Accuracy validation and user performance analysis of two new self-monitoring blood glucose systems

被引:0
|
作者
Huang, Tun-Yu [1 ]
Chen, Yu-Hsi [2 ]
Li, Yu-Ching [3 ]
Chiu, Yen-Hsin [4 ]
Kang, Hui-Jun [4 ]
Chen, Chih-Yu [5 ,6 ]
机构
[1] Show Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
[2] Show Chwan Mem Hosp, Dept Orthoped, Changhua, Taiwan
[3] Show Chwan Mem Hosp, Dept Family Med, Changhua, Taiwan
[4] China Med Univ, Dept Hlth Serv Adm, 100 Sect 1,Jingmao Rd, Taichung 406040, Taiwan
[5] Taipei Med Univ, Shuang Ho Hosp, Dept Orthoped, Taipei, Taiwan
[6] Taipei Med Univ, Coll Biomed Engn, Int Ph D Program Biomed Engn, Taipei, Taiwan
来源
ENDOCRINOLOGY DIABETES AND METABOLISM CASE REPORTS | 2025年 / 2025卷 / 01期
关键词
diabetes; self-glucose detection (monitoring) systems; user performance;
D O I
10.1530/EDM-24-0134
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with diabetes require regular blood glucose monitoring. We evaluated the precision and user performance of two blood glucose monitoring systems, the accuracy results of GlucoTeq BGM200 and DiaRite BGM300, adhering to the EN_ISO 15197:2015 standards, which require blood glucose >= 100 mg/dL, >95% of the data within +/- 15% difference, and when blood glucose <100 mg/dL, >95% of the data within +/- 15 mg/dL. This study assessed their conformity with the YSI analyzer and other leading systems. Participants (n = 101), 18 years of age were included, covering a diverse demographic. Accuracy was determined using the YSI analyzer as the standard, employing linear regression, a consensus error grid and Bland-Altman analyses. The performance of both system users showed 100% conformity within zone A of the consensus error grid, high linear regression coefficients and minimal bias in the Bland-Altman analysis. Subjective satisfaction analyses showed average scores of 4.59 and 4.62 for BGM200 and BGM300, respectively. Comparison with other systems revealed high regression coefficients (CONTOUR (R) PLUS; R-2 = 0.9960, BGM200; R-2 = 0.9927, BGM300; R-2 = 0.9915, Accu-Chek (R) Guide; R-2 = 0.9910). Both systems demonstrated reasonable accuracy and user performance comparable to competitors' products and met international standards. Their reliability in real-world scenarios and high user satisfaction make them valuable tools for diabetes management.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Self-Monitoring of Blood Glucose
    Garg, Satish K.
    Hirsch, Irl B.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : S3 - S9
  • [22] Effect of Ambient Temperature on Analytical Performance of Self-Monitoring Blood Glucose Systems
    Nerhus, Kari
    Rustad, Pal
    Sandberg, Sverre
    DIABETES TECHNOLOGY & THERAPEUTICS, 2011, 13 (09) : 883 - 892
  • [23] Self-Monitoring of the Blood Glucose - New Guidelines and Studies
    Schnell, O.
    DIABETES STOFFWECHSEL UND HERZ, 2010, 19 (01): : 43 - 44
  • [24] New optical method for blood glucose self-monitoring
    Herbrechtsmeier, P.
    Mueller, A. J.
    Hasslacher, C.
    Auffarth, G. U.
    DIABETOLOGIA, 2009, 52 : S367 - S367
  • [25] Impact of Partial Pressure of Oxygen in Blood Samples on the Performance of Systems for Self-Monitoring of Blood Glucose
    Schmid, Christina
    Baumstark, Annette
    Pleus, Stefan
    Haug, Cornelia
    Tesar, Martina
    Freckmann, Guido
    DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 (03) : 156 - 165
  • [26] BEHAVIORAL SKILLS IN SELF-MONITORING OF BLOOD-GLUCOSE - RELATIONSHIP TO ACCURACY
    WING, RR
    KOESKE, R
    NEW, A
    LAMPARSKI, D
    BECKER, D
    DIABETES CARE, 1986, 9 (04) : 330 - 333
  • [27] Accuracy, reliability and human factors associated with self-monitoring of blood glucose
    Davis, E
    Latman, NS
    FASEB JOURNAL, 2005, 19 (05): : A1530 - A1531
  • [28] VALUE OF IMPROVED ACCURACY FOR SELF-MONITORING OF BLOOD GLUCOSE DEVICES IN GERMANY
    McQueen, R. B.
    Breton, M.
    Ott, M.
    Campbell, J. D.
    VALUE IN HEALTH, 2016, 19 (07) : A696 - A696
  • [29] Accuracy of Devices for Self-Monitoring of Blood Glucose: a Stochastic Error Model
    Vettoretti, M.
    Facchinetti, A.
    Sparacino, G.
    Cobelli, C.
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 2359 - 2362
  • [30] An accuracy calibration procedure for self-monitoring of blood glucose systems using healthy venous whole blood.
    Kuwa, K
    Nakayama, T
    Hoshino, T
    Tominaga, M
    CLINICAL CHEMISTRY, 2002, 48 (06) : A197 - A197