Calculation of recovery coefficients for partial volume effect correction in PET/CT imaging using a customized anthropomorphic body phantom

被引:0
|
作者
Yavuz, Gunes [1 ]
Kovan, Bilal [2 ]
Toklu, Turkay [3 ]
Cermik, Tevfik F. [4 ]
Ozturk, Cengizhan [1 ,5 ]
机构
[1] Bogazici Univ, Inst Biomed Engn, Kandilli Campus, TR-34684 Istanbul, Turkiye
[2] Istanbul Univ, Istanbul Fac Med, Dept Nucl Med, Turgut Ozal Millet Cad 118U, TR-34104 Istanbul, Turkiye
[3] Yeditepe Univ, Kosuyolu Hosp, Dept Nucl Med, Kosuyolu Cad 168, TR-34718 Istanbul, Turkiye
[4] Acibadem Maslak Hosp, Dept Obstet & Gynecol, Buyukdere Cad 40, TR-34457 Istanbul, Turkiye
[5] Bogaz Univ, Ctr Targeted Therapy Technol CT3, Kandilli Campus, TR-34685 Istanbul, Turkiye
关键词
Partial volume effect; PVE; Partial volume effect correction; Anthropomorphic phantom; Recovery coefficients; PET/CT; Medical imaging; STRATEGIES;
D O I
10.1186/s12938-025-01330-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Positron Emission Tomography/Computed Tomography (PET/CT) combines metabolic and anatomical information improving the precision and accuracy of oncological diagnostics. The standardized uptake value (SUV) measures tumor metabolism, yet its accuracy is influenced by the partial volume effect (PVE), impacting small lesion detection. This study aims to refine PVE corrections for small lesions using an in-house customized, special anthropomorphic phantom. Scans of this phantom which contained spheres of different sizes were performed across four hospitals at different PET/CT systems from various manufacturers (Siemens and Philips analog PET/CT systems, GE analog and digital PET/CT systems). The phantom contained six custom-designed cylinders with embedded spheres simulating sub-centimeter (0.3, 0.5, 0.9) and centimeter (1.3, 1.9, 2.8) lesions. Scans were performed separately for each sphere in the thorax, abdomen, and pelvis regions at all sites. Recovery Coefficients (RCs) were calculated to correct SUV values, demonstrating that RCs vary by sphere size and anatomical region but not change significantly among scanners. RCs are approaching unity for larger spheres, ensuring accurate SUV measurements. However, small spheres (< 0.5 cm) exhibited significant measurement challenges due to PVE. The anthropomorphic phantom proved effective in obtaining realistic SUV-corrected values, offering a promising tool for enhancing the accuracy and standardization of PET imaging in oncology. This study underscores the necessity for advanced imaging technologies and standardized RC application in clinical settings to improve the quantification of PET imaging, particularly in small lesion detection.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Partial volume correction of PET scans using a segmented high resolution CT or MR scan.
    Svarer, C
    Willendrup, P
    Holm, S
    Knudsen, GM
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (05) : 192P - 192P
  • [32] Influence of partial volume correction in staging of head and neck squamous cell carcinoma using PET/CT
    Fayad, H.
    Le Pogam, A.
    Lamare, F.
    Fernandez, P.
    Pradier, O.
    Valette, G.
    Visvikis, D.
    Le Rest, C. Cheze
    QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 58 (03): : 319 - 328
  • [33] Impact of motion and partial volume effects correction on PET myocardial perfusion imaging using simultaneous PET-MR
    Petibon, Yoann
    Guehl, Nicolas J.
    Reese, Timothy G.
    Ebrahimi, Behzad
    Normandin, Marc D.
    Shoup, Timothy M.
    Alpert, Nathaniel M.
    El Fakhri, Georges
    Ouyang, Jinsong
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (02): : 326 - 343
  • [34] Phantom studies investigating extravascular density imaging for partial volume correction of 3-D PET 18FDG studies
    Wassenaar, RW
    Beanlands, RSB
    DeKemp, RA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 68 - 71
  • [35] Phantom studies investigating extravascular density imaging for partial volume correction of 3D PET 18FDG studies
    Wassenaar, RW
    Beanlands, RSB
    deKemp, RA
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 821 - 824
  • [36] Impact of motion compensation and partial volume correction for 18F-NaF PET/CT imaging of coronary plaque
    Cal-Gonzalez, J.
    Tsoumpas, C.
    Lassen, M. L.
    Rasul, S.
    Koller, L.
    Hacker, M.
    Schaefers, K.
    Beyer, T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (01):
  • [37] Impact of motion compensation and partial volume correction on 18F-NaF PET/CT imaging of coronary plaque
    Cal-Gonzalez, J.
    Tsoumpas, C.
    Lassen, M.
    Rasul, S.
    Hacker, M.
    Schaefers, K.
    Beyer, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2017, 44 : S362 - S362
  • [38] Whole-body motion correction in cardiac PET/CT using Positron Emission Tracking: A phantom validation study
    Manwell, Spencer
    Klein, Ran
    deKemp, Robert
    Xu, Tong
    2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [39] Incorporation of wavelet-based denoising in iterative deconvolution for partial volume correction in whole-body PET imaging
    Boussion, N.
    Le Rest, C. Cheze
    Hatt, M.
    Visvikis, D.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 (07) : 1064 - 1075
  • [40] Incorporation of wavelet-based denoising in iterative deconvolution for partial volume correction in whole-body PET imaging
    N. Boussion
    C. Cheze Le Rest
    M. Hatt
    D. Visvikis
    European Journal of Nuclear Medicine and Molecular Imaging, 2009, 36 : 1064 - 1075