A rare case of endotracheal tube cuff leakage with no detectable decrease in cuff pressure

被引:0
|
作者
Nagasawa, Keiichi [1 ]
Nishibata, Masayuki [1 ]
Luthe, Sarah Kyuragi [1 ]
Kawamata, Tomoyuki [1 ]
机构
[1] Wakayama Med Univ, Dept Anesthesiol, Sch Med, Wakayama 6418510, Japan
来源
JA CLINICAL REPORTS | 2024年 / 10卷 / 01期
关键词
Endotracheal tube; Cuff leak; Pilot balloon tubing; Fiber-optic light; Surgical fire; Ignition source;
D O I
10.1186/s40981-024-00754-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundCommon causes of air leakage around an endotracheal tube include insufficient endotracheal tube cuff inflation and damage to the cuff, while damage to the pilot balloon or pilot balloon tubing is relatively rare.Case presentationA 74-year-old female with vertebral osteomyelitis was scheduled for an extreme lateral interbody fusion followed posterior fixation. A fiber-optic light was utilized as part of the surgical illuminator. A sudden decrease in tidal volume and airway pressure was noted intraoperatively. We suspected leakage around the endotracheal tube cuff; however, no decrease in cuff pressure was detected. Despite the normal cuff pressure, we decided to inject a small amount of air which led to a significant increase in the cuff pressure. Upon careful inspection of the endotracheal tube, we discovered that the pilot balloon tubing was damaged as a result of thermal energy emitted by the fiber-optic light, which had ignited the surgical drape. The pilot balloon tubing was partially severed in which the section proximal to the endotracheal tube cuff was burned and punctured, causing the cuff leak. Meanwhile, the section proximal to the pilot balloon had melted and occluded the lumen, resulting in a falsely normal cuff pressure reading followed by an elevated cuff pressure when a small amount of air was injected into the pilot balloon during troubleshooting. Appropriate ventilation was resumed after extubation and re-intubation with a new endotracheal tube.ConclusionsWe experienced an endotracheal tube cuff leakage caused by a damaged pilot balloon tubing due to thermal energy of the fiber-optic light. Our case report emphasizes the importance of suspecting damage to the endotracheal tube cuff and inflation system despite a normal cuff pressure reading, given that the measurement may be falsely elevated depending on the specific location of the damage. In addition, all operating personnel should be familiarized with safety warnings and cautions related to handling.
引用
收藏
页数:4
相关论文
共 50 条
  • [31] Endotracheal Tube Cuff Pressure Monitoring in Intensive Care Units
    Zand, F.
    Nekooeian, A. A.
    Rohani, M.
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2008, 10 (03) : 220 - 224
  • [32] IS THE PRESSURE MEASUREMENT IN THE ENDOTRACHEAL TUBE CUFF A STANDARD? A SIMULATION STUDY
    Jedrzejczyk-Cwanek, Marzena
    Wozniak, Krystyna
    Niemczyk, Edyta
    Medrzycka-Dabrowska, Wioletta
    Ozga, Dorota
    JOURNAL OF EMERGENCY MEDICINE, 2018, 55 (01): : E25 - E26
  • [33] Continuous real time endotracheal tube cuff pressure waveform
    Ganigara, Anuradha
    Ramavakoda, Chandrika Y.
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2014, 28 (04) : 433 - 434
  • [34] Disposable Airway Pressure Manometers for Endotracheal Tube Cuff Inflation
    Klonner, Moriz Ettore
    Mattaliano, Giorgio
    Casoria, Vincenzo
    Vogl, Claus
    Braun, Christina
    ANIMALS, 2023, 13 (03):
  • [35] Continuous real time endotracheal tube cuff pressure waveform
    Anuradha Ganigara
    Chandrika Y. Ramavakoda
    Journal of Clinical Monitoring and Computing, 2014, 28 : 433 - 434
  • [36] Elevated endotracheal tube cuff pressure in the pediatric emergency department
    Ferenczy, Edward D.
    Stoner, Michael J.
    Spencer, Sandra P.
    Gee, Samantha W.
    Scherzer, Daniel J.
    Tobias, Joseph D.
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2018, 113 : 289 - 291
  • [37] Accuracy of an inexpensive, disposable endotracheal tube cuff pressure manometer
    Frakes, Michael A.
    Pylant, Rogers T.
    Williams, Anthony
    CRITICAL CARE MEDICINE, 2007, 35 (12) : A229 - A229
  • [38] ENDOTRACHEAL-TUBE CUFF PRESSURE - IS A SYRINGE A RELIEF VALVE
    GUYTON, DC
    BEARD, HR
    ANESTHESIOLOGY, 1994, 81 (3A) : A496 - A496
  • [39] Endotracheal tube cuff pressure changes during manual cuff pressure control manoeuvres: An in-vitro assessment
    Aeppli, Norbert
    Lindauer, Bastian
    Steurer, Marc P.
    Weiss, Markus
    Dullenkopf, Alexander
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2019, 63 (01) : 55 - 60
  • [40] The interaction of subglottic drainage, cuff pressure, and oral care on endotracheal tube fluid leakage: A benchtop study
    Chair, Sek Ying
    Chan, David Wing Keung
    Cao, Xi
    AUSTRALIAN CRITICAL CARE, 2020, 33 (04) : 358 - 363