Design, Development, and Characterization of Cross Beam Force Transducer for Low Force Measurement

被引:0
|
作者
Vikas [1 ,2 ]
Titus, S. S. K. [1 ]
Kumar, R. [1 ]
Gautam, S. K. [1 ]
机构
[1] Natl Phys Lab, Force & Hardness Metrol Sect, CSIR, Dr KS Krishnan Marg, New Delhi 110012, India
[2] Acad Sci & Innovat Res AcSIR, Human Resource Dev Ctr, CSIR, Ghaziabad 201002, UP, India
来源
关键词
Low Force Measurement; Cross Beam; Finite Element Analysis; Metrology; Strain Gauge; NONLINEARITY;
D O I
10.1007/s12647-024-00784-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This study addresses the growing need for precise static low-force measurement in industries and metrology, driven by technological advancements and a trend toward miniaturization. Existing commercially available static force transducers can do force measurements with precision but these transducers face significant challenges in their development for low force measurement due to the irregular shapes of the spring element. Additionally, in the literature, some low-force sensors are also described for static low-force measurement but they possess poor metrology characteristics such as linearity, repeatability, and reproducibility which impacts their overall performance. The use of such low-force sensors in static low force measurements related applications such as microhardness testing, tactile sensing, and automation assembly systems may lead to low product trustworthiness. To address these challenges, this study presents the design of a simple-shaped cross-beam low-force transducer specifically for static force measurement in the 5 N range. The study begins with the design of cross beam spring element whose deflection under 5 N is analyzed by analytical and computational means. For the computational analysis, Ansys software of Finite Element Analysis (FEA) is used. For the analytical analysis, an analytical expression is derived to study the deflection. Experimental results demonstrate that the developed cross-beam strain gauge force transducer exhibits good metrological performance in the low-force range.
引用
收藏
页码:127 / 137
页数:11
相关论文
共 50 条
  • [41] DESIGN AND DEVELOPMENT OF AN INSTRUMENT FOR MEASUREMENT OF BITING FORCE IN HUMAN BEINGS
    Vaz, Anand
    Jha, Mayank Shekhar
    Seth, Rishi
    Saxena, Ambuj
    PROCEEDINGS OF THE ASME 11TH BIENNIAL CONFERENCE ON ENGINEERING SYSTEMS DESIGN AND ANALYSIS, VOL 4, 2012, : 227 - 232
  • [42] Design, development and testing of a turning dynamometer for cutting force measurement
    Yaldiz, Sueleyman
    Uensacar, Faruk
    MATERIALS & DESIGN, 2006, 27 (10): : 839 - 846
  • [43] Blocked force measurement of an electro-active paper actuator using a cantilevered force transducer
    Yun, Gyu-young
    Kim, Heung Soo
    Kim, Jaehwan
    SMART MATERIALS AND STRUCTURES, 2008, 17 (02)
  • [44] Design of a three-component capacitive force transducer
    Cannella, M
    Marinozzi, F
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (08): : 3411 - 3417
  • [45] The design of a beam with controlled force units
    Kolpakov, AG
    PMM JOURNAL OF APPLIED MATHEMATICS AND MECHANICS, 2005, 69 (01): : 89 - 99
  • [46] The design of a beam with controlled force units
    Kolpakov, A.G.
    Journal of Applied Mathematics and Mechanics, 2005, 69 (01): : 89 - 99
  • [47] Design of a Beam with Controllable Force Elements
    A. G. Kolpakov
    Journal of Applied Mechanics and Technical Physics, 2003, 44 (4) : 584 - 595
  • [48] Development and Metrological Evaluation of a Force Transducer for Industrial Application
    Singh, Ajay Pratap
    Ghoshal, Sanjoy K.
    Kumar, Harish
    Yoon, Jonghun
    Yadav, Poonam
    IEEE ACCESS, 2021, 9 : 33299 - 33312
  • [49] DEVELOPMENT AND CALIBRATION OF A DYNAMIC-CONTACT-FORCE TRANSDUCER
    GOLDAR, D
    SETHI, VS
    KHURANA, OP
    VERMA, SR
    EXPERIMENTAL MECHANICS, 1984, 24 (03) : 187 - 191
  • [50] Simple, effective and high sensitivity transducer for force measurement of myofibrils
    Phan, Kien Nguyen
    Kobayashi, Takakazu
    SENSORS AND ACTUATORS A-PHYSICAL, 2008, 144 (02) : 389 - 393