A rare case of endotracheal tube cuff leakage with no detectable decrease in cuff pressure

被引:0
|
作者
Nagasawa, Keiichi [1 ]
Nishibata, Masayuki [1 ]
Luthe, Sarah Kyuragi [1 ]
Kawamata, Tomoyuki [1 ]
机构
[1] Wakayama Med Univ, Dept Anesthesiol, Sch Med, Wakayama 6418510, Japan
来源
JA CLINICAL REPORTS | 2024年 / 10卷 / 01期
关键词
Endotracheal tube; Cuff leak; Pilot balloon tubing; Fiber-optic light; Surgical fire; Ignition source;
D O I
10.1186/s40981-024-00754-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundCommon causes of air leakage around an endotracheal tube include insufficient endotracheal tube cuff inflation and damage to the cuff, while damage to the pilot balloon or pilot balloon tubing is relatively rare.Case presentationA 74-year-old female with vertebral osteomyelitis was scheduled for an extreme lateral interbody fusion followed posterior fixation. A fiber-optic light was utilized as part of the surgical illuminator. A sudden decrease in tidal volume and airway pressure was noted intraoperatively. We suspected leakage around the endotracheal tube cuff; however, no decrease in cuff pressure was detected. Despite the normal cuff pressure, we decided to inject a small amount of air which led to a significant increase in the cuff pressure. Upon careful inspection of the endotracheal tube, we discovered that the pilot balloon tubing was damaged as a result of thermal energy emitted by the fiber-optic light, which had ignited the surgical drape. The pilot balloon tubing was partially severed in which the section proximal to the endotracheal tube cuff was burned and punctured, causing the cuff leak. Meanwhile, the section proximal to the pilot balloon had melted and occluded the lumen, resulting in a falsely normal cuff pressure reading followed by an elevated cuff pressure when a small amount of air was injected into the pilot balloon during troubleshooting. Appropriate ventilation was resumed after extubation and re-intubation with a new endotracheal tube.ConclusionsWe experienced an endotracheal tube cuff leakage caused by a damaged pilot balloon tubing due to thermal energy of the fiber-optic light. Our case report emphasizes the importance of suspecting damage to the endotracheal tube cuff and inflation system despite a normal cuff pressure reading, given that the measurement may be falsely elevated depending on the specific location of the damage. In addition, all operating personnel should be familiarized with safety warnings and cautions related to handling.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Cuff leakage caused by automatic cuff pressure monitor and water condensation in the inflation tube of Microcuff® endotracheal tube
    Matsuba, Kohei
    Tsujimoto, Takafumi
    Kawaguchi, Masahiko
    Inoue, Satoki
    ANAESTHESIA AND INTENSIVE CARE, 2019, 47 (03) : 300 - 301
  • [2] Endotracheal tube cuff pressure and manometry
    Garcia, E.
    Salazar, M. F.
    Gil, S.
    Arizaga, A.
    REVISTA ESPANOLA DE ANESTESIOLOGIA Y REANIMACION, 2014, 61 (09): : 530 - 530
  • [3] Nitrous oxide and endotracheal tube cuff pressure
    Kanter, GJ
    Robbins, LD
    ANESTHESIOLOGY, 2000, 93 (3A) : U165 - U165
  • [4] Endotracheal tube cuff pressure: An overlooked risk
    Al-metwalli, Roshdi R.
    Alghamdi, Talal M.
    Fallatah, Summayah M.
    ANAESTHESIA PAIN & INTENSIVE CARE, 2021, 25 (01) : 86 - 95
  • [5] Endotracheal tube cuff pressure is unpredictable in children
    Felten, MAL
    Schmautz, E
    Delaporte-Cerceau, S
    Orliaguet, GA
    Carli, PA
    ANESTHESIA AND ANALGESIA, 2003, 97 (06): : 1612 - 1616
  • [6] Intraoperative endotracheal tube cuff pressure leak
    Aneejwal, Himanshu
    Sethi, Priyanka
    Kaur, Manbir
    ANAESTHESIA PAIN & INTENSIVE CARE, 2022, 26 (02) : 270 - 270
  • [7] Tracheobronchomegaly A Rare Cause of Endotracheal Tube Cuff Leak
    Ng, Joshua B.
    Bittner, Edward A.
    ANESTHESIOLOGY, 2011, 114 (05) : 1211 - 1211
  • [8] Endotracheal tube cuff leakage after termination of nitrous oxide
    Nomura, T
    Asano, M
    Shibata, M
    Ishi, M
    ANESTHESIOLOGY, 1999, 91 (3A) : U277 - U277
  • [9] UNUSUAL SOURCE OF LEAKAGE FROM CUFF OF AN ENDOTRACHEAL-TUBE
    WONG, RM
    ANAESTHESIA AND INTENSIVE CARE, 1977, 5 (04) : 389 - 389
  • [10] Endotracheal tube cuff pressure changes during manual cuff pressure control manoeuvres: A call for continuous cuff pressure regulation?
    Maertens, Bert
    Blot, Stijn
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2019, 63 (05) : 700 - 701