BGM200; R-2 = 0.9927, BGM300; R-2 = 0.9915, Accu-Chek (R) Guide; R-2 = 0.9910). Both systems demonstrated reasonable accuracy and user performance comparable to competitors' products and met international standards. Their reliability in real-world scenarios and high user satisfaction make them valuable tools for diabetes management.
引用
收藏
页数:5
相关论文
共 50 条
  • [41] SELF-MONITORING OF BLOOD-GLUCOSE
    MOSS, JM
    DELAWTER, DE
    AMERICAN FAMILY PHYSICIAN, 1986, 33 (02) : 225 - 228
  • [42] Overview on self-monitoring of blood glucose
    Montagnana, Martina
    Caputo, Marco
    Giavarina, Davide
    Lippi, Giuseppe
    CLINICA CHIMICA ACTA, 2009, 402 (1-2) : 7 - 13
  • [43] SELF-MONITORING OF BLOOD-GLUCOSE
    MOLITCH, ME
    BARR, J
    CALLAHAN, PL
    CAMPBELL, RK
    DELAHANTY, LM
    RIZZA, R
    TOBIN, CT
    YOUNG, DS
    DIABETES CARE, 1994, 17 (01) : 81 - 86
  • [44] The value of self-monitoring of blood glucose
    Court, M
    DIABETES CARE, 2002, 25 (07) : 1255 - 1256
  • [45] SELF-MONITORING OF BLOOD-GLUCOSE
    SKYLER, JS
    MEDICAL CLINICS OF NORTH AMERICA, 1982, 66 (06) : 1227 - 1250
  • [46] SELF-MONITORING OF BLOOD-GLUCOSE
    不详
    DIABETES, 1985, 34 (09) : 945 - 945
  • [47] Routine self-monitoring of blood glucose
    Allan, G. Michael
    Korownyk, Christina
    Ivers, Noah
    CANADIAN FAMILY PHYSICIAN, 2011, 57 (09) : 1015 - 1015
  • [48] ACCURACY OF DEVICES FOR SELF-MONITORING OF BLOOD-GLUCOSE INCLUDING HYPOGLYCEMIC BLOOD-GLUCOSE LEVELS
    ZENOBI, PD
    KELLER, A
    JAEGGIGROISMAN, SE
    GLATZ, Y
    DIABETES CARE, 1995, 18 (04) : 587 - 588
  • [49] Clinical Evaluation of a New System for Self-Monitoring of Blood Glucose
    Randoll, Randee
    Ellison, John
    DIABETES, 2011, 60 : A236 - A237
  • [50] Proposing a new design for self-monitoring blood glucose logs
    Raz, Itamar
    Zary, Raviv
    Ben Dayan, Eliad
    DIABETES-METABOLISM RESEARCH AND REVIEWS, 2016, 32 (01) : 60 - 62