Whole-body motion correction in cardiac PET/CT using Positron Emission Tracking: A phantom validation study

被引:0
|
作者
Manwell, Spencer [1 ,2 ]
Klein, Ran [3 ]
deKemp, Robert [2 ]
Xu, Tong [1 ]
机构
[1] Carleton Univ, Dept Phys, Ottawa, ON, Canada
[2] Univ Ottawa, Cardiac PET Ctr, Heart Inst, Ottawa, ON, Canada
[3] Univ Ottawa, Div Nucl Med, Ottawa, ON, Canada
基金
加拿大健康研究院;
关键词
PATIENT MOTION; STRESS; RB-82; FLOW;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A whole-body motion correction algorithm that is based on positron emission tracking (PeTrack) of fiducial markers has been validated using phantom studies in PET/CT. Patient motion during medical imaging procedures is an inevitable source of artifact and image quality reduction. A motion correction algorithm has been designed to reduce the adverse effects of misalignment artifacts between the CT attenuation map and PET data as well as reduction of spatial resolution associated with motion blurring. A phantom with a left-ventricular cardiac insert was imaged for 5 minutes in list-mode format on a GE Discovery 690 PET/CT scanner. Four acquisitions were completed in total: one without motion and three with motion in the superior-inferior and lateral directions. The list-mode data were processed offline using the PeTrack algorithm to generate 3D motion traces of a 30 kBq Na-22 fiducial marker placed on the anterior surface of the phantom. Whole-body motion triggers were created when a change in marker position, relative to t = 0, exceeded a voxel length in any direction. Motion data were used to rigidly align CT and raw emission data prior to reconstruction as well as align reconstructed PET motion frames. Images with and without motion correction were compared to the static reference image to evaluate changes in apparent left-ventricular (LV) wallthicknesses, signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR). With respect to the Reference image, Non-Corrected images exhibited average increases in LV wallthickness in the range of 9-55%, SNR reductions from 31-63% and CNR reductions of 27-80%. Motion corrected images exhibited average changes in LV wall-thickness of -0.2-6%, SNR reductions of 4-10% and CNR reductions of 13-35%. The authors concluded that motion tracking data acquired using PeTrack can be incorporated into the reconstruction algorithm to correct for rigid motion and significantly reduce the degrading factors of whole-body motion in PET/CT.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Whole-body motion correction in 13N-ammonia myocardial perfusion imaging using positron emission tracking
    Manwell, Spencer
    Klein, Ran
    deKemp, Robert
    Xu, Tong
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [2] Elastic Motion Correction For Continuous Bed Motion Whole-body PET/CT
    Hong, Inki
    Jones, Judson
    Hamill, James
    Michel, Christian
    Casey, Michael
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [3] Benefit of scatter limitation correction for cold artifacts caused by arm motion in whole-body PET/CT: physical phantom study
    Onishi, H.
    Furuta, A.
    Kangai, Y.
    Yada, N.
    Saho, T.
    Amijima, H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S503 - S504
  • [4] Impact of elastic motion correction on quantitation and image quality of whole-body PET/CT
    Meier, Joseph
    Hamill, James
    Jones, Judson
    Hong, Inki
    Mawlawi, Osama
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [5] Simulating respiratory motion in whole-body PET imaging with the MCAT phantom
    Isoardi, RA
    Comtat, C
    Frouin, V
    Delzescaux, T
    Trébossen, R
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1113 - 1115
  • [6] Ultrasensitive and multiplexed tracking of single cells using whole-body PET/CT
    Nguyen, Hieu T. M.
    Das, Neeladrisingha
    Ricks, Matthew
    Zhong, Xiaoxu
    Takematsu, Eri
    Wang, Yuting
    Ruvalcaba, Carlos
    Mehadji, Brahim
    Roncali, Emilie
    Chan, Charles K. F.
    Pratx, Guillem
    SCIENCE ADVANCES, 2024, 10 (24):
  • [7] Whole-body tracking of single cells via positron emission tomography
    Jung, Kyung Oh
    Kim, Tae Jin
    Yu, Jung Ho
    Rhee, Siyeon
    Zhao, Wei
    Ha, Byunghang
    Red-Horse, Kristy
    Gambhir, Sanjiv Sam
    Pratx, Guillem
    NATURE BIOMEDICAL ENGINEERING, 2020, 4 (08) : 835 - 844
  • [8] Effect of patient arm motion in whole-body PET/CT
    Lodge, Martin
    Mhlanga, Joyce
    Wahl, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [9] Effect of Patient Arm Motion in Whole-Body PET/CT
    Lodge, Martin A.
    Mhlanga, Joyce C.
    Cho, Steve Y.
    Wahl, Richard L.
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (12) : 1891 - 1897
  • [10] Whole-body tracking of single cells via positron emission tomography
    Kyung Oh Jung
    Tae Jin Kim
    Jung Ho Yu
    Siyeon Rhee
    Wei Zhao
    Byunghang Ha
    Kristy Red-Horse
    Sanjiv Sam Gambhir
    Guillem Pratx
    Nature Biomedical Engineering, 2020, 4 : 835 - 844