Accuracy validation and user performance analysis of two new self-monitoring blood glucose systems
被引:0
|
作者:
Huang, Tun-Yu
论文数: 0引用数: 0
h-index: 0
机构:
Show Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Huang, Tun-Yu
[1
]
Chen, Yu-Hsi
论文数: 0引用数: 0
h-index: 0
机构:
Show Chwan Mem Hosp, Dept Orthoped, Changhua, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Chen, Yu-Hsi
[2
]
Li, Yu-Ching
论文数: 0引用数: 0
h-index: 0
机构:
Show Chwan Mem Hosp, Dept Family Med, Changhua, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Li, Yu-Ching
[3
]
Chiu, Yen-Hsin
论文数: 0引用数: 0
h-index: 0
机构:
China Med Univ, Dept Hlth Serv Adm, 100 Sect 1,Jingmao Rd, Taichung 406040, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Chiu, Yen-Hsin
[4
]
Kang, Hui-Jun
论文数: 0引用数: 0
h-index: 0
机构:
China Med Univ, Dept Hlth Serv Adm, 100 Sect 1,Jingmao Rd, Taichung 406040, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Kang, Hui-Jun
[4
]
Chen, Chih-Yu
论文数: 0引用数: 0
h-index: 0
机构:
Taipei Med Univ, Shuang Ho Hosp, Dept Orthoped, Taipei, Taiwan
Taipei Med Univ, Coll Biomed Engn, Int Ph D Program Biomed Engn, Taipei, TaiwanShow Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
Chen, Chih-Yu
[5
,6
]
机构:
[1] Show Chwan Mem Hosp, Prospective Wound Med Res Ctr, Changhua 500, Taiwan
[2] Show Chwan Mem Hosp, Dept Orthoped, Changhua, Taiwan
[3] Show Chwan Mem Hosp, Dept Family Med, Changhua, Taiwan
[4] China Med Univ, Dept Hlth Serv Adm, 100 Sect 1,Jingmao Rd, Taichung 406040, Taiwan
[5] Taipei Med Univ, Shuang Ho Hosp, Dept Orthoped, Taipei, Taiwan
[6] Taipei Med Univ, Coll Biomed Engn, Int Ph D Program Biomed Engn, Taipei, Taiwan
来源:
ENDOCRINOLOGY DIABETES AND METABOLISM CASE REPORTS
|
2025年
/
2025卷
/
01期
关键词:
diabetes;
self-glucose detection (monitoring) systems;
user performance;
D O I:
10.1530/EDM-24-0134
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
Patients with diabetes require regular blood glucose monitoring. We evaluated the precision and user performance of two blood glucose monitoring systems, the accuracy results of GlucoTeq BGM200 and DiaRite BGM300, adhering to the EN_ISO 15197:2015 standards, which require blood glucose >= 100 mg/dL, >95% of the data within +/- 15% difference, and when blood glucose <100 mg/dL, >95% of the data within +/- 15 mg/dL. This study assessed their conformity with the YSI analyzer and other leading systems. Participants (n = 101), 18 years of age were included, covering a diverse demographic. Accuracy was determined using the YSI analyzer as the standard, employing linear regression, a consensus error grid and Bland-Altman analyses. The performance of both system users showed 100% conformity within zone A of the consensus error grid, high linear regression coefficients and minimal bias in the Bland-Altman analysis. Subjective satisfaction analyses showed average scores of 4.59 and 4.62 for BGM200 and BGM300, respectively. Comparison with other systems revealed high regression coefficients (CONTOUR (R) PLUS; R-2 = 0.9960, BGM200; R-2 = 0.9927, BGM300; R-2 = 0.9915, Accu-Chek (R) Guide; R-2 = 0.9910). Both systems demonstrated reasonable accuracy and user performance comparable to competitors' products and met international standards. Their reliability in real-world scenarios and high user satisfaction make them valuable tools for diabetes management.