The interaction of subglottic drainage, cuff pressure, and oral care on endotracheal tube fluid leakage: A benchtop study

被引:7
|
作者
Chair, Sek Ying [1 ]
Chan, David Wing Keung [2 ]
Cao, Xi [1 ]
机构
[1] Chinese Univ Hong Kong, Fac Med, Nethersole Sch Nursing, Shatin, Hong Kong, Peoples R China
[2] Caritas Inst Higher Educ, Sch Hlth Sci, Hong Kong, Peoples R China
关键词
Critical care; Ventilator-associated pneumonia; Endotracheal tube; Subglottic secretion drainage; Fluid leakage; VENTILATOR-ASSOCIATED PNEUMONIA; SECRETION DRAINAGE; EXPIRATORY PRESSURE; TRACHEAL CUFF; PREVENTION; MICROASPIRATION; BUNDLE;
D O I
10.1016/j.aucc.2019.05.002
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The types of endotracheal tube, positive end-expiratory pressure (PEEP) level, endotracheal tube cuff pressure level, and nursing activity may influence the occurrence of pulmonary aspiration in ventilated patients with an endotracheal tube, but the evidence on their degree of influence is still inconclusive. Aim: The aim of this study was to examine the effect of endotracheal tubes with or without subglottic secretion drainage on fluid leakage across endotracheal tube cuffs under different PEEP levels, different cuff pressures, and with or without oral care. Method: This was a benchtop study with a multifactorial experimental design using two sets of airway. The two types of endotracheal tubes were tested through 12 scenarios with 240 combinations of different PEEP levels (0, 5, and 10 cmH(2)O), endotracheal tube cuff pressures (15 and 25 cmH(2)O), and oral care (with or without) using two sets of airway manikins. Each scenario lasted for 30 min, and fluid leakage was measured at the end of each scenario. Generalised Linear Model test was used to analyse fluid leak at 30 min (with and without interaction effect). Results: A total of 100 cases showed fluid leakage, with more fluid leakage occurring in low cuff pressure (15 cmH(2)O), no PEEP, and with oral care. Results of the Generalised Linear Model revealed that endotracheal tubes with subglottic secretion drainage, high PEEP (at 10 cmH(2)O), normal cuff pressure (at 25 cmH(2)O), and no oral care demonstrated a significant effect in reducing fluid leakage than endotracheal tubes without subglottic secretion drainage, low PEEP (at 5 or 0 cmH(2)O), and low cuff pressure (at 15 cmH(2)O) (all p < 0.001). However, only the interaction effect of endotracheal tubes with subglottic secretion drainage*high PEEP showed a significant effect on fluid leakage (p < 0.001), with the combination of endotracheal tube*no PEEP producing the greatest volume of fluid leak. Conclusion: Using endotracheal tubes with subglottic secretion drainage, high PEEP, and normal cuff pressure and avoiding excessive endotracheal tube movement during oral care reduced fluid leakage. This study provided strong evidence to inform practice on reducing microaspiration in ventilated patients. (C) 2019 Australian College of Critical Care Nurses Ltd. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:358 / 363
页数:6
相关论文
共 50 条
  • [1] Influence of an endotracheal tube with polyurethane cuff and subglottic secretion drainage on pneumonia
    Lorente, Leonardo
    Lecuona, Maria
    Jimenez, Alejandro
    Mora, Maria L.
    Sierra, Antonio
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2007, 176 (11) : 1079 - 1083
  • [2] Leakage of fluid past the tracheal tube cuff in a benchtop model
    Young, PJ
    Rollinson, M
    Downward, G
    Henderson, S
    BRITISH JOURNAL OF ANAESTHESIA, 1997, 78 (05) : 557 - 562
  • [3] A rare case of endotracheal tube cuff leakage with no detectable decrease in cuff pressure
    Nagasawa, Keiichi
    Nishibata, Masayuki
    Luthe, Sarah Kyuragi
    Kawamata, Tomoyuki
    JA CLINICAL REPORTS, 2024, 10 (01):
  • [4] Assessment of the effectiveness of a ventilator associated pneumonia prevention bundle that contains endotracheal tube with subglottic drainage and cuff pressure monitorization
    Akdogan, Ozlem
    Ersoy, Yasemin
    Kuzucu, Cigdem
    Gedik, Ender
    Togal, Turkan
    Yetkin, Funda
    BRAZILIAN JOURNAL OF INFECTIOUS DISEASES, 2017, 21 (03): : 276 - 281
  • [5] Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model
    Ouanes, Islem
    Lyazidi, Aissam
    Danin, Pierre Eric
    Rana, Nerlep
    Di Bari, Annalisa
    Abroug, Fekri
    Louis, Bruno
    Brochard, Laurent
    INTENSIVE CARE MEDICINE, 2011, 37 (04) : 695 - 700
  • [6] Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model
    Islem Ouanes
    Aissam Lyazidi
    Pierre Eric Danin
    Nerlep Rana
    Annalisa Di Bari
    Fekri Abroug
    Bruno Louis
    Laurent Brochard
    Intensive Care Medicine, 2011, 37 : 695 - 700
  • [7] Cuff leakage caused by automatic cuff pressure monitor and water condensation in the inflation tube of Microcuff® endotracheal tube
    Matsuba, Kohei
    Tsujimoto, Takafumi
    Kawaguchi, Masahiko
    Inoue, Satoki
    ANAESTHESIA AND INTENSIVE CARE, 2019, 47 (03) : 300 - 301
  • [8] Endotracheal Tube Cuff Pressure Monitoring in Intensive Care Units
    Zand, F.
    Nekooeian, A. A.
    Rohani, M.
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2008, 10 (03) : 220 - 224
  • [9] A newly developed endotracheal tube offering 'pressurised sealing' outperforms currently available tubes in preventing cuff leakage: a benchtop study
    Spapen, Herbert
    Suys, Emiel
    Stiers, Wim
    Smet, Geertde
    Diltoer, Marc
    Regt, Joukede
    Honore, Patrick M.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2015, 59 : 29 - 29
  • [10] Leakage of fluid through an endotracheal tube cuff during spontaneous and intermittent positive pressure ventilation: An in vitro comparison
    Komai, M.
    Sugiura, T.
    Yoshida, A.
    Aoyama, Y.
    Kobayashi, Y.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2010, 27 (01) : 267 - 267