Current Sources for Electrical Impedance Tomography IMPETOM

被引:0
|
作者
Sanchez, Pablo [1 ,2 ]
David, Marcelo [3 ]
Gonzalez, Mariana [1 ,2 ]
Morales, Isabel [1 ,2 ]
Simini, Franco [1 ,2 ]
机构
[1] Univ Republica, Fac Med, Nucleo Ingn Biomed, Montevideo, Uruguay
[2] Univ Republica, Fac Ingn, Nucleo Ingn Biomed, Montevideo, Uruguay
[3] Jerusalem Coll Technol, Lev Acad Ctr, Dept Elect Engn, Jerusalem, Israel
关键词
Electrical impedance tomography; bandwidth; total harmonic distortion; voltage-controlled current source; step response electrical impedance tomography;
D O I
10.1007/978-3-031-61960-1_23
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Reliable current sources are imperative in Electrical Impedance Tomography (EIT) as it requires signals with minimum distortion and relatively broad bandwidth. In this work four current sources are compared, Howland modified amongst them, to select one for IMPETOM, our clinical EIT system. Bandwidth analysis (0.5 Hz-10 MHz), total harmonic distortion (up to 1.5%) and output current (1 mA to 5 mA) are studied for four current source designs. All configurations gave good results with bandwidths between 0.5 MHz and 3 MHz, driving loads of between 2 k Omega and 12 k Omega with up to 5 mA peak output currents and 1% maximum total harmonic distortion (THD). Among the options, we found the Negative Feedback current source (NFCS) to have the best performance for IMPETOM.
引用
收藏
页码:233 / 248
页数:16
相关论文
共 50 条
  • [31] A Shifting Current Mirror Driver Circuit for Electrical Impedance Tomography Applications
    Huang, Jiajie
    Wang, Chao
    Zhou, Ting
    Lu, Wangzilu
    Zhao, Yang
    Liu, Yan
    Li, Yongfu
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS, 2023, 70 (10) : 3832 - 3836
  • [32] Robust imaging using electrical impedance tomography: review of current tools
    Brazey, Benoit
    Haddab, Yassine
    Zemiti, Nabil
    PROCEEDINGS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2022, 478 (2258):
  • [33] Posterior covariance related optimal current patterns in electrical impedance tomography
    Kaipio, JP
    Seppänen, A
    Somersalo, E
    Haario, H
    INVERSE PROBLEMS, 2004, 20 (03) : 919 - 936
  • [34] DETERMINATION OF OPTIMUM INJECTED CURRENT PATTERNS IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    KOKSAL, A
    EYUBOGLU, BM
    PHYSIOLOGICAL MEASUREMENT, 1995, 16 : A99 - A109
  • [35] Dominant-Current Deep Learning Scheme for Electrical Impedance Tomography
    Wei, Zhun
    Liu, Dong
    Chen, Xudong
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (09) : 2546 - 2555
  • [36] A floating wide-band current source for electrical impedance tomography
    Khalighi, M.
    Mikaeili, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (08):
  • [37] Bipolar current injection methods for Electrical Impedance Tomography: a comparative study
    Alex, Anitha
    Reddy, Ramasubba M.
    MEDICAL IMAGING 2017: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2017, 10137
  • [38] ACTIVE CURRENT ELECTRODES FOR IN-VIVO ELECTRICAL-IMPEDANCE TOMOGRAPHY
    JOSSINET, J
    TOURTEL, C
    JARRY, R
    PHYSIOLOGICAL MEASUREMENT, 1994, 15 : A83 - A90
  • [39] Electrode Configuration with Howland Current Source and Demodulator for Electrical Impedance Tomography
    Nasiri, Hossein
    Benetti, Rafael
    Takimoto, Rogerio Y.
    Cavalheiro, Andre C. M.
    Duran, Guilherme C.
    Martins, Thiago C.
    Tsuzuki, Marcos S. G.
    IFAC PAPERSONLINE, 2021, 54 (15): : 79 - 84
  • [40] Study of Voltage Control Current Source in Electrical Impedance Tomography System
    Li Zhangyong
    Xu Zhui
    Ren Chaoshi
    Wang Wei
    Zhao Dechun
    Zhang Huiquan
    2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,