Design, Development, and Characterization of Cross Beam Force Transducer for Low Force Measurement

被引:0
|
作者
Vikas [1 ,2 ]
Titus, S. S. K. [1 ]
Kumar, R. [1 ]
Gautam, S. K. [1 ]
机构
[1] Natl Phys Lab, Force & Hardness Metrol Sect, CSIR, Dr KS Krishnan Marg, New Delhi 110012, India
[2] Acad Sci & Innovat Res AcSIR, Human Resource Dev Ctr, CSIR, Ghaziabad 201002, UP, India
来源
关键词
Low Force Measurement; Cross Beam; Finite Element Analysis; Metrology; Strain Gauge; NONLINEARITY;
D O I
10.1007/s12647-024-00784-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This study addresses the growing need for precise static low-force measurement in industries and metrology, driven by technological advancements and a trend toward miniaturization. Existing commercially available static force transducers can do force measurements with precision but these transducers face significant challenges in their development for low force measurement due to the irregular shapes of the spring element. Additionally, in the literature, some low-force sensors are also described for static low-force measurement but they possess poor metrology characteristics such as linearity, repeatability, and reproducibility which impacts their overall performance. The use of such low-force sensors in static low force measurements related applications such as microhardness testing, tactile sensing, and automation assembly systems may lead to low product trustworthiness. To address these challenges, this study presents the design of a simple-shaped cross-beam low-force transducer specifically for static force measurement in the 5 N range. The study begins with the design of cross beam spring element whose deflection under 5 N is analyzed by analytical and computational means. For the computational analysis, Ansys software of Finite Element Analysis (FEA) is used. For the analytical analysis, an analytical expression is derived to study the deflection. Experimental results demonstrate that the developed cross-beam strain gauge force transducer exhibits good metrological performance in the low-force range.
引用
收藏
页码:127 / 137
页数:11
相关论文
共 50 条
  • [31] DEVELOPMENT AND USE OF A FORCE-MOMENT TRANSDUCER
    GUPTA, BK
    EDWARDS, AD
    LOVEDAY, RW
    MAGAZINE OF CONCRETE RESEARCH, 1968, 20 (64) : 177 - &
  • [32] Design of force measurement module for force touch screens
    Lee, Yong-Min
    Lee, Sangjoon
    IEICE ELECTRONICS EXPRESS, 2017, 14 (20):
  • [33] DEVELOPMENT OF A CYTOMIC FORCE TRANSDUCER FOR EXPERIMENTAL MECHANOBIOLOGY
    Dy, E.
    Ho, C. M.
    IEEE 22ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS 2009), 2009, : 391 - 394
  • [34] Development and Metrological Evaluation of an Industrial Force Transducer
    Singh, A. P.
    Ghoshal, S. K.
    Kumar, H.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2020, 35 (02): : 165 - 176
  • [35] CAPACITIVE TRANSDUCER FOR CONTINUOUS MEASUREMENT OF VERTICAL FOOT FORCE
    MIYAZAKI, S
    ISHIDA, A
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (04) : 309 - 316
  • [36] Capacitive transducer for continuous measurement of vertical foot force
    Miyazaki, S.
    Ishida, A.
    Medical and Biological Engineering and Computing, 1984, 22 (04): : 309 - 316
  • [37] Measurement of trocar insertion force using a piezoelectric transducer
    Ng, PS
    Sahota, DS
    Yuen, PM
    JOURNAL OF THE AMERICAN ASSOCIATION OF GYNECOLOGIC LAPAROSCOPISTS, 2003, 10 (04): : 534 - 538
  • [38] Design and development of the Scanning Force Microscope for imaging and force measurement with sub-nanonewton resolution
    Brar, LK
    Paranjape, M
    Guha, A
    Raychaudhuri, AK
    CURRENT SCIENCE, 2002, 83 (10): : 1197 - 1199
  • [39] Measurement of low friction force
    Ando, Y
    JOURNAL OF JAPANESE SOCIETY OF TRIBOLOGISTS, 1998, 43 (07) : 587 - 592
  • [40] Measurement of Low Friction Force
    Ando, Yasuhisa
    Toraibarojisuto/Journal of Japanese Society of Tribologists, 1998, 43 (07): : 587 - 592